Adult Accuracy Study of the Enlite 3 Glucose Sensor

The purpose of this study is to demonstrate the performance of the Enlite 3 Sensor over 168 hours (7 days) when inserted in the abdomen and used with the Guardian Mobile App and 640G Pump in subjects aged 14-75 years with type 1 or type 2 diabetes.

Completed

Looking for future studies?

Notify Me

Key information

Age range

14 year–75 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

AMCR Institute, Escondido, California, United States

Loading trial locations.

About this study

This study is a multi-center, randomized, in-clinic trial that aims to demonstrate the performance of the Enlite 3 Sensor over 168 hours (7 days) when inserted in the abdomen and used with the Guardian Mobile App and 640G Pump in subjects aged 14-75 years with type 1 or type 2 diabetes for at least one year.

Subjects are randomized into either Group A or Group B. The groups assigned will determine when the subject will be participating in the in-clinic YSI frequent sample testing (FST). For example on Day 1: Group A begins FST 30 minutes after Enlite 3 Sensor Insertion. Group B begins FST 14 hours after Enlite 3 Sensor Insertion.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Subject is 14-75 years of age at time of screening
  • A clinical diagnosis of type 1 or type 2 diabetes for a minimum of 12 months duration as determined via medical record or source documentation by an individual qualified to make a medical diagnosis
  • Adequate venous access as assessed by investigator or appropriate staff
  • Subjects participating in the high and low glucose challenges must have an established insulin carbohydrate ratio(s) and insulin sensitivity ratio. (The term "established" refers to a ratio that has been previously defined and tested prior to screening visit). Subjects without established ratios may participate under observation only.

Exclusion criteria

  • Subject will not tolerate tape adhesive in the area of Enlite 3 Sensor placement as assessed by qualified individual
  • Subject has any unresolved adverse skin condition in the area of the Enlite 3 Sensor or device placement (e.g. psoriasis, rash, Staphylococcus infection)
  • Subject is actively participating in an investigational study (drug or device) in the last 2 weeks
  • Subject is female and has a positive pregnancy screening test
  • Females of child bearing age and who are sexually active should be excluded if they are not using a form of contraception deemed reliable by the investigator
  • Subject is female and plans to become pregnant during the course of the study
  • Subject has had a hypoglycemic seizure within the past 6 months
  • Subject has had hypoglycemia resulting in loss of consciousness within the past 6 months prior to screening visit
  • Subject has had an episode of diabetic ketoacidosis (DKA) within the past 6 months prior to screening visit
  • Subject has a history of a seizure disorder
  • Subject has central nervous system or cardiac disorder resulting in syncope
  • Subject has a history of myocardial infarction, unstable angina, coronary artery bypass surgery, coronary artery stenting, transient ischemic attack (TIA), cerebrovascular accident (CVA), angina, congestive heart failure, ventricular rhythm disturbances or thromboembolic disease
  • Subject has a hematocrit (Hct) lower than the normal reference range
  • Subject has a history of adrenal insufficiency

Treatment and study plan

Enlite 3

Device

Use of Enlite 3 Sensor over 168 hours (7 days) when inserted in the abdomen & arm used with the Guardian Mobile App and 640G Pump in subjects aged 14-75 years who have had a diagnosis of type 1 or type 2 diabetes for at least one year.

Guardian Mobile App

Device

640G Insulin Pump

Device

Primary outcomes

  1. Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD)

    Time frame: 7 Days

    Enlite 3 Sensor accuracy using two real time devices: 1) 640G Pump and 2) Guardian Mobile with the minimum calibration requirements (every 12 hours after the second calibration) will be evaluated. Enlite 3 Sensor values will be compared to YSI plasma glucose values, which is considered as the gold standard, during the frequent sample testing days (Days 1, 3 and 7). MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) * 100). Note that results from multiple testing days will be pooled together for reporting purpose.

Secondary outcomes

  1. Retrospective Re-Analysis (MARD With 1 Additional Calibration)

    Time frame: 7 Days

    Retrospective re-analysis to simulate 640G Pump and Guardian Mobile 1-minute raw data collected by GST3C Transmitters and GST4C Transmitter: Enlite 3 Sensor accuracy with 3-4 calibrations throughout the day (derived from re-analysis of Enlite 3 Sensor data using actual fingerstick values). Enlite 3 Sensor values will be compared to YSI plasma glucose values, which is considered as the gold standard, during the frequent sample testing days (Days 1, 3 and 7). MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) * 100). Note that results from multiple testing days will be pooled together for reporting purpose.

  2. Retrospective Analysis (MARD for the GSR With Minimum and 1 Additional Calibration)

    Time frame: 7 Days

    Retrospective analysis using one GSR: minimum and 3-4 calibrations will be evaluated. Enlite 3 Sensor values will be compared to YSI plasma glucose values during YSI frequent sample testing. Enlite 3 Sensor values will be compared to YSI plasma glucose values, which is considered as the gold standard, during the frequent sample testing days (Days 1, 3 and 7). MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) * 100). Note that results from multiple testing days will be pooled together for reporting purpose.

Sponsors and collaborators

Lead sponsor

Medtronic MiniMed, Inc.

Industry

Registry information

Official study title

A Performance Evaluation of the Enlite® 3 Sensor to Support a Full 168 Hours (7 Days) of Use

Acronym: E3

Important dates

Study start
2014
Primary completion
2015
Study completion
2015
First posted
Sep 23, 2014
Registry last updated
Nov 2, 2016

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.