Prospective, controlled, single-surgeon, single-center post-market clinical follow up study to investigate the clinical outcomes of a hydrophobic trifocal IOL (PhysIOL POD F GF)
Looking for future studies?
Notify Me45 year and older
All sexes
Interventional
Not applicable
Brest, 29609, France
This clinical investigation is a prospective, controlled, single-surgeon, single-center post-market clinical follow up study whereby patients undergoing routine cataract surgery will have bilateral implantation of trifocal intraocular lenses FineVision POD F GF (PhysIOL, Liège, Belgium).
The study purpose is to obtain clinical data on visual acuity, contrast sensitivity, questionnaire outcomes and PCO rate on patients implanted with FineVision POD F GF.
The device under investigation (FineVision POD F GF) is a trifocal glistening-free hydrophobic acrylic intraocular lens (IOL) manufactured by the sponsor of this study PhysIOL sa/nv. The IOL will be implanted as part of the routine cataract surgery on patients suffering from cataract development.
In total 25 patients will be recruited for this clinical study and receive a bilateral implantation of FineVision POD F GF intraocular lens.
Subjects participating in the trial will attend a total of 8 study visits (1 preoperative, 2 operative and 5 postoperative) over a period of 24 months.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Implantation of trifocal IOL POD F GF consisting of hydrophobic material
Time frame: 3 months postoperative
Statistically significant equality between literature data and data obtained in this study on binocular UDVA. UDVA is measured with ETDRS charts placed in 4m distance according to ISO 11979-7:2014
Time frame: 3 months postoperative
The manifested refraction is measured by means of a phoropter. The data contains values for sphere, cylinder and axis of cylinder according to ISO 11979-7:2014. This data will also be used to calculate the manifested refractive spherical equivalent (MRSE)
Time frame: 3 months postoperative
UDVA is measured with ETDRS charts placed in 4m distance according to ISO 11979-7:2014. This assessment is done monocularly and binocularly.
Time frame: 3 months postoperative
CDVA is measured with ETDRS charts placed in 4m distance with best aided corrective glasses according to ISO 11979-7:2014. This assessment is done monocularly and binocularly.
Time frame: 3 months postoperative
DCIVA is measured with ETDRS charts placed in 70cm distance with corrective glasses for far distance according to ISO 11979-7:2014. This assessment is done monocularly and binocularly.
Time frame: 3 months postoperative
UIVA is measured with ETDRS charts placed in 70cm distance. The examination is done without corrective glasses and according to ISO 11979-7:2014. This assessment is done monocularly and binocularly.
Time frame: 3 months postoperative
DCNVA is measured with ETDRS charts placed in 35cm distance with corrective glasses for far distance according to ISO 11979-7:2014. This assessment is done monocularly and binocularly.
Time frame: 3 months postoperative
UNVA is measured with ETDRS charts placed in 35cm distance. The examination is done without corrective glasses and according to ISO 11979-7:2014. This assessment is done monocularly and binocularly.
Time frame: 3 months postoperative
Contrast Sensitivity under photopic and mesopic light conditions using the standardized contrast sensitivity device CSV-1000 (VectorVision)
Time frame: 3 months postoperative
Aberrometry outcomes are measured with a standard aberrometer. The following values will be evaluated in this study: Spherical aberrations, high order aberrations, lens tilt.
Time frame: 2 years postoperative
The rate of patients showing posterior capsular opacities (PCO) long term after surgery will be assessed. The criteria is, if the eye requires a secondary treatment to remove the PCO.
Time frame: 3 months postoperative
Outcomes measures of a questionnaire to address the general patient satisfaction and possible side effects of the treatment. For this study, the validated and verified questionnaire VFQ-25 (National Eye Institute) will be used. The maximum score for each question is 100.
Beaver-Visitec International, Inc.
Industry
Clinical Study to Investigate Visual Performance, Patient Satisfaction and PCO Rate After Implantation of a Trifocal Hydrophobic IOL
Acronym: PHY1703
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT04699266
Cataract, Eye Diseases
Miyata, Japan
View Trial DetailsNCT03306355
Cataract, Eye Diseases
Makati City, Philippines
View Trial DetailsNCT03974451
Cataract, Eye Diseases
Alexandria, Egypt
View Trial DetailsNCT03688425
Cataract, Eye Diseases
Makati City, Philippines
View Trial Details