NCT Number: NCT04820504

Augmented Infant Resuscitator to Enhance Newborn Ventilation

Augmented Infant Resuscitator (AIR) is an inexpensive add-on, compatible with nearly every existing bag-valve mask and many types of ventilation equipment. AIR monitors ventilation quality and provides real-time objective feedback and actionable cues to clinicians to both shorten training times and improve resuscitation quality, adoption, retention, and confidence.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

About this study

Longitudinal, randomized control trial that examined the effectiveness of AIR among 270 participants, with 1,080 recorded ventilation scenarios, and across 20 sites in Uganda and the U.S. Birth attendants - including a mix of midwives, nurses, respiratory therapists, pediatricians, and neonatologists - with recent HBB and/or NRP training were recruited to participate in this randomized control trial. Participants ventilated training mannequins for a fixed duration of time. Participants were randomized to receive visual feedback from AIR (intervention) or have the AIR feedback covered (control). They were then requested to administer effective ventilation and verbally assess the mannequin condition. All mannequins appeared to be identical, but each session's mannequin was randomly selected with an upper airway that was normal, partially leaking, or partially obstructed.

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Midwives, nurses, respiratory therapists, pediatricians, and neonatologists working as birth attendants in select hospitals in Uganda and the U.S. who are either previous trained in Helping Babies Breathe (HBB) and/or have a current Newborn Resuscitation Program (NRP) certification.

Exclusion criteria

  • None

Treatment and study plan

Augmented Infant Resuscitator (AIR)

Device

Device provides visual feedback related to air leak, obstruction, hyperventilation, hypoventilation, and harsh breaths

Other names: AIR device with feedback

AIR device without feedback

Device

Device provides no visual feedback to clinician but records data on ventilation effectiveness

Primary outcomes

  1. Effective bag-mask ventilation

    Time frame: 2 minutes

    Duration of effective bag-mask ventilation by clinician during a 2-minute trial. [This will be a cumulative duration of the two minutes. This is measured internally from the AIR device. "Effective ventilation" is defined as having all three of the following conditions simultaneously: (1) Correct rate (30-60 breaths per minute); (2) Absence of significant leak (compliance >4.0mL/cm H2O); and (3) Absence of a significant airway blockage (defined as compliance <0.10 mL/cm H2O or resistance >90 cmH2O/L/s). These measurements are determined by sensors in the AIR device for both arms of the study. Their accuracy for manikin use was determined to be 100% in the following trial: Bennett el al. Anesth Analgesia. 2018 Mar;126(3):947-955. doi: 10.1213/ANE.0000000000002432.]

  2. Time to effective bag-mask ventilation

    Time frame: 2 minutes

    How long it takes for clinician to achieve effective ventilation during 2-minute trial. [This is time from when the mask first touches the face of the manikin until all three icons are green. This is measured by a Research Assistant using a stop-watch. Again, "Effective ventilation" is defined as having all three of the following conditions simultaneously and is indicated to the RA by all three icons on the AIR device turning green: (1) Correct rate (30-60 breaths per minute); (2) Absence of significant leak (compliance >4.0mL/cm H2O); and (3) Absence of a significant airway blockage (defined as compliance <0.10 mL/cm H2O or resistance >90 cmH2O/L/s). These measurements are determined by sensors in the AIR device for both arms of the study. Their accuracy for manikin use was determined to be 100% in the following trial: Bennett et al. Anesth Analgesia 2018 Mar;126(3):947-955. doi: 10.1213/ANE.0000000000002432.]

  3. Clinician's correct assessment of mannequin airway issues

    Time frame: 2 minutes

    Ability of clinician to correctly assess three different manikin airway issues: normal, air leak, or obstruction. [Manikin "conditions" are blinded to the participants and were either normal (as supplied by the manufacturer, had a "leak" insert (introduced compliance >4mL/cm H2O), or had an "obstruction" insert (introduced a reduction in compliance <0.10mL/cm H2O and resistance >90 cm H2O/L/s. As per the parameters in the following paper: Bennett et al. Anesth Analgesia. 2018 Mar;126(3):947-955. doi: 10.1213/ANE.0000000000002432.]

Sponsors and collaborators

Lead sponsor

Massachusetts General Hospital

Other

Registry information

Official study title

Augmented Infant Resuscitator (AIR): Transitioning a Novel Behavior Change Innovation to Drive Newborn Ventilation Skills Enhancement

Acronym: AIR

Important dates

Study start
2016
Primary completion
2016
Study completion
2016
First posted
Mar 29, 2021
Registry last updated
Mar 29, 2021

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.