NCT Number: NCT02326493

Tailor-CRT: Better Application of Cardiac Resynchronization Therapy

Approximately one third of patients treated with cardiac resynchronization therapy (CRT) do not derive any clinical benefit. CRT response can be improved by tailoring LV lead placement and programming of atrio-ventricular (AV) and inter-ventricular (VV) stimulation intervals to the individual patient. However, the best strategy to optimize lead positioning and device programming still remains to be established. Earlier work in our research group suggests that the vector cardiogram (VCG) can be used to determine the optimal LV lead position and AV- and VV-intervals, and pilot studies showed the feasibility to derive a VCG-like signal (D-VCG) from the implanted pacing electrodes. Other studies have suggested that the best position for the LV electrode is the region of latest electrical activation. The region of latest electrical activation can be identified by measuring the electrical delay on the LV lead (LVLED) during implantation. The objective of this study is to investigate whether D-VCG can be used to determine the optimal AV- and VV-interval and whether VCG and LVLED can be used to determine the optimal LV lead position.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year–80 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Maastricht University Medical Centre

Maastricht, Limburg, 6202 AZ, Netherlands

About this study

Cardiac resynchronization therapy (CRT) is an established treatment for heart failure (HF) patients with severe left ventricular (LV) systolic impairment and delayed electrical impulse conduction through the ventricles, such as left bundle-branch block (LBBB). Since initial approval of the therapy over 10 years ago, there have been hundreds of thousands of implants worldwide. In The Netherlands, currently more than 2000 CRT devices are implanted each year. In a heart with LBBB, electrical activation of the lateral LV free wall is delayed, which leads to dyssynchronous and inefficient LV mechanical contraction and compromised LV pump function. The positive impact of CRT on LV pump function is attributed to paced pre-excitation of the delayed activated lateral LV wall. CRT is most commonly applied by pacing the right ventricle (RV) and LV lateral wall (almost) simultaneously. This corrects the abnormal LV electrical activation and resynchronizes LV mechanical contraction, which in turn results in improved LV pump function.

Despite the striking effectiveness of CRT, 30-50% of apparently suitable patients show little or no improvement. Previous studies have shown that the response to CRT can be improved by tailoring LV lead placement and programming of atrioventricular (AV) and inter-ventricular (VV) stimulation intervals to the individual patient. In clinical practice, echocardiographic techniques are the most widely employed for CRT optimization. However these techniques are subject to large measurement errors and inter- and intra-observer variability. A more accurate technique is invasive assessment of acute hemodynamic response to CRT, with the most widely used invasive hemodynamic parameter being the maximum rate of LV systolic pressure rise (LVdP/dtmax). However, the invasive and time-consuming nature of this approach limits its use in clinical practice. Thus, the best strategy to optimize lead positioning and device programming still remains to be established.

Earlier work in our research group suggests that the vectorcardiogram (VCG) can be used to determine the optimal LV lead position and AV- and VV-intervals, and pilot studies showed the feasibility to derive a VCG-like signal (D-VCG) from the implanted pacing electrodes. Other studies have suggested that the best position for the LV electrode is the region of latest electrical activation. The region of latest electrical activation can be identified by measuring the electrical delay on the LV lead (LVLED) during implantation. Validation of these techniques for tailoring LV lead positioning and AV- and VV- stimulation intervals to the individual patient, will provide non-invasive and easy methods to optimize CRT application and improve response rate.

The objective of this study is to investigate whether D-VCG can be used to determine the optimal AV- and VV-interval and whether VCG and LVLED can be used to determine the optimal LV lead position. Validation of these techniques for tailoring LV lead positioning and AV- and VV- stimulation intervals to the individual patient, will provide non-invasive and easy methods to optimize CRT application and improve response rate.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Chronic heart failure with NYHA functional class II-IV
  • Left ventricular ejection fraction (LVEF) < 35%
  • Left bundle-branch block (LBBB) with QRS duration > 120 ms
  • In sinus rhythm

Exclusion criteria

  • Atrial fibrillation
  • ≥4 premature ventricular complexes on standard 12-lead ECG
  • Age <18 years or > 80 years
  • Incapable of giving informed consent
  • Moderate to severe aortic valve stenosis

Treatment and study plan

cardiac resynchronization therapy

Device

A CRT device will be implanted while performing extra hemodynamic (LV dP/dtmax) and electrical (LVLED, VCG, and D-VCG) measurements. Devices and leads from various vendors will be used.

Primary outcomes

  1. Correlation between the increase in LV dP/dtmax and the D-VCG derived QRS area, obtained at different AV- and VV-intervals.

    Time frame: Acute measurements are performed for the duration of the CRT implantation procedure, an expected average of three hours

    The optimal AV- and VV-interval produces the maximal increase in LV dP/dtmax. It is investigated whether the maximal increase in LV dP/dtmax also corresponds to the minimal QRS area derived from the D-VCG. The correlations will be expressed by the Pearson Correlation coefficient.

  2. Correlation between the increase in LV dP/dtmax and the LVLED or VCG derived QRS area, obtained at different potential LV lead positions

    Time frame: Acute measurements are performed for the duration of the CRT implantation procedure, an expected average of three hours

    The optimal LV lead position produces the maximal increase in LV dP/dtmax. It is investigated whether the maximal increase in LV dP/dtmax also corresponds to the longest LVLED or the minimal QRS area derived from the VCG. The correlations will be expressed by the Pearson Correlation coefficient.

Secondary outcomes

  1. Correlations between QRS vector area, -angle and -amplitude derived from VCG and from D-VCG.

    Time frame: Acute measurements are performed for the duration of the CRT implantation procedure, an expected average of three hours

    The correlations will be expressed by the Pearson Correlation coefficient.

Sponsors and collaborators

Lead sponsor

Maastricht University Medical Center

Other

Collaborators

  • Medtronic
  • University Medical Center Groningen

Registry information

Official study title

Tailor-CRT: Better Application of Cardiac Resynchronization Therapy by Automated and Improved Selection of Location and Timing of Stimulation

Important dates

Study start
2014
Primary completion
2016
Study completion
2016
First posted
Dec 29, 2014
Registry last updated
Feb 23, 2017

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.