Skip to main content
OpenTrials
Completed

NCT Number: NCT06147778

Performances and Safety of MuCopilot, a Digital Tool for the Unsupervised Objective Assessment of Cystic Fibrosis

The study aims to evaluate MuCopilot, a smartphone application to measure objective data on lung function (cough and dyspnea tests), global exercise capacity (walking test) and patient-reported outcomes (PROs) of patients with Cystic Fibrosis (CF). These data are collected during unsupervised digital tests performed in the patient's home environment between consultations.

The primary objective is to demonstrate the correlation of the MuCopilot digital tests at home (D1) with the results of their standard counterparts (D0) carried out in-clinic.

A secondary objective is to determine the accuracy, reliability and reproducibility of tests results, as well as to study the test-retest of the PRO. The study also aims to assess the safety, usability, and satisfaction of the solution.

Exploratory objectives include evaluating the relationship between MuCopilot's scores and other standards such as FVC, FEV1/FVC ratio as well as to explore the correlation with the cough monitoring with patient's perception.

Patients will be able to download MuCopilot app. They will participate in 1 inclusion visit and 7 follow-up visits, scheduled at Day 1, Day 3, Day 5, Day 7, 1 month, 2 months, 3 months - 1 day.

The study will include 70 CF patients and will be conducted in France.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Centre Hospitalier Lyon Sud (Adultes), Lyon, France

Loading trial locations.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Adult (≥ 18 years old) patients living with Cystic Fibrosis (CF)
  • Enrolled in or benefiting of a Social Security program
  • Having read the information sheet and signed the informed consent form
  • Owning a personal smartphone which Operating System (OS) is equal or above 14 for iOS (iPhone) and 8 for Android, and have access to a good internet connexion
  • Able to read language in which the mobile application is available and able to understand pictograms on the application

Exclusion criteria

  • History of lung transplantation
  • Pregnant women and women who are breastfeeding
  • Any medical condition that could interfere with the proper conduct and results of the study (investigator's judgement)
  • Illiteracy in French
  • Inability to use a smartphone or MuCopilot application
  • Persons under legal protection (i.e guardianship or curatorship)
  • Participation in another clinical interventional study within 30 days prior to selection or current participation in another study that, in the opinion of the investigator, could interfere with the proper conduct and results of the study

Treatment and study plan

MuCopilot mobile application

Device

MuCopilot is a software as medical device with three digital tests, to measure patient's with CF, lung function, global exercise capacity and a questionnaire to address CF symptoms and impacts on some aspects of patient's life

Primary outcomes

  1. To demonstrate that the results of the MuCopilot digital tests at home (D1) are associated with the results of their standard counterparts (D0) carried out in-clinic.

    Time frame: During the first in-clinic visit : Day 0 and the first at-home assessment : Day 1

    A correlation coefficient will be used to assess the relationship between MuCopilot tests (MSPT) on Day 1 (at home) and the corresponding standard test on Day 0 (in the clinic). The objective will be achieved if the correlation coefficient reaches a minimum of 0.70. A head-to-head comparison will be performed on the following tests:

    MuCopilot "Mobile Sustained Phonation Test" (MSPT) scores (e.g. maximum phonation time calculated during a sustained phonation assessment) and the FEV1 (air volume exhaled during one second) measured by spirometry at the CF centres

  2. To demonstrate that the results of the MuCopilot digital tests at home (D1) are associated with the results of their standard counterparts (D0) carried out in-clinic.

    Time frame: During the first in-clinic visit : Day 0 and the first at-home assessment : Day 1

    A correlation coefficient will be used to assess the relationship between MuCopilot tests (MEWT) on Day 1 (at home) and the corresponding standard test on Day 0 (in the clinic). The objective will be achieved if the correlation coefficient reaches a minimum of 0.70. A head-to-head comparison will be performed on the following tests:

    MuCopilot "Mobile Endurance Walking Test" (MEWT), i.e. the walking distance (in metres) walked over six minutes on a hard, flat surface at home and the walking distance (in metres) measured during the 6 minute walk test (6MWT) in-clinic

Secondary outcomes

  1. To confirm that the results of the MuCopilot digital test at-home (M3-1) are correlated with their standard clinical counterparts performed in-clinic (M3).

    Time frame: During the last in-clinic visit : at Month 3 and the last at-home assessment Month 3 - 1 day

    According to the same validation principle as for the primaries outcomes, a correlation coefficient will be used to assess the relationship between MuCopilot MSPT and MWT tests at M3 - 1 (at-home) versus standard tests at M3 (in clinic):

    • MuCopilot MSPT scores (sustained phonation time) (at-home) versus Spirometry score FEV1 (air volume exhaled during one second) (in-clinic)
    • MuCopilot MEWT (distance in metres) (at-home) scores versus the 6MWT scores (distance in metres) (in-clinic).
  2. To evaluate the correlation between MuCopilot the digital tests performed in-clinic, and the results of their standard counterparts performed in-clinic

    Time frame: During the first in-clinic visit : Day 0 and the last in-clinic visit : Month 3

    A correlation coefficient will be used to assess the individual correlation between the MuCopilot tests and standard tests performed in-clinic:

    At D0:

    • MSPT scores versus FEV1
    • MEWT scores versus 6MWT At M3
    • MSPT scores versus FEV1
    • MEWT scores versus 6MWT.
  3. To assess intra-patient test-retest reliability of the MuCopilot CF-Scan questionnaire

    Time frame: During at-home visits at Day 1 and Day 3

    Intraclass correlation coefficient (ICC) will be used to assess reliability over time between D1 and D3 (on stable patients without signs or symptoms of CF exacerbation).

    Descriptive analysis of at-home MuCopilot CF-Scan questionnaire between D1 and D3 (Bland-Altman, box plots, etc.).

  4. To assess the safety of the mobile application use.

    Time frame: Throughout the study for all visits in-clinic and at-home Day 0, Day 1, Day 3, Day 5, Day 7, Month 1, Month 2, Month 3-1 day, Month 3

    Adverse events (AEs) and serious adverse events (SAE) related to the use of the application.

  5. To assess the satisfaction and user experience with MuCopilot.

    Time frame: Throughout the study, at-home at Day 7, Month 1, Month 3 - 1 day

    Descriptive analysis of the answers collected in the patient's satisfaction and user experience questionnaires related to the use of MuCopilot.

  6. To assess at-home compliance to MuCopilot.

    Time frame: Throughout the study for all visits in-clinic and at-home Day 0, Day 1, Day 3, Day 5, Day 7, Month 1, Month 2, Month 3-1 day, Month 3

    Descriptive analysis of the number of questionnaires completed or tests performed by patients divided by the number of questionnaires or tests expected by the protocol.

  7. To evaluate the metric performance of the "Mobile Sleep Cough Monitoring Test" (MSCMT) at-home in comparison to ground truth results labelized by specialists. The data will be collected by the mobile application.

    Time frame: Throughout the study for all visits at-home Day 1, Day 3, Day 5, Day 7, Month 1, Month 2, Month 3-1 day and all voluntary tests

    Performance metrics will be used to assess the ability of the digital MSCMT (Mobile Sleep Cough Monitoring Test) trained algorithm to register and automatically detect the cough sounds during sleep to derive :

    • the time spent coughing (in second) : Automatic detection and calculation via MSCMT vs. manual detection and calculcation using annotated data from the mobile application audio record
    • the number of cough : Automatic detection and calculation via MSCMT vs. manual detection and calculation using annotated data from the mobile application audio record.
  8. To evaluate the association between the maximum phonation time determined automatically by the MuCopilot MSPT test at-home and the maximum phonation time evaluated by a trained evaluator on the same dataset.

    Time frame: Throughout the study for all visits at-home Day 1, Day 3, Day 5, Day 7, Month 1, Month 2, Month 3-1 day

    A correlation coefficient will be used to assess the association between the maximum phonation time automatically determined by the MuCopilot MSPT test at-home and the maximum phonation time evaluated by a trained evaluator on the same data (at D1, D3, D5, D7, M1, M2, M3-1).

  9. To evaluate the association between the results of the MSPT and MEWT tests scores in-clinic and the results of these same tests determined by the investigator.

    Time frame: During in-clinic visits at Day 0 and 3 months

    A correlation coefficient will be used to assess the association between the results of the MSPT and MEWT tests scores in-clinic, and the results of these same tests measured simultaneously by the investigator with a stopwatch (time) and an odometer (metres) at D0 and M3.

  10. To assess the reproducibility between MuCopilot tests scores obtained in in-clinic and MuCopilot tests scores obtained at-home

    Time frame: During in-clinic visits at Day 0 and Month 3, and during the at-home visits at Day 1 and Month 3-1 day

    The intraclass correlation coefficient (ICC) will be used to assess the reproducibility of MuCopilot tests MSPT and MEWT performed at :

    • D1 (at-home) versus D0 (in-clinic)
    • M3 -1 (at-home) versus M3 (in-clinic) The minimum ICC to reach is 0.8.
  11. To assess the test-retest reliability of MuCopilot tests scores obtained at-home

    Time frame: During at-home visits at Day 1, Day 3, Day 5 and Day 7

    The intraclass correlation coefficient (ICC) will be used to assess the reliability of MuCopilot tests performed at home at D1, D3, D5 and D7, for MSPT and MEWT digital tests.

    The minimum ICC to reach is 0.8.

Sponsors and collaborators

Lead sponsor

Ad scientiam

Other

Collaborators

  • Vertex Pharmaceuticals Incorporated

Registry information

Acronym: MuControl

Important dates

Study start
2024
Primary completion
2025
Study completion
2025
First posted
Nov 28, 2023
Registry last updated
Nov 28, 2025

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.