Daxor Corporation
Oak Ridge, Tennessee, 37830, United States
NCT Number: NCT06227819
The proposed study is a prospective, observational, investigator and patient blinded study comparing the results of blood volume measurement using the FDA-cleared Daxor BVA-100 device to the Daxor BVA-200 device. Data from this study may be pooled with data from studies with similar design conducted at other sites, for the purposes of an FDA medical device submission.
Looking for future studies?
Notify Me18 year and older
All sexes
Observational
Oak Ridge, Tennessee, 37830, United States
Measurements of human blood volume will be made via the indicator dilution method, using gamma counter readings of the concentration of I-131 in a subject's blood sample taken after radio-iodinated Human Serum Albumin (HSA) has been injected into the bloodstream via venous access. Each subject receiving an injection will have five separate blood samples taken, which will be analyzed separately. Each of the five measurements will be conducted in duplicate with the BVA-100 instrument and the BVA-200 instrument. Sufficient blood will be withdrawn at each phlebotomy to permit such duplicate measurement.
Subjects will be enrolled who have been prescribed a blood volume procedure using the BVA-100, as a result either of their ordinary diagnostic care or of their participation in another study that uses blood volume measurement. Subjects will be adults under treatment for conditions normally associated with the need for volume assessment, such as Heart Failure, or scheduled to receive BVA-100 measurement as part of another study. An effort will be made to ensure a range of ages and a balance of the sexes.
Subjects will be enrolled and consented sequentially. The University of Tennessee Medical Center site will document any patient who is not enrolled in the study sequentially and the reason why or who declines enrollment. Demographics of patients not enrolled will be collected to permit identification of any trends in non-enrollment. We do not anticipate any sequential patient not being offered the opportunity to participate in order to assure that there is no bias in selection of patients ("all-comers").
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Blood samples collected for measurement with the BVA-100 will also be tested with the BVA-200. The BVA-200 uses whole blood collection cartridges which allow for a precise amount of blood to be measured without the need to first centrifuge and then pipette plasma aliquots. A small amount of whole blood from each sample will be set aside for BVA-200 use, while the remainder of the blood sample is used for BVA-100 measurement. The BVA-200 device was developed with the support of the Department of Defense through Phase I (W81XWH19C0048) and Phase II (W81XWH20C0025) SBIR contracts.
Time frame: Approximately 6 months.
Comparison of a sufficient number of indicator-dilution volume measurements between BVA-100 and BVA-200 devices to establish substantial equivalence.
Time frame: Approximately 6 months.
Comparison of a sufficient number of full 5-point time-zero blood volume measurements between BVA-100 and BVA-200 devices, to establish substantial equivalence, including measurement of:
Time frame: Approximately 6 months.
Validation of filling consistency of at least 100 whole blood quantitative collection cartridges, in at least 40 sets of 5 samples from the same patient
Time frame: Approximately 6 months.
Validation of the ability of personnel who are not laboratorians to understand the test procedures and perform the measurement after receiving training materials (which may be a combination of written instructions, video instructions, and individual instruction) in the use of the instrument. Usability is evaluated through a User Validation Survey consisting of yes or no questions, as well as questions answered on a scale of 1 (strongly disagree) to 5 (strongly agree), where 5 is the most favorable outcome.
Daxor Corporation
Industry
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT02742597
Alzheimer Disease, Alzheimer's Disease
London, Ontario, Canada
View Trial DetailsNCT02789800
Alzheimer Disease, Alzheimer's Disease
Chicoutimi, Quebec, Canada
View Trial DetailsNCT03847753
Acquired Immunodeficiency Syndrome, Allergy
View Trial DetailsNCT03892265
Alcohol Drinking, Alcohol Use, Unspecified
Port-au-Prince, West, Haiti
View Trial Details