Icare Versus Haag-Streit Applanation Tonometer

The clinical performance of Icare ONE Tonometer TA02 was compared to a Goldmann type applanation tonometer (Haag-Streit). Data was collected according to ANSI Z80.10-2003 (ISO 8612.2) standard. Subjects were eligible for the study if they had no exclusion criteria specified in the ANSI/ISO standard. One eye from 260 subjects was selected for the study. The minimum sample size according to the standard is 150. One eye of each subject was selected for the study by an investigator, based on the history of the patient, results of an external examination of the eye with slit lamp, keratometry readings, and prior knowledge of the intraocular pressure (IOP) range. The Icare ONE Tonometer TA02 in comparison to the reference tonometer (Goldmann type tonometer).

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Observational

Primary location

Glaucoma Service, Department of Opthalmology, Helsinki University Central Hospital

Helsinki, 00029, Finland

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • subjects were eligible for the study if they had no exclusion criteria as specified by the ANSI/ISO standard

Subjects were selected according to the following exclusion criteria (as given in ANSI Z80.10-2003 and ISO 8612):

  • subjects with only one functional eye
  • those with one eye having poor or eccentric fixation
  • high corneal astigmatism (i.e. those eyes displaying an oval contact image with the Goldmann tonometer)
  • those with corneal scarring or who have had corneal surgery, including corneal laser surgery
  • microphthalmos
  • buphthalmos
  • contact lens wearers
  • dry eyes
  • lid squeezers (blepharospasm)
  • nystagmus
  • keratoconus
  • any other corneal or conjunctival pathology or infection

Sponsors and collaborators

Lead sponsor

Icare Finland Oy

Industry

Registry information

Official study title

Design Compliance Testing of the Icare One Rebound Tonometer According to the American National Standard ANSI Z80.10-2001 and International Standard ISO 8612.2 for Tonometers

Important dates

Study start
2009
Study completion
2009
First posted
Jul 9, 2010
Registry last updated
Jul 9, 2010