Perin Health Devices
Woodland Hills, California, 91364, United States
NCT Number: NCT07147920
The goal of this clinical trial is to test how accurately and reliably a new chest-worn device called the Perin Health Patch measures heart rhythms (ECG) in adults aged 22 and older. The main questions it aims to answer are:
* How accurately does the Perin Health Patch measure heart rhythm compared to standard hospital ECG monitoring equipment? * Does the patch still work properly if it's placed slightly off from the exact recommended position on the chest? * Does the patch maintain good performance when worn continuously for 15 days?
The study has three parts that test different aspects of the device. Participants will:
* Wear the Perin Health Patch on their chest along with standard ECG monitoring equipment for comparison * Have ECG recordings taken while sitting quietly for 15 minutes * For the placement study: have the patch placed in several different positions to test if small placement errors affect performance * For the long-term study: wear the patch continuously for 15 days at home with regular check-ups on days 1, 7, 10, and 15
The study will include both healthy adults and people with heart or lung conditions like heart failure, COPD, high blood pressure, or diabetes to test how well the device works for different types of patients. Researchers will compare the heart rhythm measurements from the chest patch to standard medical monitoring equipment to determine if the patch is accurate enough for medical use. Some participants will complete just one part of the study, while others may participate in multiple parts depending on which aspect of the device is being tested.
Looking for future studies?
Notify Me22 year–90 year
All sexes
Observational
Woodland Hills, California, 91364, United States
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
The Perin Health Patch is a non-invasive, chest-worn wearable device designed to continuously monitor electrocardiogram (ECG) signals using dry conductive adhesive electrodes. In this validation study, the device will be evaluated for its ability to accurately measure ECG parameters including heart rate, QT interval, QRS duration, and R-R intervals compared to FDA-cleared Holter monitors.
The study intervention involves three components: (1) simultaneous ECG recording with the patch and reference Holter monitor for 15 minutes while participants remain seated, (2) placement tolerance testing where the patch is positioned at various locations (±20mm from optimal placement) to assess performance under suboptimal positioning, and (3) long-term wear evaluation where participants wear the patch continuously for 15 days with periodic assessments. The intervention is limited to device application and ECG data collection under controlled conditions, with no therapeutic treatment provided.
Time frame: During single 15-minute recording session and up to 15 days of continuous monitoring
Equivalence of ECG measurements between the investigational chest patch and FDA-cleared Holter monitor using Two One-Sided Tests (TOST) statistical methodology. ECG recordings are taken simultaneously from both devices while participants sit quietly for comparison. Equivalence is demonstrated when confidence intervals for mean differences fall within pre-specified clinical margins for key ECG timing intervals including QT interval, QRS duration, and R-R interval. Signal morphology similarity is assessed using Pearson correlation coefficient (where 1.0 indicates perfect similarity) and normalized root mean square error (where 0% indicates identical signals).
Time frame: During single 15-minute recording session and up to 15 days of continuous monitoring
Accuracy of heart rate measurements from the investigational device compared to the reference Holter monitor. Heart rate is calculated from R-R intervals during simultaneous ECG recordings. Accuracy is assessed using mean absolute error between devices, where lower values indicate better agreement.
Time frame: During single 15-minute recording session and up to 15 days of continuous monitoring
Comparison of signal quality between the investigational device and reference monitor using signal-to-noise ratio analysis. Signal quality is measured in decibels (dB), where higher values indicate cleaner signals with less noise interference. Equivalence is assessed using statistical comparison methods to determine if signal quality differences fall within clinically acceptable ranges.
Time frame: Over 15-day continuous monitoring period
Percentage of participants who successfully complete the full monitoring period without device adhesion failure. Adhesion failure is defined as device detachment that prevents reliable ECG signal acquisition. Success is measured by the proportion of participants maintaining adequate device adherence throughout the intended wear duration.
Perin Health Devices
Industry
Clinical Validation of the Perin Health Patch Electrocardiogram Monitoring Device: A Multi-Component Study Evaluating Signal Quality, Placement Tolerance, and Long-Term Performance
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07533838
ECG, Nursing Skills
Bartın, Turkey (Türkiye)
View Trial DetailsNCT07057440
ECG, Heart Rate
Hvidovre, Denmark
View Trial DetailsNCT06335498
Afib, Arrhythmias, Cardiac
Gig Harbor, Washington, United States
View Trial DetailsNCT06076798
Afib, Arrhythmias, Cardiac
San Diego, California, United States
View Trial Details