Joslin Diabetes Center
Boston, Massachusetts, 02215, United States
NCT Number: NCT02733588
The primary objective of this study is to test an optimized control system for sensor-guided (physician administered) delivery of glucagon, and test Proof-of-Concept (POC) in a clinical setting in patients with severe hypoglycemia following bariatric surgery.
Looking for future studies?
Notify Me18 year–65 year
All sexes
Interventional
Phase 2
Boston, Massachusetts, 02215, United States
This is a Phase 2a, single-center, open-label, proof-of-concept study designed to test the ability of the control algorithm to detect and direct timing of G-Pump™ glucagon infused from an OmniPod® pump to prevent hypoglycemia in patients with post-bariatric hypoglycemia syndrome. While the algorithm will provide an alert as to when glucagon should be dosed to prevent hypoglycemia, there will be no automation in this clinical trial and it will ultimately be up to the physician to initiate dosing via the OmniPod® controller.
Participant on continuous glucose monitoring will arrive at the clinic and IV line will be inserted for venous access. Subject will then be asked to drink a liquid mixed meal containing 60 g of carbohydrates, e.g. Boost® Nutritional Drink, over 10 minutes. Blood samples will be collected for glucose and hormone measurement.
The open-loop system will be set to recognize low sensor glucose values, triggering an alert to the physician, who will deliver a bolus of 150 or 300 µg of glucagon via the pump, with the goal of preventing further decline in glucose values. Depending on response, a second bolus dose of 150 or 300 µg of glucagon may be administered. Plasma glucose will be measured after glucagon administration to ensure successful treatment and glucose stability, and glucagon levels will be analyzed concurrently to determine magnitude of increase above baseline. Sensors will be downloaded for subsequent analysis of appropriateness of alert timing and trigger for glucagon bolus delivery.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
0.15 or 0.3 mg G-Pump™ (glucagon infusion) administered from OmniPod® pump
Other names: glucagon
Time frame: 0 - 120 minutes following dosing
Frequency with which the device controller software correctly identifies impending hypoglycemia (glucose < 75 mg/dl) and notifies the investigator to initiate treatment. Reported as the number of successful identifications.
Time frame: 0 - 120 minutes following dosing
Frequency of severe hypoglycemia defined as glucose levels below 60 mg/dl. Reported as the number of subjects with severe hypoglycemia.
Time frame: 0 - 120 minutes following dosing
Frequency of rebound hyperglycemia defined as glucose levels above 180 mg/dl. Reported as the number of subjects with rebound hyperglycemia.
Time frame: 0 - 120 minutes following dosing
Time glucose remains in goal range, 60-180 mg/dl, reported in minutes
Xeris Pharmaceuticals
Industry
A Phase 2 Proof-of-Concept Study of Sensor-Guided, Clinician-Administered Delivery of G-Pump™ (Glucagon Infusion) From an OmniPod® to Prevent Post-Prandial Hypoglycemia in Post-Bariatric Surgery Patients
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT03734718
Autoimmune Diseases, Diabetes Mellitus
Hellerup, Denmark
View Trial DetailsNCT06575426
Autoimmune Diseases, Diabetes Mellitus
Chicago, Illinois, United States
View Trial DetailsNCT05557227
Glucose Metabolism Disorders, Hyperglycemia
Aarhus, Region Midt, Denmark
View Trial DetailsNCT07614139
Carbohydrate Metabolism, Inborn Errors, Congenital Hyperinsulinism
Padova, Italy
View Trial Details