A Comparative Clinical Trial of the TensorTip, a Novel Non Invasive Device for Measurement Blood Parameters

1. Aim:

The aim of the clinical trial is to study the accuracy of the TensorTip device compared with registered (FDA-approved) invasive and non-invasive devices. 2. Hypothesis:

Human physiological biomarkers may be measured from the color distribution of the internal or external (skin) tissue. The technology of the TensorTip finger-mounted device is based on the color distribution of the peripheral blood tissue, which enables the measurement of certain biomarkers and vital signs under consideration.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Atlantic Health - Morristown Memorial Hospital- Cardiovascular Medicine

Morristown, New Jersey, 07962-1905, United States

About this study

The TensorTip (previously named Soft Touch) is a novel non invasive Finger tip device that. The Tensor Tip is a non-invasive device which was designed and aimed to measure a wide range of physiological and bio-parameters such as blood glucose, hemoglobin and hematocrit, blood PH, oxygen saturation SpO2, blood carbon dioxide, blood pressure, peripheral pulse and more based on the temporarily color distribution of the tissue under consideration. The Tensor Tip is based on real time color image sensor, real time photographing the fingertip tissue. A color image sensor of the type used in the Tensor Tip enables a wide range of information in the spectral, resolution, dynamic range and time domains enabling further investigation of the blood chromatic changes as a function body physiology. The concept behind this investigation reflects the idea that a change in human physiology condition would temporarily change the blood pigmentation.

The present study is aimed to assess the accuracy of the novel device in measuring all the above mentioned parameters by recording AS IS the change of the fingertip color tissue while simultaneous recording the standard invasive measured parameters on daily or periodic routine in the hospitalized patients.

This vast data recordings (tissue and references) enable the Tensor tip to analyze a particular bio parameter. The tensor tip is calibrated to a certain bio parameter based on the vast data set collected.

The bio parameter under consideration are: Arterial Systolic and Diastolic, MAP, PA systolic, PA Diastolic, CO, Heart rate, Stroke volume, WBC, RBC, Hgb, HCT ,Iron, PLT, PH, PCO2, PO2, HCO3, SaO2 , Na, K, Cl, CO2, Bun, Cr, Gluc, HGB (A1C),TBiLi, LDL, HDL.

This study hopes to demonstrate the connection between blood color pigmentation and certain bio parameters for simplifying monitoring procedure and particular home monitoring.

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Male or female, age > 18
  • Patients with cardiovascular disease and/or with diabetes mellitus type I or II and/or with COPD.
  • Patients receiving coronary artery bypass with or without valve repair.
  • Able to sign an informed consent.

Exclusion criteria

  • Refusal to sign an informed consent and to participate in the study.
  • Participant plans to enroll in another other clinical trial of an investigational agent while participating in this study.
  • Below the age of 18.

Treatment and study plan

Non invasive peripheral blood monitoring

Device

non invasive monitoring

Primary outcomes

  1. Comparison of the TensorTip Accuracy Against Hospital Periodical Readings.

    Time frame: one year

    To validate the parameters with similar measures obtained with standard invasive techniques in hospitalized patients.

    Methodology:

    Eligible real time color signal obtained by the TensorTip shall be recorded simultaneously during the monitoring performed.

    An algorithm shall be designed according to the blood color distribution to each parameter and a final test shall be recorded signals.

    Statistical analysis for each parameter shall perform on the entire eligible recorded signals.

    Determination of Accuracy:

    Err = √(1/N ∑N (Ref(k)-NewDevice(k)) ^2 ) (k=1)

    For each parameter a satisfactory result is considered when Err satisfies the industry requirement.

    Results viewing Each parameter comparative study shall be presented on a (X,Y) plan versus a regression line where X - represents the results measured by TensorTip and Y - the reference results.

Sponsors and collaborators

Lead sponsor

Cnoga Medical Ltd.

Industry

Registry information

Official study title

A Comparative Clinical Trial to Assess the Accuracy of the TensorTip, a Novel Non Invasive Device for the Measurement of Physiological and Blood Chemistry Parameters

Important dates

Study start
2010
Primary completion
2013
Study completion
2014
First posted
Sep 8, 2010
Registry last updated
Jan 31, 2019

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.