The investigators evaluate the effect of postural change on the bronchial cuff pressure (BCP) of double-lumen endotracheal tube (DLT) in patients undergoing thoracic surgery, by observing the pressure of the bronchial cuff before and after lateral decubitus positioning.
Looking for future studies?
Notify Me18 year–70 year
All sexes
Observational
Daegu, 42472, South Korea
It is essential to maintain proper cuff pressure in the endotracheal tube (ETT) when placing the ETT in the trachea in patients undergoing general anesthesia. Most of the literature recommend maintaining a pressure of 20-30 cmH2O to prevent cuff-related complications such as micro-aspiration or airway trauma. Proper pressure is also important for DLTs used in thoracic surgery. Furthermore, the cuff pressure of the ETT can change during the patient's positional changes. All things considered, we hypothesized that the change from supine to lateral decubitus position, which is essential for thoracic surgery would affect the BCP of the DLT.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
After the DLT intubation, the patient is placed in lateral decubitus position
The pressure of the bronchial cuff should be measured with cuff-manometer in supine position, and then measurement should be repeated after lateral positioning.
Time frame: 1. In supine position, 2 minutes after the completion of confirming the double-lumen endotracheal tube (DLT) position via fiberoptic bronchoscope (FOB) 2. In lateral position, 2 minutes after the completion of confirming the DLT position via FOB
A cuff-manometer will be connected to the valve of the pilot balloon of bronchial cuff via a three-way stopcock. And then, the BCP will be assessed, while inflating the cuff with air in 0.5 ml increments from 0 to 3.0 ml. If the BCP exceeds 40 cmH2O during expansion by increasing 0.5 ml from 0 to 3.0 ml, we will stop inflating the bronchial cuff with air and record the bronchial cuff volume (BCV) and BCP up to the last numerical value. During the pressure measurement, air leak or seal around the bronchial cuff will be assessed at each time point when inflating the cuff in 0.5 ml increments.
Time frame: 1. In supine position, 2 minutes after the completion of confirming the DLT position via FOB 2. In lateral position, 2 minutes after the completion of confirming the DLT position via FOB
Time frame: 1. In supine position, 2 minutes after the completion of confirming the DLT position via FOB 2. In lateral position, 2 minutes after the completion of confirming the DLT position via FOB
Daegu Catholic University Medical Center
Other
Effect of Lateral Positioning on Bronchial Cuff Pressure of Left-sided Double-lumen Endotracheal Tube During Thoracic Surgery: a Prospective Observational Study
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT05342675
Lung Cancer, Lung Diseases
San Francisco, California, United States
View Trial DetailsNCT05594485
Artificial Intelligence, Atelectasis
Havířov, Czechia
View Trial DetailsNCT05904574
Lung Cancer, Lung Diseases
Edirne, Turkey (Türkiye)
View Trial DetailsNCT05222568
Lung Cancer, Lung Diseases
Daegu, South Korea
View Trial Details