ARGOS-IO system
DeviceThis study will enroll a minimum of 10 and a maximum of 15 patients. It is anticipated to be large enough to provide an initial estimate of common safety events and assessment of performance. Any adverse event (AE), serious adverse event (SAE), ADE (adverse device event) and SADE (serious adverse device event) will be listed. Incidence will be estimated with a 95% confidence interval (Pearson-Clopper, two-sided). The Bland-Altman method will be used to assess the limits of agreement between the IOP measurements ARGOS-IO and surgical manometry. When appropriate, two-sided 95% confidence intervals, for these limits will be calculated accounting for repeated measurements based on the method proposed by Zou (2011). Other secondary performance endpoints will be analyzed by descriptive and explorative statistical methods.
Other names: ARGOS-IO pressure sensor implant