Bioscience of Bozeman, LLC
Bozeman, Montana, 59715, United States
NCT Number: NCT07188129
The purpose of this study is to demonstrate that the ZywieAI analysis algorithm can be used to analyze ECG data collected from ZywieZ3 devices with non-traditional lead placement over a 24-hour simulated use period by human subjects.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Bozeman, Montana, 59715, United States
At least 150 subjects will participate in this study over a 24-hour test period. The subjects must have been diagnosed with a known condition such as Atrial Fibrillation (AF), Tachycardia, Bradycardia, Premature Ventricular Contractions (PVC/VEB), and/or Premature Atrial Contractions (PAC), etc.
A Testing Facility analyst will conduct an initial visual evaluation of the level of skin adhesion for the Test Device. The Testing Facility analyst will apply the Test device adhesive on the subject and initiate ECG data collection. Subjects will observe a 5-minute stabilization period to ensure the device is connected appropriately and has a stable input signal. Subjects will be allowed to leave the Testing Facility for the next 24 hours. During this period, subjects will follow the instructions for use provided by the Sponsor and engage in normal routine activities. After the 24-hour recording, subjects will return to the Testing Facility, and an analyst will then remove the Test Device.
Two cardiologists will review the collected ECG data from the 24-hour wear time to identify episodes of arrhythmia. A third cardiologist will be used to review the ECG data collected if the first two do not agree with their assessment. The Sponsor will download the ECG data for the Test Device from a cloud-based data repository to complete the data analysis according to the Sponsor's statistical analysis plan. The cardiologist's assessment will be compared with the analysis by the ZywieAI algorithm. A summary of the ECG statistical analysis comparing the agreement of the cardiologists will be included as an appendix to the final report for the study.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Test Algorithm analysis of 24-hour ECG Data collected with a Test Device
Time frame: 24 Hours after Application of Test Device
Two cardiologists will review the collected ECG data from the 24-hour wear time to identify episodes of arrhythmia. A third cardiologist will be used to review the ECG data collected if the first two do not agree with their assessment. The Sponsor will download the ECG data for the Test Device from a cloud-based data repository to complete the data analysis by ZywieAI algorithm identifying cardiac anomalies. The cardiologist's assessment will be compared with the analysis by the ZywieAI.
Zywie, Inc.
Industry
Assessing Compatibility of the ZywieAI Algorithm to Analyze ECG Data Collected From the Zywiez3 Sensor and Adhesive Over a 24-hour Period
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07018648
Arrhythmias, Cardiac, Atrial Fibrillation
Olomouc, Czechia
View Trial DetailsNCT06892938
Arrhythmias Cardiac, Arrhythmias, Cardiac
La Jolla, California, United States
View Trial DetailsNCT07220525
Apnea, Arrhythmias, Cardiac
San Francisco, California, United States
View Trial DetailsNCT07217392
Arrhythmias, Cardiac, Atrial Fibrillation
Rio de Janeiro, Brazil
View Trial Details