NCT Number: NCT07166341

Accuracy Verification of the INVOS PM7100

Prospective, observational, pre-market, verification study to confirm accuracy of the INVOS 7100 system in compliance with ISO 80601-2-85:2021 standard

Completed

Looking for future studies?

Notify Me

Key information

Conditions

Age range

18 year–46 year

Sex eligibility

All sexes

Study type

Observational

Primary location

Duke University Medical Center

Durham, North Carolina, 27710, United States

About this study

Verify that the INVOS PM7100 system in conjunction with the INVOS Adult Sensors meets the requirements for tissue saturation accuracy.

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Healthy subjects between the ages of ≥18 to <47 years
  • Completion of a health screening for medical history by a licensed physician, nurse practitioner, or physician assistant
  • Subject is a non-smoker or has not smoked within 2 days prior to the study
  • Minimum weight 40kg
  • BMI within range 18.0-- 35.0
  • Reference tissue hemoglobin oxygen saturation over the range 50% to 85%

Exclusion criteria

  • Compromised circulation, injury, or physical malformation of fingers, hands, or forehead/skull or other sensor sites which would limit the ability to test sites needed for the study (Note: certain malformations may still allow subjects to participate if the condition is noted and would not affect the sites utilized)
  • Prior or known allergies to lidocaine (or similar pharmacologic agents, e.g., Novocain) [self-reported]
  • Prior known severe allergies to medical grade adhesive/tape (Band-Aid) [self-reported]
  • Subjects taking certain medication(s) beyond birth control may be excluded from participating at the PI discretion [self-reported]
  • Subject participation will be at the PI's discretion if they are currently participating in, or if they have recently participated in (discontinued within 30 days prior to the hypoxia procedure for this study) in an investigational drug, device, or biologic study [self-reported]
  • Negative Allen's test for radial and ulnar patency
  • Has made a whole blood donation or has had at least 450 ml of blood drawn within 8 weeks prior to the study procedure [self-reported]
  • Individuals of childbearing potential who are pregnant, who are trying to get pregnant, or who have a urine test positive for pregnancy on the day of the study and is unwilling to use effective birth control between the time of screening and study procedure, or is breastfeeding.
  • Has anemia [lab values specific for gender]
  • Has a history of sickle cell trait or thalassemia [self-reported]
  • Has an abnormal hemoglobin electrophoresis test [lab measurement]
  • Has a positive urine cotinine test or urine drug screen or oral ethanol test [Point of Care (POC) testing]
  • Has a room air saturation less than 95% by pulse oximetry [measurement by PI or delegate]
  • Has a clinically significant abnormal ECG [assessment by PI or delegate]
  • Has a clinically significant abnormal pulmonary function test via spirometry [assessment by PI or delegate]
  • Has a COHb greater than 3%, or MetHb greater than 2% [measured by venous blood sample co-oximetry]
  • Has tHb < 10 g/dl as assessed by co-oximetry during the procedure

Treatment and study plan

observational

Device

The study objectives require placement of a jugular venous bulb catheter and radial arterial line placement with ultrasound guidance. Two INVOS™ sensors (PMSENS71-A) will be placed bilaterally on the subject's forehead and two on the calf. The INVOS™ device will be evaluated for a targeted range of 70-100% SpO2 levels. Hypoxic mixtures of gas are delivered, and data are collected at approximately (~) 5-minute intervals during steady periods of increasing and decreasing oxygen concentration.

Primary outcomes

  1. Accuracy in terms of the root mean square error

    Time frame: Performance evaluation is expected take approximately 2.5 hours

    Performance assessed by comparing INVOS™ rSO2 to the calculated global field saturation (fSO2) value obtained from simultaneous arterial and jugular venous blood samples during normocapnia.

Secondary outcomes

  1. Trend accuracy

    Time frame: Evaluation is expected take approximately 2.5 hours

    • Cerebral site: Complete system in conjunction with an adult sensor (model PMSENS71-A) shall have an rSO2 trending accuracy over the declared range of 50% to 85% compared to reference hemoglobin tissue oxygen saturation.
    • Somatic site: System shall be able to calculate rSO2 on somatic sites.
  2. Trend accuracy in a hypercapneic state

    Time frame: Evaluation is expected take approximately 2.5 hours

    Device performance will be evaluated during hypercapnia

Sponsors and collaborators

Lead sponsor

Medtronic - MITG

Industry

Registry information

Official study title

Accuracy Performance Verification of the INVOS™ PM7100 System, Comparing rSO2 Values to Reference Blood Oxygen Saturation Measurements in Adult Human Volunteers

Important dates

Study start
2025
Primary completion
2025
Study completion
2026
First posted
Sep 10, 2025
Registry last updated
May 14, 2026

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.