The purpose of this study is to compare blood pressure values (systolic, diastolic, mean arterial pressure) and waveform as recorded by a radial arterial line catheter to the waveform and values as calculated by the Sensifree's algorithm from a PPG sensor from a variety group of patients during elective surgeries.
Looking for future studies?
Notify Me18 year–80 year
All sexes
Observational
Holon, 58100, Israel
Before induction of anesthesia, an arterial line will be inserted to the patient radial artery using a designated short, stiff tubing kit according to catheter IFU (Instructions for Use), preferably under US guidance. Removal of air bubbles from the catheter-transducer system is mandatory prior to connecting the tubing to the catheter.
Study nurse will connect a Sensifree laptop with designated data acquisition software (Medicollector BEDSIDE) to the multi-parameter monitor communication port and start the recording of invasive blood pressure and pulse oximeter pleth (PPG sensor) per software IFU. Prior to connecting the arterial line system to the patient, a calibration (flush test) will be done and recorded. During the study, any deterioration in the waveform recorded by the intra-arterial catheter will be noted and appropriate corrective measures will be taken as soon as possible (e.g., flushing or adjusting the position of the catheter).
Preferable NIBP (Non Invasive Blood Pressure) measurement rate will be every 10 minutes, unless the medical procedure requires otherwise.
Calibration procedure will be repeated every 60 +/- 5 minutes and at the end of the surgery. Recording will take place for the duration of the surgery.
NIBP cuff will be placed over the arterial-line arm, NIBP air hose will be replaced with an air hose supplied by Sensifree and the pressure gauge in it will be connected to Sensifree laptop via analog to digital converter in order to capture the data.
Medication used during the surgery will be listed and documented in the CRF (Case Report Form) including drugs name, dose, time and method of administration.
Offline data analysis will compare the blood pressure values as calculated by the Sensifree's algorithm based on the PPG signal, to the values obtained from the arterial line. The comparison results will be measured against the accuracy performance requirements per ISO 81060-2.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Invasive monitoring of blood pressure via catheterization of the radial artery, displaying a continuous pressure waveform
PPG based, fingertip mounted sensor, that measures the oxygen level (oxygen saturation) of the blood and generates a waveform
A blood pressure cuff is used to measure blood pressure. The cuff has an inflatable rubber bladder that is typically fastened around the arm. A pressure meter indicates the cuff's pressure.
Time frame: 3 months after last patient enrolled
Blood pressure waveform will be acquired from the PPG while blood pressure waveform and values (systolic, diastolic, mean) will be acquired from A-line. PPG data will be analyzed by the Sensifree's algorithm and compared to the A-line measurement results according to ISO 81060-2 accuracy specifications
Sensifree Ltd.
Industry
Evaluation of Sensifree's Algorithm for Continuous Blood Pressure Measurements Based on a PPG Sensor's Signal Compared to Invasive Radial Arterial Line
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06915376
Blood Pressure Measurement
Konya, Konya/Meram, Turkey (Türkiye)
View Trial DetailsNCT07024472
Blood Pressure, Blood Pressure Measurement
Helsinki, Finland
View Trial DetailsNCT04976452
Blood Pressure Measurement
Youngstown, Ohio, United States
View Trial DetailsNCT04510753
Blood Pressure Measurement
San Francisco, California, United States
View Trial Details