Owlet Smart Sock 3 Accuracy on Infants

The purpose of this research study is to evaluate the Owlet Smart Sock 3 wireless pulse oximeter performance in assessing the heart rate and oxygen saturation in neonatal population, while measuring the same parameter with a standard wired pulse oximeter simultaneously. The monitoring duration with the study device will be a maximum of 60 minutes, after which the recorded data from the Owlet Smart Sock 3 and the standard wired pulse oximeters will be de-identified and analyzed.

Completed

Looking for future studies?

Notify Me

Key information

Sex eligibility

All sexes

Study type

Observational

Primary location

University of Utah NICU

Salt Lake City, Utah, 84112, United States

About this study

This prospective observational study compared the HR and SpO2 measurements of the OSS3 (Owlet, Utah, USA) with the Masimo SET® (Masimo Corporation, California, USA) in infants admitted to the University of Utah NICU. The inclusion criteria were infants who weighed 1.7-2.5 kg at the time of monitoring and were medically stable to tolerate routine care. The exclusion criteria were: infants with significant anomalies, congenital deformities, or medical instability. Infants with an intravenous catheter (IV) and skin lesions on either foot were excluded from the study. The wireless device being tested was the OSS3. The gold standard for validating the pulse oximeter's accuracy is arterial blood oxygen saturation (SaO 2 ) at the pulse oximeter sensor site. [6] However, obtaining SaO2 is an invasive arterial blood test for infants. To avoid this conflict, the OSS3 was tested against the reference FDA-approved wired pulse oximeter Masimo.

The study was approved by the University of Utah Institutional Review Board, and informed consent was obtained from parents. During the measurement, OSS3 and Masimo were placed on the infants with one device on each foot. Since the data obtained from the study devices can not be recorded in the hospital electronic medical record, the hospital wired pulse oximeter (Philips, Amsterdam, Netherlands) continued to be used on one of the infant's hands during the study period. OSS3 and Masimo were calibrated appropriately against the hospital wired pulse oximeter with matching HR and SpO 2 readings. Pulse oximetry data were collected simultaneously for 60 minutes on the quiet infants. The two devices (OSS3 and Masimo) continuously recorded HR (beats per minute, bpm) and SpO2 (%) every two seconds. Data from the 2 devices were uploaded and compared according to their time stamped measurements. The paired data agreement was compared using Pearson's correlation coefficient (r), Bland-Altman, Average Root Mean Square (ARMS), and Prevalence and Bias Adjusted Kappa (PABAK) analyses.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Infant who weigh between 1.7kg and 2.3kg
  • Infant who is admitted to U of U NICU

Exclusion criteria

  • Infant who has injuries, deformities, abnormalities, or intravenous catheter in the extremities that may prevent proper application of device
  • Infant who has unstable vital signs or clinical condition
  • No parental consent
  • Inability to achieve adequate fit with the OSS3

Treatment and study plan

Owlet OSS V3

Device

A wearable pulse oximeter designed for monitoring infants and children in a home environment

Primary outcomes

  1. Owlet Oxygen Saturation Accuracy Performance

    Time frame: During the monitoring period (approx. 1 hour)

    Conduct a single-arm observational study to monitor oxygen saturation (SpO2) while a standard wired pulse oximeter (WPOx) will simultaneously evaluate the same parameters in the University of Utah (U of U) Neonatal Intensive Care Unit (NICU).

  2. Owlet Pulse Rate Accuracy Performance

    Time frame: During the monitoring period (approx. 1 hour)

    Conduct a single-arm observational study to monitor heart rate (HR), while a standard wired pulse oximeter (WPOx) will simultaneously evaluate the same parameters in the University of Utah (U of U) Neonatal Intensive Care Unit (NICU).

Sponsors and collaborators

Lead sponsor

Owlet Baby Care, Inc.

Industry

Registry information

Important dates

Study start
2021
Primary completion
2021
Study completion
2021
First posted
Dec 2, 2022
Registry last updated
Feb 22, 2023

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.