AcrySof® IQ ReSTOR® +2.5 D Multifocal IOL Model SN6AD2 [SV25T0]
DeviceMultifocal IOL with extended secondary focal point implanted for long-term use over the lifetime of the cataract patient
NCT Number: NCT01510717
The purpose of this study is to compare the safety and efficacy of an investigational multifocal intraocular lens (IOL) to an FDA-approved monofocal IOL.
Looking for future studies?
Notify Me21 year and older
All sexes
Interventional
Not applicable
Participants 21 years and older in need of cataract extraction were consented and randomized in a 1:1 ratio at each investigative site to receive either the AcrySof® IQ ReSTOR® +2.5 D Multifocal IOL Model SN6AD2 [SV25T0] or the AcrySof® IQ Monofocal IOL Model SN60WF in both eyes. Each participant completed a preoperative examination of both eyes, implantation of IOL at the operative visit for each eye, and up to 7 postoperative visits, with monocular testing at Day 1-2, Day 7-14, and Day 30-60, and monocular and binocular testing at Day 120-180 following the second implantation. The first eye to be implanted was considered the primary eye. The second eye was implanted 7 and 30 days of the first. If the first eye was not implanted during surgery, the second eye was not eligible. If the second eye was not implanted during surgery, only the first eye was followed.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Multifocal IOL with extended secondary focal point implanted for long-term use over the lifetime of the cataract patient
Monofocal IOL implanted for long-term use over the lifetime of the cataract patient
Time frame: Day 120-180 from second eye implantation
Visual acuity (VA) was tested monocularly (each eye separately) using the manifest refraction adjusted for optical infinity and the hand-held, 100% contrast, Early Treatment Diabetic Retinopathy Study (ETDRS) chart set at 53 centimeters (cm) on the nearpoint rod. VA was measured in logMAR (logarithm of the minimum angle of resolution), with 0.1 logMAR increment corresponding to 5 letters, or 1 line, on an ETDRS chart. A lower numeric value represents better visual acuity. This analysis was prespecified for the primary eye.
Time frame: Day 0 first operative eye visit, up to Day 120-180 from second eye implantation
Cumulative and persistent adverse events were collected. This outcome measure was prespecified for the multifocal IOL.
Time frame: Day 120-180 from second eye implantation
Contrast sensitivity (ie, the ability to detect objects by distinguishing them from their background) was assessed binocularly with the participant's best spectacle correction under photopic (bright) conditions at a distance of 8 feet at spatial frequencies of 3, 6, 12, and 18 cycles per degree (cpd) using the Vector Vision CSV 1000 without a glare source. Raw scores from contrast sensitivity testing were transformed to log units. Scores of (-1) were set to missing; hence, the mean measures may be overestimated and the variability measures may be underestimated. A higher numeric value represents better contrast sensitivity.
Time frame: Day 120-180 from second eye implantation
Contrast sensitivity was assessed binocularly with the participant's best spectacle correction under photopic (bright) conditions at a distance of 8 feet at spatial frequencies of 3, 6, 12, and 18 cycles per degree (cpd) using the Vector Vision CSV 1000 with a glare source. Raw scores from contrast sensitivity testing were transformed to log units. Scores of (-1) were set to missing; hence, the mean measures may be overestimated and the variability measures may be underestimated. A higher numeric value represents better contrast sensitivity.
Time frame: Day 120-180 from second eye implantation
Contrast sensitivity was assessed binocularly with the participant's best spectacle correction under mesopic (dim lighting) conditions at a distance of 8 feet at spatial frequencies of 1.5, 3, 6, and 12 cycles per degree (cpd) using the Vector Vision CSV 1000 without a glare source. Raw scores from contrast sensitivity testing were transformed to log units. Scores of (-1) were set to missing; hence, the mean measures may be overestimated and the variability measures may be underestimated. A higher numeric value represents better contrast sensitivity.
Time frame: Day 120-180 from second eye implantation
Contrast sensitivity was assessed binocularly with the participant's best spectacle correction under mesopic conditions at a distance of 8 feet at spatial frequencies of 1.5, 3, 6, and 12 cpd using the Vector Vision CSV 1000 with a glare source. Raw scores from contrast sensitivity testing were transformed to log units. Scores of (-1) were set to missing; hence, the mean measures may be overestimated and the variability measures may be underestimated. A higher numeric value represents better contrast sensitivity.
Time frame: Day 120-180 from second eye implantation
VA was tested monocularly (each eye separately) using the correction obtained from the manifest refraction and 100% contrast, ETDRS charts at a distance of 4 meters. VA was measured in logMAR, with 0.1 logMAR increment corresponding to 5 letters, or 1 line, on an ETDRS chart. A lower numeric value represents better visual acuity. This analysis was prespecified for the primary eye.
Time frame: Day 120-180 from second eye implantation
VA was tested monocularly (each eye separately) using the manifest refraction adjusted for optical infinity and the hand-held, 100% contrast, ETDRS chart set at 40 cm on the nearpoint rod. VA was measured in logMAR, with 0.1 logMAR increment corresponding to 5 letters, or 1 line, on an ETDRS chart. A lower numeric value represents better visual acuity. This analysis was prespecified for the primary eye.
Time frame: Day 120-180 from second eye implantation
Overall Spectacle Independence was rated using SILVER (Spectacle Independence Lens Vision Evaluation and Repurchase), a new patient reported outcome questionnaire. The participant was asked, "How often do you wear eyeglasses or contact lenses overall?"
Time frame: Day 120-180 from second eye implantation
Near Spectacle Independence was rated using SILVER, a new patient reported outcome questionnaire. The participant was asked, "How often do you wear eyeglasses or contact lenses for seeing objects up close?"
Alcon Research
Industry
Clinical Investigation of AcrySof® IQ ReSTOR® +2.5 D Multifocal Intraocular Lens (IOL) Model SN6AD2
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06377514
Cataract, Cataracts
Bakersfield, California, United States
View Trial DetailsNCT05575063
Cataract, Cataracts
Bakersfield, California, United States
View Trial DetailsNCT02535819
Cataract, Cataracts
Metz, France
View Trial DetailsNCT06250322
Cataract, Cataracts
Chatswood, New South Wales, Australia
View Trial Details