The First Affiliated Hospital of Zhejiang University
Hangzhou, Zhejiang, 310000, China
NCT Number: NCT03828357
This clinical study is to evaluate the safety and effectiveness of the Ellipse VR/DR implantable cardiac defibrillators (ICDs) and the Quadra Assura MP cardiac resynchronization therapy defibrillators (CRT-Ds), with Durata or Optisure defibrillation leads, Tendril STS or Isoflex pacing leads, and the Quartet quadripolar leads in a 1.5T MRI environment for MR-conditional labeling expansion of these market-approved ICD/CRT-D systems in China.
Note: Protocol updated: Fortify Assura VR/DR ICD removed; Quadra Assura MP CRT-D and Quartet leads added. The Detailed Description Section below was updated upon protocol amendment.
Looking for future studies?
Notify Me18 year and older
All sexes
Observational
Hangzhou, Zhejiang, 310000, China
Endpoints, and contents of clinical study
Objectives:
The objective of this clinical study is to evaluate the safety and effectiveness of the Ellipse VR/DR implantable cardiac defibrillators (ICDs) and the Quadra Assura MP cardiac resynchronization therapy defibrillators (CRT-Ds), with Durata or Optisure defibrillation leads, Tendril STS or Isoflex pacing leads, and the Quartet quadripolar leads in a 1.5T MRI environment for MR-conditional labeling expansion of these market-approved ICD/CRT-D systems in China.
There are four device combination groups in this clinical study as outlined below:
Endpoints
Primary Endpoints
Descriptive Endpoint(s)
Descriptive endpoints are reported using only summary statistics and no hypothesis tests will be performed.
The following data will be collected:
This is a prospective, multi-center, single-arm study to evaluate the safety and effectiveness of the Ellipse VR/DR ICDs, and the Quadra Assura MP CRT-D, with Durata or Optisure defibrillation leads, Tendril STS or Isoflex pacing leads, and the Quartet quadripolar LV lead in a 1.5T MRI environment for MR-conditional labeling expansion in China.
This study will be conducted in at least 6 centers in China and at least 60 subjects who provide consent for participation will be enrolled. Subjects will be considered enrolled after providing consent. Subjects will be implanted with any one of the device/lead combination groups outlined in Figure 1, depending on their clinical need as determined by the Investigators. At least 15 subjects will be enrolled in each group.
The enrollment group and group 4 lead pairing will be at the discretion of the Investigator, dependent on patient needs in conjunction with study needs to fulfill study goals.
At least forty-five (45) days after system implantation or after a system revision, subjects will have a baseline visit followed by an MRI visit; subjects will undergo a study-related MRI scan at 1.5T. Subjects will have a 1-month follow-up after the study MRI scan. The minimum duration of each subject's participation is approximately 2.5 months from enrollment. The expected duration of enrollment is approximately 14 months. The total duration of clinical study is expected to be 16.5 months.
The study population includes male and female Chinese nationals that meet the eligibility requirements and give consent for the study. Vulnerable subjects, such as minors or those unable to provide consent are excluded from participating.
Inclusion criteria
Assessment for general eligibility criteria is based on medical records of the site and interview with a candidate patient. If some of the clinical and laboratory tests are not included in site standard tests, they must be done but after written informed consent is obtained. Patients must meet ALL the inclusion criteria to be considered for the clinical study. If ANY of the exclusion criteria are met, the patient is excluded from the clinical study and cannot be enrolled.
To participate in this clinical study, the subject must meet all the following inclusion criteria:
Exclusion criteria
Subjects who meet any of the following exclusion criteria must be excluded from the clinical investigation:
Effectiveness evaluation method
There are two effectiveness endpoints to be evaluated in this study.
The effectiveness parameters are recorded directly from the device. The device measurements will be obtained using the Merlin programmer to interrogate the device during study visits. The analysis of the effectiveness parameters will be conducted on all subjects with complete data from pre- to 1-month post-MRI scan.
Safety evaluation method There is one safety primary endpoint to be evaluated in this study.
Safety surveillance within this study and the safety reporting, both performed by the investigator, starts as soon as the subject completes the pre-MRI visit testing and the MRI Settings have been programmed according to the protocol. The safety surveillance and the safety reporting will continue until the last study visit has been performed, the subject is deceased, the subject/investigator concludes his participation into the clinical study or the subject withdrawal from the clinical study.
Monitoring plan
Sponsor and/or designee will monitor the clinical study over its duration according to the CIP-specific monitoring plan which will include the planned extent of source data verification.
Prior to initiating any procedure, the Sponsor monitor (or delegate) will ensure that the following criteria are met:
Statistical considerations
This study is an observational study to meet the requirements for device approval by the National Medical Products Administration (NMPA, previously China FDA). There is no endpoint hypotheses testing planned in this study. The endpoints, demographics, safety and device data will be summarized using descriptive statistics as outlined in the Statistical Analysis Plan (SAP).
Any major changes to the SAP will be documented in an amendment to the SAP. Less significant changes to the planned analyses will be documented in the final report(s).
Data management
Study data will be handled according to China Device GCP (Order 25). The Leading PI and his/her associated organization (leading Unit) will be responsible for compiling and submitting all required reports to governmental agencies.
Before the submission, Leading PI and his/her associated organization (leading Unit) should allow sponsor to have a copy of the above reports for reviewing, and providing comments for Leading PI and his/her associated organization consideration.
Data will be analyzed by the Leading PI and his/her associated organization (leading Unit) and/or its affiliates. The Leading PI and his/her associated organization (leading Unit)'s also agree that data may be transferred to the Sponsor's locations worldwide and/or any other worldwide regulatory authority in support of a market-approval application.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
To participate in this clinical study, the subject must meet all of the following inclusion criteria:
Exclusion criteria
Subjects who meet any of the following exclusion criteria must be excluded from the clinical investigation:
At least forty-five (45) days after system implantation or after a system revision, Subjects will undergo a study-related MRI scan at 1.5T.
Time frame: 1 month from MRI scan
Freedom from Magnetic Resonance Imaging (MRI) scan related complications related to the Implantable Cardiac Difibrillator (ICD) device and/or leads from the time of the MRI scan to 1-month post-MRI scan testing.
Time frame: pre-MRI scan to 1 month post-MRI scan
Proportion of leads with a capture threshold increase of ≤ 0.5V at 0.5ms from pre-MRI scan to 1-month post-MRI scan testing.
Time frame: pre-MRI scan to 1 month post-MRI scan
Proportion of subjects with a sensing amplitude decrease ≤ 50% from pre-MRI scan to 1-month post MRI scan % (n/N)
Time frame: Pre-MRI scan, Post MRI scan, 1 month follow-up
Summaries of electrical measurements taken in the Analysis Cohort at Pre-MRI scan, Post-MRI scan, and 1-month post MRI scan. The stable electrical measurements from pre-MRI scan to post-MRI scan and 1-month post MRI scan indicate that the MRI scan did not impact the electrical performance of the Ellipse ICD or Quadra Assura CRT-D system
Time frame: during MRI scan
Whole body (WB) SAR values from the RF intensive MRI scans.
Abbott is seeking MR-conditional labeling for MRI scans in Normal Operating Mode (WB Specific Absorption Rate (SAR) up to 2 W/kg). Achieving a WB SAR of exactly 2 W/kg is not feasible since the MR technologist must adjust multiple interdependent scab parameters to increase the WB SAR. Therefore, an acceptable limit of ≥ 1.8 W/kg (within 10%) was established as part of the MRI scan sequence guideline document for this study prior to site enrollments to account for expected user variability.
Time frame: MRI scan visit
Subjects that returned to usual programming after the MRI scan and the number of subjects experiencing delays in disabling MRI settings after the MRI scan
Abbott Medical Devices
Industry
The Evaluation of ICD Defibrillation Systems in a 1.5T Magnetic Resonance Imaging Environment in China
Acronym: CN_MRI_ICD
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07258628
Arrhythmias, Cardiac, Cardiac Conduction System Disease
Ankara, Turkey (Türkiye)
View Trial DetailsNCT07385040
Arrhythmias, Cardiac, Cardiac Conduction System Disease
Karachi, Sindh, Pakistan
View Trial DetailsNCT06038123
Arrhythmias, Cardiac, Bradycardia
Guangzhou, Guangdong, China
View Trial DetailsNCT01259648
Arrhythmias, Cardiac, Cardiac Conduction System Disease
Nîmes, Gard, France
View Trial Details