Skip to main content
OpenTrials
Not Yet Recruiting

NCT Number: NCT06729255

A Prospective Cohort Study to Evaluate the Safety, Efficacy, and Usability of the EpiZact Device

The purpose of the study is to learn more about a new device, the EpiZact Epidural Device, that may be helpful for reducing the likelihood of wet tap when a labor epidural is placed. When epidurals are placed correctly the needle stops just in front of the space in the body that contains cerebrospinal fluid. When the epidural is advanced a short distance further, leakage of cerebrospinal fluid can occur. This is known as a wet tap. A wet tap can result in a severe headache and at times other complications. The goal of this clinical trial is to learn if the use of the EpiZact device prevents wet taps in pregnant women receiving epidurals for relief of labor pain. The main questions this study aims to answer are:

1. What the rate of wet tap with the EpiZact device? 2. What is the rate of failed epidural placement with the EpiZact device?

All patients in the study will receive an epidural with the EpiZact device. The investigators will compare the results with patients not in the study that receive an epidural without the EpiZact device.

Not Yet Recruiting

Trial opening soon.

Get Notified

Key information

Age range

15 year–50 year

Sex eligibility

Female

Study type

Interventional

Phase

Phase 4

Primary location

About this study

The purpose of the study is to learn if use of the EpiZact device decreases the number of wet taps that occur when epidurals are placed. The EpiZact device has a technology that automatically stops the epidural needle when it enters the epidural space. In theory, this should decrease the likelihood of a wet tap. The standard epidural to treat labor pain does not have this capability. The study device may cause fewer wet taps than the standard epidural. All study participants will receive an epidural to treat labor pain with the EpiZact device. Results will be compared with patients receiving epidurals around the same time without the EpiZact device.

If a participant's screening questions show that they can take part in the study, and the participant chooses to be a part of the study, medical records from OHSU or other facilities may be reviewed. Information reviewed may include treatment history, medications, surgeries, allergies, scans, and lab results. The participant will then receive an epidural with the EpiZact device. After the epidural is placed, the participant will be asked a few questions to determine how well the epidural is working and if a complication occurred. The questions are expected to take less than 5 minutes to answer. Participants will also be called 1 week after receiving an epidural to determine if a complication occurred. This call is expected to last less than 5 minutes.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Pregnant women requesting labor epidural analgesia
  • Women pregnant with a single baby (not twins or triplets)

Exclusion criteria

  • Women requesting not to have a labor epidural
  • Height less than 150 cm
  • A bleeding disorder, nerve problem, or skin infection that contraindicates receiving an epidural
  • A women receiving a combined spinal epidural (CSE) or a dural puncture epidural (DPE)
  • A history of spine surgery
  • Allergy to local anesthetics (numbing medicine) or skin cleaning solutions
  • Decisionally impaired individuals

Treatment and study plan

EpiZact-assisted epidural placement

Device

The standard epidural kit used with the EpiZact device to place a midline epidural using the landmark technique between the L2 and L5 vertebral bodies. The patient's back will be sterilized with a 2% chlorhexidine/70% isopropyl alcohol applicator. The epidural will be placed with a 17-gauge 3.5-inch Tuohy needle. Once the Tuohy needle is appreciated to be engaged in ligament the EpiZact device will be filled with sterile and attached to the Tuohy via a Luer connector. The green button on the EpiZact device will then be pressed to prime the device. The EpiZact device and Tuohy needle will be advanced as a unit in continuous fashion or with intermittent advancement. Upon sensing a loss of resistance, the EpiZact device will provide a visual and tactile signal, halting needle advancement and signaling that the epidural catheter should be threaded through the Tuohy needle.

Primary outcomes

  1. Proportion of Participants with an Unintentional Dural Puncture

    Time frame: 1 week

    Unintentional dural puncture will be a composite outcome diagnosed by any of the following: (1) free flow of cerebrospinal fluid from the Tuohy needle at any time during the epidural technique, (2) requirement of blood patch within one week of neuraxial procedure, or (3) postural headache that worsens with upright posture and improves with supine position diagnosed within one week of neuraxial placement.

Secondary outcomes

  1. Proportion of Participants with Failed Labor Epidural Placement

    Time frame: 2 hours

    Failed epidural placement will be defined as the inability to place a catheter in the epidural place or the need to replace the epidural within 2 hours of insertion. Decision to replace the epidural will be according to standard practice on labor and delivery at OHSU and will typically involve assessment of loss of cold to ice sensation in the lower thoracic and lumbar dermatomes, 50% decrease in labor contraction pain, and new lower extremity paresthesia.

Study contacts

Contact information is provided by the study sponsor or research team.

Brandon M Togioka, MD

CONTACT

[email protected]

503-494-4572

Sarah L Feller

CONTACT

[email protected]

503-494-6233

Sponsors and collaborators

Lead sponsor

Oregon Health and Science University

Other

Collaborators

  • GuideStar Medical Devices

Registry information

Official study title

A Single Academic Center, Prospective Cohort With Retrospective Loss-of-resistance Comparator, Proof of Concept Study to Evaluate the Safety, Efficacy, and Usability of the EpiZact Device for Parturients Undergoing Epidural Labor Analgesia

Important dates

Study start
2025
Primary completion
2026
Study completion
2026
First posted
Dec 11, 2024
Registry last updated
Feb 14, 2025

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.