SE Health
Markham, Ontario, L3R 6H3, Canada
NCT Number: NCT04082819
The goal of the study was to compare the current gold standard blood pressure instrument (a mercury-filled sphygmomanometer with cuff and stethoscope) with the new experimental algorithm developed by HeartBeat Technologies Ltd. To that end, participants were recruited from Markham, Ontario and participated in a series of alternating blood pressure measurements over a period of 45 minutes. Comparisons were made between manual measurements and device measurements to determine the quality of the device readings. Measurements were taken by trained nurses and staff were consistent across the study duration.
Looking for future studies?
Notify Me25 year and older
All sexes
Interventional
Not applicable
Markham, Ontario, L3R 6H3, Canada
The gold standard instrument for blood pressure measurement is a mercury-filled sphygmomanometer with cuff and stethoscope (Canadian Agency for Drugs and Technologies in Health; Canadian Electronic Library, 2012). However, such a device is not the most convenient for client self-monitoring or virtual (remote) care. Thus, there is a desire for an accurate, reliable alternative for the traditional mercury-containing sphygmomanometer.
HeartBeat Technologies Ltd. ("Heartbeat") has developed a novel approach to measuring blood pressure after an initial blood pressure reading using the conventional measurement method, supplemented by specific characteristics of a person (age, gender, height, weight, and heart rate), has established a "baseline measurement" for the person. The "novel approach" uses a finger pulse oximeter, the Contec CMS50EW device (Figure 1) (manufactured by Contec Medical Systems Co., Ltd., Shanghai, People's Republic of China). The finger pulse oximeter detects the changes in blood volume directly below the person's skin and indirectly measures oxygen saturation in the blood. The measurement is in the form of a photoplethysmogram (PPG) which is captured by the Contec oximeter and then, by Bluetooth technology, transmitted to a smartphone or tablet. A HeartBeat application, called MediBeat, on the smartphone or tablet then transmits the PPG to a server where a proprietary algorithm analyzes the baseline measurement for the person and the PPG to calculate the person's current blood pressure.
Heartbeat's intent is that the current blood pressure reading would then be made available to both the client and the health provider through separate web-based interfaces. Providers would be able to log onto the HeartBeat web-based application from anywhere to review and monitor the blood pressure of multiple clients, while clients can log onto the application to track their own measurements.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Participants' blood pressure will be measured 3 times with the wearable finger pulse oximeter at their study visit.
Participants' blood pressure will be measured 4 times manually with a blood pressure cuff at their study visit.
Time frame: 45 - 60 minutes
Each device reading was compared to the manual blood pressure reading taken before (R1) and after it (R2). The degree of alignment for systolic blood pressure was calculated by taking the difference between the two scores (manual and device). The reading with the best alignment was used.
Time frame: 45-60 minutes
Each device reading was compared to the manual reading before (R1) and after (R2) it, to calculate the degree of alignment. The difference of the device and R1, and the device and R2 were taken. The smaller number was used for results reporting.
SE Health
Other
Determining the Accuracy of the HeartBeat Algorithm for Calculating Blood Pressure
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06491433
Cardiovascular Diseases, Hypertension
Seoul, South Korea
View Trial DetailsNCT02452125
Basal Ganglia Diseases, Brain Diseases
Old Westbury, New York, United States
View Trial DetailsNCT03182335
Cardiovascular Diseases, Confusion
Chicago, Illinois, United States
View Trial DetailsNCT06319248
Cardiovascular Diseases, Hypotension
Scottsdale, Arizona, United States
View Trial Details