Comparison of a Torso-Based Derived 6-Lead Electrocardiogram (ECG) Device With Standard 12-Lead Electrocardiogram (ECG) in Healthy Volunteers

This study aims to compare electrocardiographic interval measurements obtained from a torso-placed six-lead electrocardiogram (ECG) device (torso-derived ECG; T-ECG) with those obtained from a standard 12-lead electrocardiogram (ECG) in healthy adult volunteers. The torso-derived device synthesizes limb-lead signals using electrodes positioned on the thorax, potentially allowing for faster and more practical ECG acquisition in emergency department settings, operating rooms, and prehospital environments.

A total of 24 participants undergo both standard 12-lead ECG recording and torso-derived 6-lead ECG recording during the same study session. Heart rate (HR), PR interval, QRS duration, QT interval, and corrected QT (QTc) values are analyzed using Bland-Altman agreement analysis and correlation testing. The study also evaluates whether anthropometric characteristics-including body mass index (BMI)-influence the compatibility and agreement between the two measurement systems.

Completed

Looking for future studies?

Notify Me

Key information

Conditions

Age range

18 year–90 year

Sex eligibility

All sexes

Study type

Observational

Primary location

Haydarpaşa Numune Training and Research Hospital - Emergency Medicine Department

Istanbul, 34668, Turkey (Türkiye)

About this study

This observational, single-center clinical study evaluates the agreement between a torso-based derived six-lead electrocardiogram (ECG) system and a standard 12-lead electrocardiogram (ECG) in healthy adult volunteers. The study is conducted at Haydarpasa Numune Training and Research Hospital following ethics committee approval (Approval No: 2023/193). Volunteers aged 18 to 65 years without known cardiac disease are enrolled.

Each participant undergoes one standard 12-lead ECG recording followed by a torso-based six-lead ECG recording under identical resting conditions. Electrocardiographic measurements include heart rate (HR), RR interval, PR interval, QRS duration, QT interval, and corrected QT (QTc). Anthropometric characteristics-body weight, height, and body mass index (BMI)-are collected to assess whether body composition influences measurement consistency between the two systems.

Primary analysis includes Bland-Altman agreement statistics and correlation testing between ECG modalities. No drug administration, invasive procedure, randomization, or therapeutic intervention is applied. The study involves minimal risk and is limited to non-invasive surface ECG acquisition.

The expected clinical significance is that, if the torso-based ECG demonstrates measurement compatibility with the standard 12-lead ECG, simplified electrode placement may enable faster ECG acquisition in emergency departments, prehospital evaluations, and clinical scenarios in which limb lead placement is impractical.

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

Age between 18 and 65 years

No known cardiovascular disease

No arrhythmia, conduction disorder, or ischemic electrocardiogram (ECG) changes at baseline

Willingness to participate and ability to provide written informed consent

Exclusion criteria

Known cardiac disease, including coronary artery disease (CAD), arrhythmia, heart failure, or valvular heart disease

Pacemaker rhythm or baseline abnormal electrocardiogram (ECG) findings

Thoracic deformity or prior thoracic surgery that prevents appropriate electrode placement

Unwillingness to participate or inability to provide informed consent

Treatment and study plan

Torso-Based 6-Lead ECG Recording

Device

Non-invasive ECG measurement with torso electrode placement for comparison with standard 12-lead ECG.

Primary outcomes

  1. PR Interval Agreement Between Torso-Derived and Standard 12-Lead Electrocardiogram (ECG)

    Time frame: Single-session baseline ECG recording.

    Mean difference in PR interval between torso-derived six-lead electrocardiogram (ECG) and standard 12-lead ECG recordings obtained during the same resting session.

    Single-session baseline ECG recording.

  2. QRS Duration Agreement Between Torso-Derived and Standard 12-Lead Electrocardiogram (ECG)

    Time frame: Single-session baseline ECG recording.

    Mean difference in QRS duration between torso-derived six-lead ECG and standard 12-lead ECG recordings obtained during the same resting session

  3. QT Interval Agreement Between Torso-Derived and Standard 12-Lead Electrocardiogram (ECG)

    Time frame: Single-session baseline ECG recording

    Mean difference in QT interval between torso-derived six-lead ECG and standard 12-lead ECG recordings obtained during the same resting session.

  4. Heart Rate (HR) Agreement Between Torso-Derived and Standard 12-Lead Electrocardiogram (ECG)

    Time frame: Single-session baseline ECG recording.

    Mean difference in heart rate (HR) between torso-derived six-lead ECG and standard 12-lead ECG recordings obtained during the same resting session.

Secondary outcomes

  1. Agreement Between Torso-Derived 6-Lead Electrocardiogram (ECG) and Standard 12-Lead Electrocardiogram (ECG) Measurements

    Time frame: Single-session recording (baseline only)

    Description:

    Comparison of PR interval, QRS duration, QT interval, and heart rate (HR) measurements obtained from torso-derived 6-lead electrocardiogram (ECG) recordings and standard 12-lead electrocardiogram (ECG) recordings to assess measurement concordance.

    Time Frame:

    Single-session recording (baseline only)

Sponsors and collaborators

Lead sponsor

Haydarpasa Numune Training and Research Hospital

Other

Registry information

Official study title

Comparison of 6-Lead Electrocardiogram (ECG) Device Data With Standard 12-Lead Electrocardiogram (ECG) Data in Individuals With Different Anthropometric Characteristics

Acronym: T-ECG

Important dates

Study start
2023
Primary completion
2024
Study completion
2024
First posted
Dec 11, 2025
Registry last updated
Dec 11, 2025

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.