Asian Eye Institute
Makati City, 1200, Philippines
NCT Number: NCT03688425
Prospective, randomised, controlled, single-surgeon, single-center clinical study to compare the clinical outcomes of two trifocal IOLs differing in the dominance of additional power. The investigational device POD L GF shows dominance for the intermediate addition (+1.75 D), whereas the control device POD F GF shows dominance for the near addition (+3.5 D). Implantation of the IOLs is bilaterally.
Looking for future studies?
Notify Me40 year and older
All sexes
Interventional
Not applicable
Makati City, 1200, Philippines
This is a prospective, randomised, controlled, single-surgeon, single-center clinical study whereby patients undergoing routine cataract surgery will have bilateral implantation of trifocal intraocular lenses. The patients will either be implanted with the hydrophobic IOL FineVision POD L GF or the hydrophobic IOL FineVision POD F GF (both lenses: (PhysIOL, Liège, Belgium).
The devices under investigation (FineVision POD L GF and POD F GF) are trifocal glistening-free acrylic intraocular lenses (IOLs) manufactured by the sponsor of this study PhysIOL sa/nv. Both consist of the same hydrophobic material. The main difference between the lenses is the dominance of additional power. The investigational device POD L GF shows dominance for the intermediate addition (+1.75 D), whereas the control device POD F GF shows dominance for the near addition (+3.5 D). The IOLs will be implanted as part of the routine cataract surgery on patients suffering from cataract development.
Subjects participating in the trial will attend a total of 10 study visits (1 preoperative, 2 operative and 7 postoperative) over a period of 6 months. Subjects would have the option for unscheduled visits if required medically.
Primary and secondary endpoint data will be collected at the 6 months follow up visit. Data analyses will be done after the last patient finished the final examination to support the study publication plan.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Implantation of trifocal IOL POD F GF consisting of light distribution order: far > intermediate > near
Implantation of trifocal IOL POD F GF consisting of light distribution order: far > near > intermediate
Time frame: 3 months postoperative
No statistically significant difference between the two study groups on monocular CDVA at 3 months follow up visit. A significance level of 0.05 or lower (p < 0.05) will be considered statistically significant. CDVA is measured with ETDRS charts placed in 4m distance with best aided corrective glasses according to ISO 11979-7:2014.
Time frame: 3 months postoperative
The manifested refraction is measured by means of a phoropter. The data contains values for sphere, cylinder and axis of cylinder according to ISO 11979-7:2014. This data will also be used to calculate the manifested refractive spherical equivalent (MRSE)
Time frame: 3 months postoperative
UDVA is measured with ETDRS charts placed in 4m distance according to ISO 11979-7:2014. This assessment is done monocularly and binocularly under photopic light conditions.
Time frame: 6 months postoperative
UDVA is measured with ETDRS charts placed in 4m distance according to ISO 11979-7:2014. This assessment is done monocularly under mesopic light conditions.
Time frame: 3 months postoperative
CDVA is measured with ETDRS charts placed in 4m distance with best aided corrective glasses according to ISO 11979-7:2014. This assessment is done binocularly under photopic light conditions.
Time frame: 6 months postoperative
CDVA is measured with ETDRS charts placed in 4m distance with best aided corrective glasses according to ISO 11979-7:2014. This assessment is done monocularly under mesopic light conditions.
Time frame: 3 months postoperative
DCIVA is measured with ETDRS charts placed in 70cm distance with corrective glasses for far distance according to ISO 11979-7:2014. This assessment is done monocularly and binocularly under photopic light conditions.
Time frame: 6 months postoperative
DCIVA is measured with ETDRS charts placed in 70cm distance with corrective glasses for far distance according to ISO 11979-7:2014. This assessment is done monocularly under mesopic light conditions.
Time frame: 3 months postoperative
UIVA is measured with ETDRS charts placed in 70cm distance. The examination is done without corrective glasses and according to ISO 11979-7:2014. This assessment is done monocularly and binocularly under photopic light conditions.
Time frame: 6 months postoperative
UIVA is measured with ETDRS charts placed in 70cm distance. The examination is done without corrective glasses and according to ISO 11979-7:2014. This assessment is done monocularly under mesopic light conditions.
Time frame: 3 months postoperative
DCNVA is measured with ETDRS charts placed in 35cm distance with corrective glasses for far distance according to ISO 11979-7:2014. This assessment is done monocularly and binocularly under photopic light conditions.
Time frame: 6 months postoperative
DCNVA is measured with ETDRS charts placed in 35cm distance with corrective glasses for far distance according to ISO 11979-7:2014. This assessment is done monocularly under mesopic light conditions.
Time frame: 3 months postoperative
UNVA is measured with ETDRS charts placed in 35cm distance. The examination is done without corrective glasses and according to ISO 11979-7:2014. This assessment is done monocularly and binocularly under photopic light conditions.
Time frame: 6 months postoperative
UNVA is measured with ETDRS charts placed in 35cm distance. The examination is done without corrective glasses and according to ISO 11979-7:2014. This assessment is done monocularly under mesopic light conditions.
Time frame: 3 months postoperative
Contrast Sensitivity under photopic and mesopic light conditions
Time frame: 3 months postoperative
Aberrometry outcomes are measured with a standard aberrometer. The following values will be evaluated in this study: Spherical aberrations, high order aberrations, lens tilt.
Time frame: 3 months postoperative
Measurement of Ocular Scatter Index (OSI) score to evaluate and compare the scattering of the implanted lenses.
Time frame: 3 months postoperative
Outcomes measures of a questionnaire to address the general patient satisfaction and possible side effects of the treatment. For this study, the validated and verified questionnaire VFQ-25 (National Eye Institute) will be used. The maximum score for each question is 100.
Time frame: 1 month postoperative
To assess the visual acuity for different distances, defocus curves under photopic light conditions are measured. This test is performed with best distance corrected refraction and spherical additions ranging from -5.0 D to +1.5 D. This examination is performed monocularly.
Time frame: 3 months postoperative
To assess the visual acuity for different distances, defocus curves under photopic light conditions are measured. This test is performed with best distance corrected refraction and spherical additions ranging from -5.0 D to +1.5 D. This examination is performed binocularly.
Beaver-Visitec International, Inc.
Industry
Clinical Study to Compare Clinical Outcomes of Trifocal IOLs FineVision POD L GF and FineVision POD F GF After Bilateral Implantation in Asian Eyes
Acronym: PHY1802
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT04699266
Cataract, Eye Diseases
Miyata, Japan
View Trial DetailsNCT03306355
Cataract, Eye Diseases
Makati City, Philippines
View Trial DetailsNCT03974451
Cataract, Eye Diseases
Alexandria, Egypt
View Trial DetailsNCT03306342
Cataract, Eye Diseases
Brest, France
View Trial Details