NCT Number: NCT06609486
According to INTERNATIONAL STANDARD (hereafter ISO) blood pressure (BP) monitor clinical evaluation protocol (ISO 81060-2:2018, Non-invasive sphygmomanometers- Part2: Clinical investigation of intermittent automated measurement type) for BN1 (BNiBP), an electronic sphygmomanometer module newly developed by Bionet Co., Ltd., the clinical group and evaluation criteria presented in, the effectiveness of blood pressure measurement performance was evaluated.
The auscultation method used as a reference (gold standard) of a non-invasive blood pressure (NIBP) measurement method was used. The first blood pressure measurement was performed once each (reference BP not used for evaluation, R0; test BP not used for evaluation, T0) with auscultation and the test medical device (BM5 equipped with BN1 (BNiBP)). The blood pressure used for evaluation was measured four times by auscultation (first reference BP to fourth reference BP, R1 to R4) and three times by the test medical device (first test BP to third test BP, T1 to T3) alternately. A minimum interval of 1 min was allowed between each blood pressure measurement.
Three data pairs (mean value of R1, R2 & T1 / R2, mean value of R3 & T2 / R3, mean value of R4 & T3) were obtained from one study subject. This study aims to verify that the mean value of the differences and standard deviations of data pairs meet the criteria of ISO 81060-2:2018.
Looking for future studies?
Notify Me3 year and older
All sexes
Interventional
Not applicable
Seoul, 04763, South Korea
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
In this study, the protocol applied to automatic electronic blood pressure system module BN1 for performance evaluation is ISO's proposed "ISO 81060-2:2018 Non-invasive sphygmomanometers- Part2: Clinical investment of internal automatic measurement type". The initial BP measurement is performed alternately, measuring the blood pressure by the measurer and the test device once each, and measuring blood pressure by the measurer four times by the observer and three times by the test device.
Time frame: Right after the experiment
The blood pressure used for evaluation was measured four times by auscultation (first reference BP to fourth reference BP, R1 to R4) and three times by the test medical device (first test BP to third test BP, T1 to T3) alternately. A minimum interval of 1 min was allowed between each BP measurement. Three data pairs (mean value of R1, R2 & T1 / R2, mean value of R3 & T2 / R3, mean value of R4 & T3) were obtained from one study subject.
Criterion 1 is that for valid data pairs, the mean value of the difference between the test medical device and the auscultation method BP (test device BP - auscultation BP) should be 5 mmHg or less and the standard deviation (SD) should be 8 mmHg or less.
Criterion 2 is that, for study subjects, the ISO 81060-2:2018 for presenting different SD thresholds depending on the mean value of the difference should be satisfied. When mean value of the difference is ±1.6, ±0.0 mmHg, the SD should be less than 6.76 and 6.95 mmHg, respectively.
Hoon-Ki Park
Other
Clinical Evaluation of Bionet Sphygmomanometer Module BN1 Based on ISO 81060-2:2018 Protocol
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT05725148
Blood Pressure, Cardiovascular Diseases
Gyeonggi-do, Pangyo, South Korea
View Trial DetailsNCT04082819
Blood Pressure, Cardiovascular Diseases
Markham, Ontario, Canada
View Trial DetailsNCT05367544
Blood Pressure, Cardiovascular Diseases
Birmingham, Alabama, United States
View Trial DetailsNCT06513975
Blood Pressure, Blood Pressure, High
Seoul, South Korea
View Trial Details