A Multicenter, Open Label, Prospective, Non-randomized Study of the InCraft® Stent Graft System in Subjects With Abdominal Aortic Aneurysms (INSPIRATION)
The purpose of this study is to evaluate the safety and effectiveness of InCraft® in subjects with abdominal aortic aneurysms requiring endovascular repair.
Looking for future studies?
Notify MeKey information
Conditions
Age range
20 year and older
Sex eligibility
All sexes
Study type
Interventional
Phase
Not applicable
Primary location
Takao Ohki, MD, Minato-ku, Tokyo, Japan
About this study
In the US, abdominal aortic aneurysms (AAA) are found in 4-8% of older men and 0.5-1.5% of older women, resulting in 30,000-40,000 elective procedures and 1,400 peri-operative deaths. The most significant complication of AAA is an aneurysm sac rupture from which more than 15,000 patients die annually and is the 15th leading cause of death in elderly between 60 to 85 years of age . In Japan, it is estimated that approximately 14,000 endovascular aneurysm repair (EVAR) and open surgical repair cases were performed in the year 2010 .
Abdominal aortic aneurysms can be treated three ways: (1) Medical management; (2) Open surgical repair; and (3) Endovascular aneurysm repair (EVAR). EVAR has emerged as an alternative treatment of AAA for most patients. It is less invasive than open repair and carries lower rates of early mortality and morbidity . It has also extended treatment options to patients who cannot undergo conventional surgical procedures due to a high operative risk. As EVAR technology evolves, it allows treatment of AAA with increasing complexity of the aortic neck and access vessels.
The InCraft® AAA Stent Graft System is designed for endovascular repair of infrarenal AAAs with complex aortic anatomies. This stent-graft system utilizes nitinol stent and polyester graft technology in an ultra-low profile delivery system, which assists the physician in deploying the device in a controlled, consistent, and precise manner within the aortic neck and iliac arteries. By isolating the aneurysmal sac, the system provides an alternative blood flow path to relieve pressure on the arterial vessel walls and minimize aneurysm growth and the potential for aneurysm sac rupture.
Who can participate
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
- Subject must meet ALL of the following inclusion criteria to be enrolled in the study:
- Male or Female age 20 years or older;
- Proximal aortic neck is 17-31mm in diameter;
- Supra-renal aorta, at 20mm above the anticipated landing location, is smaller than the nominal diameter of the aortic bifurcate prosthesis to be used;
- Infra-renal aortic neck is ≥10mm in length with supra-renal and infra-renal angulations ≤60°;
- Subject has at least one of the following:
- AAA size > 5.0 cm;
- Increase of the AAA diameter of > 0.5 cm over the last 6 months;
- Abdominal treatment length (lowest renal artery origin to the aortic bifurcation) ≥ 9.4cm;
- Aortic bifurcation >18mm in diameter;
- Iliac landing zone ≥15mm in length;
- Iliac landing zone 7-22mm in diameter;
- Minimum access vessel size of ≥ 5mm;
- Minimum overall AAA treatment length (from lowest renal artery to distal landing zone) of 128 mm;
- Women of child bearing potential must be non-pregnant, non-lactating, and not planning to become pregnant during the course of the trial; and have a negative urine or serum pregnancy test within 7 days prior to index procedure;
- Provide written informed consent and as applicable written HIPAA authorization (For US sites only) prior to initiation of study procedures;
- Willing to comply with the specified follow-up evaluation schedule.
Exclusion criteria
- Subjects will be excluded if ANY of the following exclusion criteria apply:
- Vascular anatomy in which the placement of the stent-graft will cause occlusion of both internal iliac arteries or necessitates surgical occlusion of both internal iliac arteries;
- Subject has one of the following:
- Aneurysm sac rupture or leaking abdominal aortic aneurysm;
- Mycotic, dissecting, or inflammatory abdominal aortic aneurysm;
- Clinically significant acute vascular injury due to trauma;
- Significant aortic or iliac mural thrombus, plaque or calcification that would compromise fixation and seal of the device;
- A conical aortic neck defined as >3mm distal increase over a 10mm length in the planned seal zone;
- Thoracic aortic aneurysm ≥45mm;
- Any aortic dissection;
- Morbid obesity (BMI >40.0 kg.m2) or other clinical conditions that limit required imaging studies or visualization of the aorta;
- Renal insufficiency (Creatinine > 2.0mg/dL) or subject on renal dialysis;
- Known allergy or intolerance to nickel titanium (nitinol) , Polyethylene terephthalate (PET), or polytetrafluoroethylene (PTFE);
- Known contraindication to undergoing angiography or anticoagulation (e.g. contrast allergies which cannot be treated);
- Connective tissue disorder (such as Marfan's Syndrome or Ehlers-Danlos Syndrome);
- Coagulopathy, bleeding disorder, or other hypercoagulable state;
- Organ transplant recipient or subject requiring systemic immunosuppressant therapy;
- Cerebral vascular accident (CVA), MI, or intracranial bleeding within 3 months prior to the procedure;
- Active infection or chronic systemic illness at the time of index procedure that may interfere with the study objectives;
- Major surgical procedure within 1 month prior to the index procedure or pre-planned within 1 month afterwards;
- Co-existing condition with a life expectancy of less than 2 years at time of procedure;
- Current or planned participation in any other investigational drug or medical device clinical study that has not completed primary endpoint(s) evaluation;
- Existing AAA surgical graft and/or a AAA stent-graft system;
- Other medical, social, or psychological issues that in the opinion of the investigator preclude the subjects from receiving this treatment, and the procedures and evaluations pre- and posttreatment.
Treatment and study plan
Primary outcomes
-
Number of Participants With Incidence of MAE (Major Adverse Events) Based Upon a Composite MAE Rate
Time frame: 30 days post-procedure
defined as a composite of:
- Death
- Stroke
- Myocardial Infarction
- New Onset Renal Failure (requiring dialysis)
- Respiratory Failure (requires mechanical ventilation)
- Paralysis/paraparesis
- Bowel ischemia (requiring surgical intervention)
- Procedural blood loss ≥1,000 cc
-
Number of Participants With Successful Aneurysm Treatment
Time frame: Up to 1 year post-procedure
Successful aneurysm treatment which is a composite endpoint of the following:
- Technical Success at the conclusion of the index procedure, defined as successful insertion of the delivery system through the vasculature and successful deployment of the device at the intended location. The endovascular graft must be patent, with absence of types I or III endoleaks or aneurysm sac rupture, at the time of procedure completion as confirmed by angiography or other imaging modality;
- Absence of post-operative aneurysm enlargement (growth > 5 mm) or stent graft migration (> 10mm) compared to the one month size measurement at anytime up to 1-year;
- Absence of post-operative conversion to open surgery, sac rupture, endoleak Type I / III, or graft occlusion (including unilateral or bilateral limb occlusion) at any time up to 1-year.
Secondary outcomes
-
Procedure-related Complications
Time frame: Through 1 month, 180-days, 360-days and annually to 5-years post-procedure
Adverse events that are a result of the procedure itself.
-
Aneurysm-related Mortality
Time frame: At 30-days, 180-days, 360-days and annually to 5-years post-procedure
-
Incidence of Secondary Interventions
Time frame: through 5 years
The incidence of secondary interventions within 1 year post-procedure, needed to prevent the occurrence of a significant event. Significant event being defined as: aneurysm enlargement (growth > 5 mm), stent graft migration (> 10mm) compared to the one month size, endoleak type I / III, graft occlusion, sac rupture.
-
Major Adverse Events (MAEs) and Individual Components of the MAEs
Time frame: At 180-days, 360-days and annually to 5-years post-procedure
-
Device-related Events
Time frame: At 1 month, 6 months, 1 year and annually to 5-years post-procedure
Endoleak(s), Aneurysm sac rupture, Fracture(s, Delivery System Malfunction, Device Malfunction, Stent Graft Migration - evidence of proximal or distal movement of the stent graft >10mm relative to fixed anatomic landmarks compared with 1 month, Graft Occlusion (including unilateral or bilateral limb occlusion, Conversion to open surgery, and Aneurysm Enlargement - defined as an increase in maximum aneurysm cross sectional diameter > 5mm compared to the 1-month measurement.
-
Technical Success Confirmed by CT or Other Imaging Modality
Time frame: At 30-days
The study evaluated technical success at 30-days as one of the secondary effectiveness endpoints. The definition of technical success is defined as patency of the endovascular graft, with the absence of Type I or III endoleaks or aneurysm sac rupture, up to 30-days post-procedure completion as confirmed by CT or other imaging modality.
-
Length of Hospital Stay (Days) Post Index Procedure
Time frame: up to 17 days
-
Length of Intensive Care Unit (ICU) Stay (Hours) Post Index Procedure
Time frame: up to 48 hours
-
Length of the Index Procedure (Minutes)
Time frame: 1 day
Other outcomes
-
Pain Post-Operatively
Time frame: Baseline, 1 month, 6 months and 1 year post-procedure
Outcome measured by the SF36v2 Health Status Survey where all items are scored so that a high score defines a more favorable health state. Each item is scored on a 0 to 100 range so that the lowest and highest possible scores are 0 and 100, respectively.
-
Physical Functioning Post-operatively
Time frame: Screening, 1 month, 6 months and 1 year post-procedure
Physical Functioning (PF) Outcome measured by the SF36v2 Health Status Survey where all items are scored so that a high score defines a more favorable health state. Each item is scored on a 0 to 100 range so that the lowest and highest possible scores are 0 and 100, respectively.
Sponsors and collaborators
Lead sponsor
Cordis US Corp.
Industry
Collaborators
- Carelon Research
- Quintiles, Inc.
Registry information
Acronym: INSPIRATION
Important dates
- Study start
- 2012
- Primary completion
- 2014
- Study completion
- 2018
- First posted
- Aug 14, 2012
- Registry last updated
- Jul 3, 2024
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Related clinical trials
Published trials that share one or more normalized conditions with this study.
Follow-up After Endovascular Repair of Abdominal Aortic Aneurysm
NCT00488943
Abdominal Aortic Aneurysm, Aneurysm
San Francisco, California, United States
View Trial DetailsNellix Registry Study: EVAS-Global
NCT02996396
Abdominal Aortic Aneurysm, Aneurysm
Augsburg, Germany
View Trial Details3D Holographic Guidance, Navigation, and Control for Endovascular Aortic Repair
NCT06008613
Abdominal Aortic Aneurysm, Aneurysm
Cleveland, Ohio, United States
View Trial DetailsResults Wide and Conical Neck EVAR Procedures
NCT06972095
Abdominal Aortic Aneurysm, Aneurysm
Arnhem, Netherlands
View Trial Details