Guillaume Emeriaud
Montreal, Quebec, H3T 1C5, Canada
NCT Number: NCT02163382
Mechanical ventilation permits to support the work of breathing in case of respiratory failure, but therapy also has many side effects. Non-invasive ventilation (NIV), which delivers the ventilatory assist via a face mask or nasal canula, permits to decrease these complications. However, NIV is not always successful and half of children in respiratory failure finally require invasive ventilation. A major cause of NIV failure is the ventilator inability to detect patient efforts. The new ventilatory mode NAVA (neurally adjusted ventilatory assist) improves the detection of patient efforts during mechanical ventilation. The hypothesis of this study is that NAVA improves synchrony during pediatric NIV and therefore permits to unload the patient ventilatory efforts.
Looking for future studies?
Notify Me3 day–18 year
All sexes
Interventional
Not applicable
Montreal, Quebec, H3T 1C5, Canada
Nasogastric tube installation to monitor electrical activation of the diaphragm (EAdi):
A specific nasogastric tube equipped with an array of microelectrodes (Edi catheter, Maquet, Solna, Sweden) will be installed after inclusion. This tube has an external aspect and size (6F, 8F, or 12F depending on the patient size) similar to usual nasogastric tubes. The gastric tube installation is extremely frequent in pediatric ICU, and systematic in patients with ventilatory support, to empty the gastric gas and permit the feeding. The new tube will stay in place after the study and could be used as a classic tube. The catheter position in esophagus will be adjusted using a special window of the Servo i ventilator screen, which permits to confirm the correct positioning close to the diaphragm [Barwing 2009]. It has recently been confirmed that Edi catheter placement is not difficult and that the correct placement of the nasogastric tube is actually facilitated by the possibility of activity monitoring [Green 2011].
RIP jacket installation:
A special jacket adapted to the patient size will be installed to monitor the lung volume changes by impedance plethysmography (RIP). This sleeveless jacket is constituted of a large band of distensible cloth in which 2 metallic coils are coated to record the impedance changes created by the ventilation. The RIP coils will be connected to the RIP monitoring module, and no volume calibration will be done to avoid the manipulation of patient airway. RIP monitoring is completely non-invasive and no complications are expected [Emeriaud 2010][Emeriaud 2008].
Baseline measurements:
The Edi tube is connected to the Servo i to monitor EAdi, but the patient remains connected to its usual ventilator with unmodified ventilator settings. Airway pressure, respiratory volume and flow will be recorded simultaneously with EAdi during 30 minutes.
Ventilator change:
If a different ventilator was used prior to the study, the patient will be installed on a Servo i ventilator, initially with the same ventilatory settings.
Determination of NAVA parameters:
The positive end expiratory pressure (PEEP) will not be modified. The NAVA level - the proportionality factor that converts EAdi (microV) into pressure support (cmH2O) - will be set using a specific window which simulates the pressure that would be delivered in NAVA. The NAVA level will be initially adjusted to match the simulated pressure with the actual assist pressure. If the patient is on CPAP (i.e. no assist pressure), the NAVA level will be initially set to match an assist pressure of 5 cmH2O above PEEP. The maximal pressure alarm will be set at 30 cmH2O.
NAVA period:
The NAVA mode will be activated for one hour, under the continuous supervision by a physician involved in the study, and with a continuous monitoring of vital signs including cardiac and respiratory rate, SaO2, respiratory distress signs, patient agitation, and the normal function of the ventilator. The NAVA level will be adjusted if the patient breathing frequency is > 40/min (progressive increase of NAVA level) or < 12/min (decrease of NAVA level), or if the delivered pressure is low (<3 cmH2O) due to air leaks. Ventilatory pressure, flow and volumes and EAdi will be continuously recorded during the last 30 minutes.
Second period with conventional NIV settings:
The patient will be installed again with the ventilatory settings prescribed prior to the study, and the ventilatory data will be recorded again for 30 minutes.
During the entire study, "usual" modifications of settings (e.g. adaptation of assist level, FiO2, or PEEP) considered by the attending physician, the nurse, or a respiratory therapist, will be permitted and documented.
End of the study:
At the end of the study, if a clear clinical benefit of one mode or setting is observed during the study, the attending physician will be informed to permit the optimal adaptation of the ventilatory assist.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Nasogastric tiube installation to monitor EAdi
Other names: Health Canada approval number: 186536
Time frame: up to 28 days
The analysis of respiratory recordings will be conducted on 15 minute period during which the patient is not agitated, at the end of the 3 ventilation periods: baseline period with conventional parameters, NAVA period, and 2nd period with conventional parameters. The mean EAdi will be calculated as the area under the curve of EAdi versus time
Time frame: Up to 28 days
The inspiratory and expiratory times of the patient and of the ventilator will be identified cycle-by-cycle from the tracing of EAdi and airway pressure, respectively. The percentage of time spent in asynchrony (discordance between the patient and the ventilator actions) will be calculated [Beck 2004], as well as the delay to activate or inactivate the ventilatory support. The number of non-assisted breaths (increase in EAdi without pressure elevation) and of auto-triggered breaths (assist delivered in absence of EAdi activation) will also be counted [Thille 2006; Vignaux 2009].
Time frame: Up to 28 days
St. Justine's Hospital
Other
Acronym: NAVANI
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07376525
Mechanical Ventilation, Respiration Disorders
Istanbul, Sisli, Turkey (Türkiye)
View Trial DetailsNCT07411976
Critical Illness, Disease Attributes
Nîmes, France
View Trial DetailsNCT07658131
Critical Illness, Disease Attributes
Taichung, Taiwan
View Trial DetailsNCT07620912
Chronic Disease, Chronic Obstructive Pulmonary Disease
Asyut, Egypt
View Trial Details