NCT Number: NCT02293122
The purpose of this study is to evaluate the precision and agreement of the study device, a specular microscope, in comparison to a currently commercially available specular microscope in establishing substantial equivalence of the two devices in clinical use for diagnostic purposes of the human cornea.
Looking for future studies?
Notify Me18 year–80 year
All sexes
Observational
Columbus, Ohio, 43210, United States
The observational clinical investigation is intended to measure agreement between two diagnostic instruments of the eye's corneal endothelium. The study device is the EM-3000 specular microscope and the control device is the NonCon Robo Pachy (F&A) (CellChek XL) specular microscope. Both instruments measure corneal endothelial cell layer parameters including cell density, coefficient of variation, percent hexagonality, and central thickness.
The primary variable is the measurement of agreement between the two instruments for the four parameters described. The secondary variables include repeatability (intra-investigator) and reproducibility both inter-investigator, and inter-instrument.
Additionally, measurements of agreement will be made on the EM-3000 using automated analysis with and without touch-up.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Non-pathologic younger eyes (18-50 yrs) Non-pathologic older eyes (51-80 yrs) Pathologic adult eyes (29-80)
Exclusion criteria
-Pathologic subjects
Intervention is observational to two different diagnostic devices for which comparative readings of the corneal endothelium are being measured; the test device- the EM-3000 Specular Microscope against the Konan Robo-pachy F&A Specular Microscope.
Time frame: 1 day visit per subject.
For Agreement between the two instruments, outcome measures include endothelial cell density (/mm2), coefficient of variation between the two instruments (SD x CD x 10-6, and percent hexagonality of cells in the measured visible field of view between the two instruments.
Time frame: 1 day visit per subject.
To investigate repeatability (intra-investigator) in the EM-3000 specular microscope in the four outcome measures of the corneal endothelium as well as corneal thickness.
To investigate reproducibility (inter-investigator/instrument) of the EM-3000 specular microscope in the four outcome measures of the corneal endothelium and corneal thickness.
Tomey Corporation
Industry
To Investigate Agreement Between the EM-3000 and the Predicate Device NonCon Robo Pachy (F&A) (CellChek XL).
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07693673
Cornea, Corneal Diseases
Sohag, Sohag Governorate, Egypt
View Trial DetailsNCT05204329
Cornea, Corneal Defect
Chicago, Illinois, United States
View Trial DetailsNCT06759207
Cornea, Corneal Diseases
Sohag, Egypt
View Trial DetailsNCT04626583
Cornea, Corneal Defect
Chicago, Illinois, United States
View Trial Details