NCT Number: NCT06609486

Clinical Evaluation of the Performance of Bionet Sphygmomanometer Modules Based on Protocol Presented by the International Organization for Standardization

According to INTERNATIONAL STANDARD (hereafter ISO) blood pressure (BP) monitor clinical evaluation protocol (ISO 81060-2:2018, Non-invasive sphygmomanometers- Part2: Clinical investigation of intermittent automated measurement type) for BN1 (BNiBP), an electronic sphygmomanometer module newly developed by Bionet Co., Ltd., the clinical group and evaluation criteria presented in, the effectiveness of blood pressure measurement performance was evaluated.

The auscultation method used as a reference (gold standard) of a non-invasive blood pressure (NIBP) measurement method was used. The first blood pressure measurement was performed once each (reference BP not used for evaluation, R0; test BP not used for evaluation, T0) with auscultation and the test medical device (BM5 equipped with BN1 (BNiBP)). The blood pressure used for evaluation was measured four times by auscultation (first reference BP to fourth reference BP, R1 to R4) and three times by the test medical device (first test BP to third test BP, T1 to T3) alternately. A minimum interval of 1 min was allowed between each blood pressure measurement.

Three data pairs (mean value of R1, R2 & T1 / R2, mean value of R3 & T2 / R3, mean value of R4 & T3) were obtained from one study subject. This study aims to verify that the mean value of the differences and standard deviations of data pairs meet the criteria of ISO 81060-2:2018.

Completed

Looking for future studies?

Notify Me

Key information

Age range

3 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Hanyang University Seoul Hospital

Seoul, 04763, South Korea

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Recruitment is made through the recruitment notice for patients and their guardians who visit Hanyang University Hospital. Subjects who meet the criteria presented in ISO 81060-2:2018 Protocol 5.1 Subject requirements are recruited.

Exclusion criteria

  • Emergency surgery patient
  • Persons who have mental disorders such as mental retardation, autism, etc. and who have difficulty measuring blood pressure or are expected to have difficulties
  • A person who disagreed with the study

Treatment and study plan

Blood pressure measurement group

Device

In this study, the protocol applied to automatic electronic blood pressure system module BN1 for performance evaluation is ISO's proposed "ISO 81060-2:2018 Non-invasive sphygmomanometers- Part2: Clinical investment of internal automatic measurement type". The initial BP measurement is performed alternately, measuring the blood pressure by the measurer and the test device once each, and measuring blood pressure by the measurer four times by the observer and three times by the test device.

Primary outcomes

  1. Test Device's Blood Pressure of Resting Condition

    Time frame: Right after the experiment

    The blood pressure used for evaluation was measured four times by auscultation (first reference BP to fourth reference BP, R1 to R4) and three times by the test medical device (first test BP to third test BP, T1 to T3) alternately. A minimum interval of 1 min was allowed between each BP measurement. Three data pairs (mean value of R1, R2 & T1 / R2, mean value of R3 & T2 / R3, mean value of R4 & T3) were obtained from one study subject.

    Criterion 1 is that for valid data pairs, the mean value of the difference between the test medical device and the auscultation method BP (test device BP - auscultation BP) should be 5 mmHg or less and the standard deviation (SD) should be 8 mmHg or less.

    Criterion 2 is that, for study subjects, the ISO 81060-2:2018 for presenting different SD thresholds depending on the mean value of the difference should be satisfied. When mean value of the difference is ±1.6, ±0.0 mmHg, the SD should be less than 6.76 and 6.95 mmHg, respectively.

Sponsors and collaborators

Lead sponsor

Hoon-Ki Park

Other

Collaborators

  • Bionet Co., Ltd

Registry information

Official study title

Clinical Evaluation of Bionet Sphygmomanometer Module BN1 Based on ISO 81060-2:2018 Protocol

Important dates

Study start
2020
Primary completion
2021
Study completion
2021
First posted
Sep 24, 2024
Registry last updated
Mar 3, 2025