NCT Number: NCT04433520

Trevisio Post-Approval Study

A single-arm, non-randomized, multi-center clinical study of the Amplatzer™ Trevisio™ Intravascular Delivery System for facilitating percutaneous, transcatheter implantation of the Amplatzer™ Occluder Devices.

Completed

Looking for future studies?

Notify Me

Key information

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Patient was indicated for implantation with the Amplatzer Septal Occluder for occlusion of a secundum atrial septal defect (Note: This does not include the indication for closure of a fenestration following a fenestrated Fontan procedure) OR subject was indicated for implantation with the Amplatzer PFO Occluder
  • Patient was of legal age and has provided his/her own written, informed consent

OR

Patient was a minor and has provided verbal and/or written informed consent or assent per local EC requirements, and his/her legally authorized representative, or representatives, had provided written informed consent on behalf of the minor according to local EC requirements

  • Patient was indicated for implantation with the Amplatzer Muscular VSD Occluder OR patient was indicated for implantation with the Amplatzer Post-Infarct Muscular VSD Occluder

Exclusion criteria

  • Presence of anatomic or comorbid conditions, or other medical, social, or psychological conditions that, in the Principal Investigator's opinion, could limit the subject's ability to participate in the clinical study or to comply with follow-up requirements.
  • Exclusion Criteria for Patients Undergoing ASD Closure with the Amplatzer ASO or Amplatzer ASD-MF Occluder
  • Patients known to have extensive congenital cardiac anomaly that can only be adequately repaired by cardiac surgery
  • Patients known to have sepsis within 1 month prior to implantation, or any systemic infection that cannot be successfully treated prior to device placement
  • Patients known to have demonstrated intracardiac thrombus on echocardiography (especially left atrial or left atrial appendage thrombi)
  • Patients whose size or condition (e.g., too small for transesophageal echocardiography [TEE] probe, catheter size, vasculature size, active infection) would cause the patient to be a poor candidate for cardiac catheterization
  • Patients with defect margins less than 5 mm to the coronary sinus, inferior vena cava rim, an atrioventricular valve, or the right upper lobe pulmonary vein
  • Exclusion Criteria for Patients Undergoing PFO Closure
  • Presence of thrombus at the intended site of implant, or documented evidence of venous thrombus in the vessels through which access to the defect is gained. Thrombus must be ruled out prior to introducing the delivery system.
  • Active endocarditis or other infections producing bacteremia
  • Patients whose vasculature, through which access to the defect is gained, is inadequate to accommodate the appropriate sheath size
  • Anatomy in which the Amplatzer™ PFO device size required would interfere with other intra-cardiac or intravascular structures, such as valves or pulmonary veins
  • Patients with known hypercoagulable states
  • Patients with intra-cardiac mass or vegetation, thrombus, or tumor
  • Body weight <8 kg
  • Tetralogy of Fallot
  • Intracardiac thrombi on echocardiography

Treatment and study plan

Amplatzer™ Trevisio™ Intravascular Delivery System using Amplatzer PFO Occluder, ASO or Amplatzer Cribriform Occluder

Device

Amplatzer Trevisio Delivery System is used for facilitating percutaneous, transcatheter implantation of the Amplatzer PFO Occluder, ASO or Amplatzer Cribriform Occluder.

Amplatzer™ Trevisio™ Intravascular Delivery System using Amplatzer Muscular VSD Occluder or Amplatzer Post-Infarct Muscular VSD Occluder

Device

Amplatzer Trevisio Delivery System is used for facilitating percutaneous, transcatheter implantation of the Amplatzer Muscular VSD Occluder or Amplatzer Post-Infarct Muscular VSD Occluder.

Primary outcomes

  1. Effectiveness Endpoint: Rate of Technical Success

    Time frame: at pre-hospital discharge or 7 days after the procedure

    The Technical success, defined as successful deployment and release of at least one device

  2. Safety Endpoint : Device- or Procedure-related Serious Adverse Events.

    Time frame: through discharge or 7 days after the procedure

    The primary safety endpoint is device- or procedure-related serious adverse events through discharge or 7 days, whichever is earlier, including:

    • Cardiac perforation
    • Sustained atrial fibrillation requiring intervention
    • Device thrombus
    • Device erosion
    • Device embolization
    • Vascular complication requiring surgical intervention
    • Device- or procedure-related serious adverse event leading to death

Sponsors and collaborators

Lead sponsor

Abbott Medical Devices

Industry

Registry information

Official study title

Amplatzer™Trevisio™Delivery System Post-Approval Study

Acronym: TrevisioPAS

Important dates

Study start
2020
Primary completion
2023
Study completion
2023
First posted
Jun 16, 2020
Registry last updated