NCT Number: NCT03796624
Prospective, non-randomised, open, controlled, single-center post-market clinical follow study about Micropure 1.2.3. and PODEYE intraocular lenses.
Looking for future studies?
Notify Me45 year and older
All sexes
Interventional
Not applicable
Zlín, 76001, Czechia
This clinical investigation is a prospective, non-randomised, open, controlled, single-center post-market clinical follow whereby study patients undergoing routine cataract surgery will have implantation of monofocal intraocular lens Micropure 1.2.3. (PhysIOL, Liège, Belgium) in one eye and implantation of monofocal intraocular lens PODEYE (PhysIOL, Liège, Belgium) in the contralateral eye of the same patient.
The study purpose is to obtain clinical data on visual acuity and contrast sensitivity on patients implanted with Micropure 1.2.3. and PODEYE
The devices under investigation (Micropure 1.2.3. and PODEYE) are a monofocal glistening-free hydrophobic acrylic intraocular lenses (IOLs) manufactured by the sponsor of this study PhysIOL sa/nv. The optical properties of the lenses are very comparable. The main difference is the mechanical design of the haptics, that is not expected to have an influence on the clinical outcomes.
The IOLs will be implanted as part of the routine cataract surgery on patients suffering from cataract development.
In total 76 patients will be recruited for this clinical study and receive the implantation of Micropure 1.2.3. and POPDEYE intraocular lenses.
Subjects participating in the trial will attend study visits over a period of 12 months.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Implantation of intraocular lens (IOL). Name: "MicroPure 1.2.3." It is a monofocal intraocular lens consisting of hydrophobic acrylic material. One IOL per patient will be implanted
Implantation of intraocular lens (IOL). Name: "PODEYE" It is a monofocal intraocular lens consisting of hydrophobic acrylic material. The IOL will be implanted in the contralateral eye of the same patient already implanted with Micropure 1.2.3.
Time frame: 6 months postoperative
Statistically non-inferior visual acuity outcomes on monocular Corrected Distance Visual Acuity (CDVA) under photopic light conditions compared to eyes implanted with the comparator PODEYE at the 6 months follow up visit.
Time frame: preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
The manifested refraction is measured by means of a phoropter. The data contains values for sphere, cylinder and axis of cylinder according to ISO 11979-7:2018.
Time frame: preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
UDVA is measured with ETDRS charts placed in 4m distance according to ISO 11979-7:2018. This assessment is performed under photopic light conditions
Time frame: preoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
UDVA is measured with ETDRS charts placed in 4m distance according to ISO 11979-7:2018. This assessment is performed under mesopic light conditions
Time frame: preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
CDVA is measured with ETDRS charts placed in 4m distance with best aided corrective glasses according to ISO 11979-7:2018. This assessment is performed under photopic light conditions
Time frame: preoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
CDVA is measured with ETDRS charts placed in 4m distance with best aided corrective glasses according to ISO 11979-7:2018. This assessment is performed under mesopic light conditions
Time frame: 6 months postoperative
Contrast Sensitivity under photopic light conditions using the standardized contrast sensitivity device CSV-1000 (VectorVision)
Time frame: 6 months postoperative
Contrast Sensitivity under mesopic light conditions using the standardized contrast sensitivity device CSV-1000 (VectorVision)
Time frame: preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
Time frame: preoperative, 12 months postoperative
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
Time frame: preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
Time frame: preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
Time frame: preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
Time frame: preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
Time frame: preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
Time frame: preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
Time frame: preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
Time frame: preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
Time frame: preoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative
The IOP will be measured with non-contact tonometer as part of the routine follow up examinations
Time frame: peroperative
A questionnaire will be handed out to the surgeon right after the surgery to document the ease of use and possible issues during IOL implantation. These data will be used to compare the outcomes to different surgery techniques or injectors. The questionnaire is not validated and the outcomes serve only for the sponsor to receive feedback if one or the other IOL is easier to implant.
Time frame: preoperative
Keratometric measurements are performed to calculate the required IOL power
Time frame: preoperative
Biometry measurements are performed to calculate the required IOL power
Beaver-Visitec International, Inc.
Industry
Comparison of Clinical Outcomes After Implantation of MICROPURE (PhysIOL) and PodEye (PhysIOL) Monofocal Intraocular Lenses Into the Two Eyes of the Same Patient
Acronym: PHY1803
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06399211
Aphakia, Aphakia, Postcataract
View Trial DetailsNCT04249492
Cataract, Eye Diseases
Vienna, Austria
View Trial DetailsNCT05875922
Cataract, Eye Diseases
Visp, Switzerland
View Trial DetailsNCT04465344
Cataract, Eye Diseases
Ostrava, Czechia
View Trial Details