Time frame: The evaluation will be compared from pre-MRI to 1-month post-MRI scan
There was no hypothesis for this secondary objective. Pacing capture thresholds (PCTs) to be characterized include:
APCT of atrial lead for subjects with an IPG device; APCT of atrial lead and VPCT of right ventricular (RV) lead for subjects with an ICD device; APCT of atrial lead and VPCTs of right ventricular (RV) and left ventricular (LV) leads for those with a CRT-D device.
PCTs of each subject will be measured prior to the 3.0T MRI scan and at the one-month post-MRI. Change in PCT is defined as PCT measured at one-month post-MRI subtracts that measured prior to a 3.0T MRI scan. Proportion of subjects with an increase less than or equal to 0.5V in APCT or VPCTs from pre-MRI to one-month post-MRI will also be used to characterize the efficacy performance of SureScan with respect to PCTs. If the 1-month post-MRI APCT or VPCT is missing, the PCT measured immediately post-MRI will be used instead.