This is a prospective randomised open-label study with the aim to compare the visual performance of the premium monofocal Isopure IOL versus the extended depth of focus Eyhance® IOL after phaco-emulsification cataract surgery.
Looking for future studies?
Notify Me50 year and older
All sexes
Interventional
Not applicable
Visp, 3930, Switzerland
This is a single center, prospective, randomised, controlled, double-blind post-market clinical follow-up (PMCF) study whereby patients undergoing routine cataract surgery will have bilateral implantation of premium monofocal intraocular lenses (PhysIOL ISOPURE 1.2.3.) or EDOF lenses (TECNIS Eyhance). The patients will be randomized in a 1:1 ratio to receive the study or control lenses. Both IOLs, investigational device and control device, are CE approved and commercially available in the countries this clinical investigation is being carried out. The investigational device and all study products, including the devices used for the study examinations, will be used within the intended use specifications from the manufacturer. In addition, no invasive or other burdening examinations will occur for the patient.
The device under investigation (ISOPURE 1.2.3.) is a hydrophobic, glistening-free, acrylic premium monofocal intraocular lens (IOL) manufactured by the sponsor of this study, PhysIOL sa/nv. The control lens (TECNIS Eyhance) is a non-diffractive Extended Vision Posterior Chamber IOL.
The IOLs will be implanted as part of the routine cataract surgery on patients suffering from cataract development.
Subjects participating in the trial will attend a total of 6 study visits (1 preoperative, 1 operative and 4 postoperative) over a period of 6 months. Subjects would have the option for unscheduled visits if required medically.
Primary and secondary endpoint data will be collected at the 120-180 days follow up visit.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
In addition to above mentioned in- and exclusion criteria, subjects shall be discontinued when certain conditions are present at the time of surgery, including:
Implantation of premium monofocal IOL ISOPURE 1.2.3.
Implantation of EDOF IOL TECNIS Eyhance
Time frame: 6 months (120-180 days) postoperative
Non inferiority of ISOPURE 1.2.3. outcomes compared to the TECNIS Eyehance outcomes in terms of monocular Best Corrected Distance Visual Acuity (CDVA) by means of statistical significance.
A significance level of 0.05 or lower (p < 0.05) will be considered statistically significant. CDVA is measured with ETDRS charts placed in 4m distance with best aided corrective glasses according to ISO 11979-7:2018.
Time frame: 1 day (1-2 days) postoperative, 1 week (7-14 days) postoperative, 1 month (30-60 days) postoperative, 6 months (120-180 days) postoperative
The objective is to compare the SPE (Safety and Performance Endpoints) rates of the investigational product to the results of historical data of EN ISO 11979-7:2018, Annex E (Table E.2)
Time frame: Preoperative, 1 day (1-2 days) postoperative, 1 week (7-14 days) postoperative, 1 month (30-60 days) postoperative, 6 months (120-180 days) postoperative
The IOP will be measured with non-contact tonometer as part of the routine follow up examinations. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
Time frame: Preoperative
Anterior Chamber Depth measurement is performed using the IOLMaster device. This measurement is required to calculate the required IOL power. The measurement unit is mm. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
Time frame: Preoperative
Axial Length measurement is performed using the IOLMaster device. This measurement is required to calculate the required IOL power. The measurement unit is mm. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
Time frame: Preoperative
Lens Thickness measurement is performed using the IOLMaster device to asses the dimension of the natural lnes. This measurement is required to calculate the required IOL power. The measurement unit is mm. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
Time frame: Preoperative
Keratometric measurements are performed to calculate the required IOL power. The measurement is performed using the IOLMaster device. This specific outcome measure is to measure the K value in the steep axis (K1) and flat axis (K2). The measurement unit is diopter (D). The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
Time frame: Operative (day of surgery)
The IOL spherical power of the implanted IOL must be recorded.
Time frame: Operative (day of surgery)
The target refraction given by the IOL calculator must be recorded. This parameter is needed to calculate the accuracy of achieving the target refraction. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
Time frame: Preoperative, 1 month (30-60 days) postoperative
An OCT (optical coherence tomography) image will be taken at the preoperative visit and 1 months postoperative visit to identify AMD suspicious eyes that need to be excluded from this study. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
Time frame: preoperative
Mesopic pupil diameters are measured with a precision of at least +/-0.5 mm. For the mesopic pupil diameters, eye illumination should be identical to that used for mesopic contrast sensitivity testing. It is required that pupil measurements be made with an infrared pupillometer / CCD camera. Pupil measurements should only be made after the eye has had time to fully adapt to the testing conditions (at least 10 minutes).
The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
Time frame: Preoperative, 1 week (7-14 days) postoperative, 1 month (30-60 days) postoperative, 6 months (120-180 days) postoperative
The manifested refraction is measured by means of a phoropter. The data contains values for sphere, cylinder and axis of cylinder according to ISO 11979-7:2018. This data will also be used to calculate the manifested refractive spherical equivalent (MRSE)
Time frame: Preoperative, 1 day (1-2 days) postoperative, 1 week (7-14 days) postoperative, 1 month (30-60 days) postoperative, 6 months (120-180 days) postoperative
monocular and binocular UDVA is measured with ETDRS charts placed in 4m distance according to ISO 11979-7:2018. This assessment is done monocularly under photopic light conditions.
Time frame: 1 month (30-60 days) postoperative, 6 months (120-180 days) postoperative
Binocular UDVA is measured with ETDRS charts placed in 4m distance according to ISO 11979-7:2018. This assessment is done binocularly under photopic light conditions.
Time frame: 1 month (30-60 days) postoperative, 6 months (120-180 days) postoperative
Binocular UDVA is measured with ETDRS charts placed in 4m distance according to ISO 11979-7:2018. This assessment is done binocularly under mesopic light conditions.
Time frame: Preoperative, 1 week (7-14 days) postoperative, 1 month (30-60 days) postoperative, 6 months (120-180 days) postoperative
Monocular CDVA is measured with ETDRS charts placed in 4m distance with best aided corrective glasses according to ISO 11979-7:2018. This assessment is done monocularly under photopic light conditions.
Time frame: 6 months (120-180 days) postoperative
Binocular CDVA is measured with ETDRS charts placed in 4m distance with best aided corrective glasses according to ISO 11979-7:2018. This assessment is done binocularly under photopic light conditions.
Time frame: 1 week (7-14 days) postoperative, 1 month (30-60 days) postoperative, 6 months (120-180 days) postoperative
Monocular DCIVA@66cm is measured with ETDRS charts placed in 66cm distance with corrective glasses for far distance according to ISO 11979-7:2018. This assessment is done monocularly under photopic light conditions.
Time frame: 1 month (30-60 days) postoperative, 6 months (120-180 days) postoperative
Binocular UIVA@66cm is measured with ETDRS charts placed in 66cm distance. The examination is done without corrective glasses and according to ISO 11979-7:2018. This assessment is done binocularly under mesopic light conditions.
Time frame: 1 month (30-60 days) postoperative, 6 months (120-180 days) postoperative
Monocular DCNVA is measured with ETDRS charts placed in 40cm distance with corrective glasses for far distance according to ISO 11979-7:2018. This assessment is done monocularly under photopic light conditions.
Time frame: 1 week (7-14 days) postoperative, 1 month (30-60 days) postoperative, 6 months (120-180 days) postoperative
Monocular UNVA is measured with ETDRS charts placed in 40cm distance. The examination is done without corrective glasses and according to ISO 11979-7:2018. This assessment is done binocularly under photopic light conditions.
Time frame: 1 month (30-60 days) postoperative
To assess the visual acuity for different distances, defocus curves under photopic light conditions are measured. This test is performed with best distance corrected refraction and spherical additions ranging from -2.5 D to +1.5 D in 0.5 D steps. This examination is performed binocularly.
Time frame: 6 months (120-180 days) postoperative
Contrast Sensitivity under photopic light conditions using the standardized contrast sensitivity device CSV-1000 (Vector Vision). This examination is performed binocularly.
Time frame: 6 months (120-180 days) postoperative
Contrast Sensitivity under mesopic light conditions using the standardized contrast sensitivity device CSV-1000 (Vector Vision). This examination is performed binocularly.
Time frame: Preoperative, 1 month (30-60 days) postoperative, 6 months (120-180 days) postoperative
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
Time frame: Preoperative, 1 month (30-60 days) postoperative, 6 months (120-180 days) postoperative
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
Time frame: Preoperative, 1 month (30-60 days) postoperative, 6 months (120-180 days) postoperative
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
Time frame: Preoperative, 1 month (30-60 days) postoperative, 6 months (120-180 days) postoperative
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
Time frame: Preoperative, 1 week (7-14 days) postoperative, 1 month (30-60 days) postoperative, 6 months (120-180 days) postoperative, 12 months (330-420 days) postoperative
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
Time frame: 1 month (30-60 days) postoperative, 6 months (120-180 days) postoperative
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
Time frame: 6 months (120-180 days) postoperative
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
Time frame: 1 month (30-60 days) postoperative, 6 months (120-180 days) postoperative
The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
Time frame: 6 months (120-180 days) postoperative
To assess the subjective perception of disturbances by photic phenomena. Subjects answer one question per category on their individual perception of glare, halo and starbursts, respectively. Answers are graded in 5 steps between always and never.
Descriptive analysis will be performed on the occurence of photic phenomena per category. This questionnaire does not consist of a scaling.
Time frame: 6 months (120-180 days) postoperative
The Patient-Reported Spectacle Independence Questionnaire (PRSIQ) is a patient-reported measure assessing spectacle independence following cataract surgery. The analyses conducted provide evidence for the use of the PRSIQ total score as a measure of spectacle independence [Morlock et al, 2017].
In the first category, subjects are asked if they need glasses for distance vision, intermediate vision and near vision with 'yes' and 'no' being the only possible answers.
In the second category, patients are asked to scale their frequency of using spectacles over the last 7 days for three different distances with 5 ratings ranging from '1: All of the time' to '5: None of the time'. A higher rating means higher rate of spectacle independence.
In addition, patients are asked to scale if they are able to function without spectacle aid for three different distances with 5 ratings ranging from '1: All of the time' to '5: None of the time'. A higher rating means higher rate of spectacle freedom.
Beaver-Visitec International, Inc.
Industry
Prospective Post-Market Clinical Follow-Up (PMCF) Study to Investigate the Clinical Outcomes of a ISOPURE and EYHANCE Intraocular Lenses (IOL)
Acronym: PHY2301
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06399211
Aphakia, Aphakia, Postcataract
View Trial DetailsNCT04249492
Cataract, Eye Diseases
Vienna, Austria
View Trial DetailsNCT04465344
Cataract, Eye Diseases
Ostrava, Czechia
View Trial DetailsNCT04176965
Aphakia, Aphakia, Postcataract
Northridge, California, United States
View Trial Details