Investigation of novel non-CE marked intermittent catheters. The study was a randomized, single blinded, cross-over investigation comparing two new catheters for males with a comparator catheter
Looking for future studies?
Notify Me18 year and older
Male
Interventional
Not applicable
Copenhagen, Denmark
The CP323 study investigated a novel non-CE marked intermittent catheter, designed in two different variants, Variant 1 and Variant 2, respectively, and was conducted in Denmark. The study was a randomized, single blinded, cross-over investigation, comparing the two new variant catheters with a comparator catheter in 15 male IC users.
For each participant, the study thus contains three test visits/treatment arms (Comparator, Variant 1 and Variant 2, respectively), to which the 15 participants were randomized using a randomization sequence of six (i.e., there were six different random options for the order in which the participants tested the three catheters). There were 4-14 days between the test visits.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Intermittent catheterization through the urethra draining the bladder, using SpeediCath® Standard male.
Intermittent catheterization through the urethra draining the bladder, using the new Variant 1 catheter.
Intermittent catheterization through the urethra draining the bladder, using the new Variant 2 catheter.
Time frame: Immediately after the procedure/catheterization, up to 5 min.
The residual urine at 1st flow-stop during nurse catheterization (i.e., catheterization performed by the nurse) represents the volume of urine left in the bladder, the first time the urine stops running out of the catheter during bladder emptying/urination. This was assessed using a pressure sensor with time-logged weighing, meaning that a flow-stop was registered measuring the hydrostatic pressure at the outlet of an intermittent urinary catheter. Simultaneously, the collected urine was weighed and converted to volume, and the residual volume at first flow stop was calculated as the difference between the total urine volume collected and the volume collected at the first flow-stop.
Time frame: Immediately after the procedure/catheterization, up to 5 min.
The residual urine at 1st flow-stop during nurse catheterization (i.e., catheterization performed by the participant) represents the volume of urine left in the bladder, the first time the urine stops running out of the catheter during bladder emptying/urination. This was assessed using a pressure sensor with time-logged weighing, meaning that a flow-stop was registered measuring the hydrostatic pressure at the outlet of an intermittent urinary catheter. Simultaneously, the collected urine was weighed and converted to volume, and the residual volume at first flow stop was calculated as the difference between the total urine volume collected and the volume collected at the first flow-stop.
Time frame: Immediately after the procedure/catheterization, up to 15 min.
The volume of residual urine (urine left in the bladder) post catheterization (performed by the nurse) measured in triplicates using an ultrasound scanner.
Time frame: Immediately after the procedure/catheterization, up to 15 min.
The volume of residual urine (urine left in the bladder) post catheterization (performed by the participant) measured in triplicates using an ultrasound scanner.
Time frame: Immediately after the procedure/catheterization, up to 5 min.
Discomfort measured at catheter insertion (performed by nurse), using the Visual Analogue Scale (VAS). The VAS is a tool used to rate the intensity of certain sensations and feelings, e.g. discomfort/pain. It is a straight line of 10 cm (in this case horizontal) with one end meaning no discomfort/pain (0 cm) and the other end meaning the worst discomfort/pain imaginable (10 cm). The subject visualizes and marks a point on the line that matches the amount of discomfort/pain experienced. A high score reflects a worse outcome than a lower score.
Time frame: Immediately after the procedure/catheterization, up to 5 min.
Discomfort measured during catheter urination/emptying (performed by nurse), using the Visual Analogue Scale (VAS). The VAS is a tool used to rate the intensity of certain sensations and feelings, e.g. discomfort/pain. It is a straight line of 10 cm (in this case horizontal) with one end meaning no discomfort/pain (0 cm) and the other end meaning the worst discomfort/pain imaginable (10 cm). The subject visualizes and marks a point on the line that matches the amount of discomfort/pain experienced. A high score reflects a worse outcome than a lower score.
Time frame: Immediately after the procedure/catheterization, up to 5 min.
Discomfort measured during catheter withdrawal (performed by nurse), using the Visual Analogue Scale (VAS). The VAS is a tool used to rate the intensity of certain sensations and feelings, e.g. discomfort/pain. It is a straight line of 10 cm (in this case horizontal) with one end meaning no discomfort/pain (0 cm) and the other end meaning the worst discomfort/pain imaginable (10 cm). The subject visualizes and marks a point on the line that matches the amount of discomfort/pain experienced. A high score reflects a worse outcome than a lower score.
Time frame: Immediately after the procedure/catheterization, up to 5 min.
Discomfort measured for the entire catheterization procedure (performed by nurse), using the Visual Analogue Scale (VAS). The VAS is a tool used to rate the intensity of certain sensations and feelings, e.g. discomfort/pain. It is a straight line of 10 cm (in this case horizontal) with one end meaning no discomfort/pain (0 cm) and the other end meaning the worst discomfort/pain imaginable (10 cm). The subject visualizes and marks a point on the line that matches the amount of discomfort/pain experienced. A high score reflects a worse outcome than a lower score.
Time frame: Immediately after the procedure/catheterization, up to 5 min.
Discomfort measured at catheter insertion (performed by participant), using the Visual Analogue Scale (VAS). The VAS is a tool used to rate the intensity of certain sensations and feelings, e.g. discomfort/pain. It is a straight line of 10 cm (in this case horizontal) with one end meaning no discomfort/pain (0 cm) and the other end meaning the worst discomfort/pain imaginable (10 cm). The subject visualizes and marks a point on the line that matches the amount of discomfort/pain experienced. A high score reflects a worse outcome than a lower score.
Time frame: Immediately after the procedure/catheterization, up to 5 min.
Discomfort measured during catheter urination/emptying (performed by participant), using the Visual Analogue Scale (VAS). The VAS is a tool used to rate the intensity of certain sensations and feelings, e.g. discomfort/pain. It is a straight line of 10 cm (in this case horizontal) with one end meaning no discomfort/pain (0 cm) and the other end meaning the worst discomfort/pain imaginable (10 cm). The subject visualizes and marks a point on the line that matches the amount of discomfort/pain experienced. A high score reflects a worse outcome than a lower score.
Time frame: Immediately after the procedure/catheterization, up to 5 min.
Discomfort measured during catheter withdrawal (performed by participant), using the Visual Analogue Scale (VAS). The VAS is a tool used to rate the intensity of certain sensations and feelings, e.g. discomfort/pain. It is a straight line of 10 cm (in this case horizontal) with one end meaning no discomfort/pain (0 cm) and the other end meaning the worst discomfort/pain imaginable (10 cm). The subject visualizes and marks a point on the line that matches the amount of discomfort/pain experienced. A high score reflects a worse outcome than a lower score.
Time frame: Immediately after the procedure/catheterization, up to 5 min.
Discomfort measured for the entire catheterization procedure (performed by participant), using the Visual Analogue Scale (VAS). The VAS is a tool used to rate the intensity of certain sensations and feelings, e.g. discomfort/pain. It is a straight line of 10 cm (in this case horizontal) with one end meaning no discomfort/pain (0 cm) and the other end meaning the worst discomfort/pain imaginable (10 cm). The subject visualizes and marks a point on the line that matches the amount of discomfort/pain experienced. A high score reflects a worse outcome than a lower score.
Coloplast A/S
Industry
Exploratory Investigation on Performance and Safety of Newly Developed Intermittent Catheters in Male Users of Intermittent Catheters
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06985888
Female Urogenital Diseases, Female Urogenital Diseases and Pregnancy Complications
Copenhagen, Denmark
View Trial DetailsNCT05485922
Female Urogenital Diseases, Female Urogenital Diseases and Pregnancy Complications
Gandrup, Denmark
View Trial DetailsNCT04557787
Female Urogenital Diseases, Female Urogenital Diseases and Pregnancy Complications
Copenhagen, Denmark
View Trial DetailsNCT05314582
Blood Transfusion Complication, Female Urogenital Diseases
Istanbul, Turkey (Türkiye)
View Trial Details