Completed

NCT Number: NCT03901092

A Reader Study to Assess Accuracy and Reliability of Flortaucipir F 18 Positron Emission Tomography (PET) Scan Interpretation

This study will evaluate the performance of physician readers trained to read flortaucipir-PET (positron emission tomography) scans.

Completed

Looking for future studies?

Notify Me

Key information

Sex eligibility

All sexes

Study type

Interventional

Phase

Phase 3

Primary location

American College of Radiology

Philadelphia, Pennsylvania, 19103, United States

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Scan Reader Criteria (5 total):

  • Board-certified in radiology or nuclear medicine
  • Professional experience interpreting PET scans
  • Naive to study protocol
  • No previous training or exposure to Avid Flortaucipir F 18 read methodology

Scan Criteria:

  • Previous enrollment in Study A05 confirmatory cohort (NCT02016560), or A16 (NCT02516046)

Scan Study Population Criteria for FR01 (A05 confirmatory cohort):

  • Cognitively-impaired
  • mild cognitive impairment (MCI) or dementia with suspected neurodegenerative cause
  • mini-mental status exam (MMSE) score of 20-27, inclusive

Scan Study Population Criteria for FR01 (Study A16):

  • Subjects at end of life (less than or equal to 6 months)
  • Imaged with flortaucipir F18 and came to autopsy

Treatment and study plan

flortaucipir F 18

Drug

No study drug will be administered.

Other names: 18F-AV-1451, [F-18]T807, LY3191748

Primary outcomes

  1. Primary Objective 1 Analysis 1: Diagnostic Performance of Individual Readers (NFT Score)

    Time frame: baseline scan

    Sensitivity and specificity of 5 independent readers' interpretations of ante-mortem 18F-AV-1451 PET imaging for detection of a pattern of 18F-AV-1451 neocortical uptake that corresponds to neurofibrillary tangles (NFT) Score of B3 (Hyman et al., 2012; Montine et al., 2012). NFT B scores range from B0 (no NFTs in the brain) to B3 (widespread NFTs in the brain). Sensitivity and specificity are percentages that can range from 0 to 100%. The hypothesis tested was that, of the 5 independent imaging physicians, at least 3 will have the lower bounds of 2-sided 95% CIs ≥50%, for both sensitivity and specificity.

  2. Primary Objective 1 Analysis 2: Diagnostic Performance of Individual Readers (NIA-AA Autopsy Diagnosis)

    Time frame: baseline scan

    Sensitivity and specificity of 5 independent readers' interpretations of ante-mortem 18F-AV-1451 PET imaging for detection of a pattern of 18F-AV-1451 neocortical uptake that corresponds to high levels of AD neuropathologic change (High ADNC) as defined by National Institute on Aging-Alzheimer's Association (NIA-AA) criteria (see Hyman et al. 2012). ADNC categories are None, Low, Intermediate and High, with High indicating the most severe level of AD-related pathology changes in the brain. The hypothesis tested was that, of the 5 independent imaging physicians, at least 3 will have the lower bounds of 2-sided 95% CIs ≥50%, for both sensitivity and specificity.

  3. Primary Objective 2: Inter-reader Reliability of Reader Interpretation of Flortaucipir-PET Imaging

    Time frame: baseline scan

    As measured by Fleiss' Kappa across all scans read. Fleiss' kappa is a statistical measure for assessing the reliability of agreement between a fixed number of raters when assigning categorical ratings to a number of items or classifying items. Fleiss' kappa can range from 0 to 1 with 1 indicating perfect agreement between the readers. Scan results binarized as positive AD pattern versus negative AD pattern. Results are displayed as percentage of agreement for individual reader pairs, computed as follows: number of images for which reader had the same interpretation divided by the total number of images evaluated, multiplied by 100%.

Secondary outcomes

  1. Secondary Objective 1 Analysis 1: Diagnostic Performance of τAD++ Flortaucipir PET Images to Detect B3 NFTs

    Time frame: baseline scan

    Flortaucipir F 18 PET imaging will be classified by each reader as either neocortical uptake not consistent with AD (τAD-), neocortical uptake consistent with AD (τAD+), or neocortical uptake consistent with AD with uptake beyond the temporal/occipital regions (τAD++). NFT scoring is according to Hyman, et al 2012. Truth positive is a NFT B3 score. Truth negative is NFT <B3 score.

  2. Secondary Objective 1 Analysis 2: Diagnostic Performance of τAD++ Flortaucipir PET Images to Detect High ADNC

    Time frame: baseline scan

    Sensitivity and specificity of 5 independent readers' interpretations of ante-mortem 18F-AV-1451 PET imaging for detection of a pattern of 18F-AV-1451 neocortical uptake that corresponds to high levels of AD neuropathologic change (High ADNC) as defined by National Institute on Aging-Alzheimer's Association (NIA-AA) criteria (Hyman et al. 2012). Truth positive is a High ADNC score. Truth negative is No/Low/Intermediate ADNC score.

  3. Secondary Objective 2: Inter-reader Reliability of Reader Interpretation of τAD++ Flortaucipir PET Images

    Time frame: baseline scan

    Overall reader agreement as measured by Fleiss' Kappa statistic. Scan results binarized as τAD++ versus τAD+/τAD-. Results are displayed as percentage of agreement for individual reader pairs, computed as follows: number of images for which reader had the same interpretation divided by the total number of images evaluated, multiplied by 100%.

  4. Secondary Objective 3: Inter-reader Reliability of Flortaucipir PET Scan Interpretation in the Population of Intended Use

    Time frame: baseline scan

    Reader agreement measured by Fleiss' Kappa across scans from non-autopsy cases from Study A05. Scan results binarized as positive for AD pattern versus negative for AD pattern. Results are displayed as percentage of agreement for individual reader pairs, computed as follows: number of images for which reader had the same interpretation divided by the total number of images evaluated, multiplied by 100%.

  5. Secondary Objective 4: Intra-reader Reliability of Reader Interpretation of Flortaucipir-PET Imaging

    Time frame: baseline scan

    Cohen's Kappa will be calculated for each of 5 readers to assess the intra-reader reliability of flortaucipir F 18 PET scan visual interpretation. The statistic takes into account the possibility of the agreement occurring by chance. Cohen's kappa values range from 0 to 1 with 1 representing perfect agreement. Results are displayed as percentage of agreement within a reader, computed as follows: number of images for which reader had the same interpretation at initial and second read divided by the total number of images evaluated twice by a reader, multiplied by 100%.

Sponsors and collaborators

Lead sponsor

Avid Radiopharmaceuticals

Industry

Registry information

Official study title

A Reader Study to Assess Accuracy and Reliability of Flortaucipir F 18 PET Scan Interpretation

Important dates

Study start
2019
Primary completion
2019
Study completion
2019
First posted
Apr 3, 2019
Registry last updated
Sep 9, 2020