NCT Number: NCT04380389
Performance testing of pulse oximeters. The protocol involves brief stable arterial oxygen desaturation in healthy volunteers and sampling arterial blood when a stable level of hypoxia has been attained. The blood sample is analyzed for oxygen saturation with a gold standard bench CO-oximeter. This protocol is aligned with the latest ISO and FDA guidance documents for pulse oximeter testing.
Looking for future studies?
Notify Me18 year–49 year
All sexes
Interventional
Not applicable
San Francisco, California, 94133, United States
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Brief stable arterial oxygen desaturation in healthy volunteers and sampling arterial blood when a stable level of hypoxia has been attained.
Time frame: 1 hour
Root Mean Squared Error between the functional blood oxygen saturation (SpO2) provided by the device under test and the arterial blood oxygen measured with the CO-oximeter.
Withings
Industry
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06015204
Brachial Plexus Block, Oximetry
Daegu, South Korea
View Trial DetailsNCT03534271
Oximetry
Chicago, Illinois, United States
View Trial DetailsNCT04964609
Oximetry
San Francisco, California, United States
View Trial DetailsNCT04562584
Oximetry
San Francisco, California, United States
View Trial Details