NCT Number: NCT04313595

Device for Breathing Frequency Monitoring in Muscular Dystrophy

Patients at risk of developing respiratory dysfunctions, such as patients with severe forms of muscular dystrophy, need a careful respiratory assessment, and periodic follow-up visits to monitor the progression of the disease. Continuous monitoring of respiratory activity pattern at home could give additional understandings about disease progression, flanking traditional, intermittent, cardiopulmonary evaluations, allowing prompt clinical intervention, and anticipating respiratory dysfunction. The main objective of the present study is thus to investigate the feasibility of using an innovative wearable device for respiratory monitoring, especially breathing frequency variation assessment, in patients with muscular dystrophy. The comparison between the measurements of breathing frequency obtained by using the IMU-based device and by using the reference method provided optimal results, in terms of accuracy errors, correlation and agreement. Participants positively evaluated the device for what concerns ease of use, comfort, usability and wearability. Moreover, preliminary results confirmed that breathing frequency is an interesting breathing parameter to monitor, at the clinic and at home, because it strongly correlates with the main indexes of respiratory function

Completed

Looking for future studies?

Notify Me

Key information

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

IRCCS Medea

Bosisio Parini, Lecco, 23842, Italy

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • documented DMD or LGMD2
  • loss of independent ambulation (wheelchair-bound patients)
  • ability to understand and follow test instructions and to report pain and discomfort.

Exclusion criteria

  • presence of metal implants and cardiac pacemakers
  • relevant concomitant comorbidities (e.g.epilepsy)
  • behaviour and/or psychiatric disorders (e.g. emotional problems, anxiety, panic attacks).

Treatment and study plan

RespirHó

Device

The system is a wearable, unobtrusive inertial-sensor-based device for long-term breathing pattern monitoring, also during daily life activities. It consists of three IMU-sensor units (3-axis accelerometer, 3-axis gyroscope, 3-axis magnetometer), positioned respectively on the patient's abdomen and thorax, and on a body area integral with thorax but not affected by espiratory movements. The peripheral units, placed on thorax and abdomen are used to record orientation changes during respiratory movements. The third unit is a central reference unit (hereafter CRU) that receives data from the other two units, save them on an SD card and communicate via Bluetooth with a smartphone / tablet / PC. Moreover, this unit detects only non-respiratory movement, representing not only a pure source of "noise" that must be removed from the thoracic and bdominal signals, but also a pure source of additional information regarding the state of ctivity of the subject .

Primary outcomes

  1. Safety of the device

    Time frame: After 24 hours of continued use

    Recording of any adverse events

  2. Usability of the device (System Usability Scale)

    Time frame: After 24 hours of continued use

    System Usability Scale (SUS acceptance and wearability of the device):the scores range from 0 to 100 in 2.5-point increments, with higher values meaning high perceived usability of the system

  3. Acceptability of the device (Technological Acceptance Measure)

    Time frame: After 24 hours of continued use

    Technological Acceptance Measure (TAM) questionnaire appropriate for the medical device: with higher values meaning high acceptance of the system

Secondary outcomes

  1. Accuracy in measuring the breathing frequency

    Time frame: Immediately after application of the device

    Comparison between the measures of respiratory rate by using the device and by using already validated instrumentation: Optoelectronic Plethysmography (OEP), optoelectronic movement analysis system.

    Chest wall movements during breathing are simultaneously recorded by the RespirHó device and by Optoelectronic Plethysmography (OEP), in static conditions, and in particular, in supine and seated positions.

    OEP allows assessment of ventilatory and breathing pattern by measuring chest-wall movements related to breathing, by using motion capture principles.

    Using OEP as reference method it is possible to compare the data recorded with the thoracic and abdominal units of the device RespirHó with those obtained by using OEP for the thoracic and abdominal compartments

Sponsors and collaborators

Lead sponsor

IRCCS Eugenio Medea

Other

Registry information

Official study title

A Wearable Device for Breathing Frequency Monitoring: a Pilot Study on Patients With Muscular Dystrophy

Important dates

Study start
2018
Primary completion
2019
Study completion
2019
First posted
Mar 18, 2020
Registry last updated
Mar 18, 2020

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.