A Clinical Trial to Verify the Accuracy of Seers Technology's Pulse Oximeter (mobiCARE +Pulse, MP100W) in Healthy Volunteers

It is important to evaluate the accuracy of the pulse oximeter, which is commonly used in clinical practice. The US FDA only allows the use of devices that meet the ISO 80601-2-61:2017 standard in the United States. According to the ISO 80601-2-61:2017 standard, the accuracy of the oxygen saturation (SpO2) of the pulse oximeter should have an error range of less than 4.0% compared to the actual arterial blood oxygen saturation (SaO2) in the range of 70-100%. To prove compliance with this requirement, it is necessary to conduct a clinical trial that induces hypoxemia in healthy adults and compares and evaluates the arterial blood oxygen saturation values of the pulse oximeter's SpO2 and the arterial blood oxygen saturation values of the carbon monoxide-oxygen meter (CO-oximeter).

Completed

Looking for future studies?

Notify Me

Key information

Conditions

Age range

20 year–50 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Seers Technology Co., Ltd.

Pyeongtaek-si, Gyeonggi-do, 17707, South Korea

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • healthy adults between 20 and 50 years
  • volunteers who have plasma proportion of carboxyhemoglobin 3%
  • volunteers who have plasma proportion of methemoglobin 2%
  • volunteers who have total plasma hemoglobin concentration over 10 g/dL

Exclusion criteria

  • History of respiratory diseases
  • History of cardiovascular disease
  • Smokers (including ex-smokers)
  • Pregnant
  • History of fainting
  • History of diabetes
  • Obesity (BMI> 30 kg/m2)
  • Allen's test results, those who do not recover the color of their fingers and palms within 10 seconds
  • History of allergic reaction to lidocaine, a local anesthetic

Treatment and study plan

mobiCARE +Pulse (MP100W)

Device

Volunteers were monitored using end-tidal carbon dioxide partial pressure and fraction of inspired oxygen. Each volunteer was placed in a semi-Fowler's position and connected to a breathing circuit to administer the nitrogen-air-carbon dioxide mixtures. A nose clip was applied to prevent breathing of room air. For frequent blood sampling, an arterial cannula was placed in the radial artery of each volunteer. Pulse oximeter probes were simultaneously attached to each volunteer's fingers. Each volunteer was exposed to various levels of induced hypoxia from 70~100% of SaO2. Each plateau of oxygen saturation was maintained for at least 30 s until stabilization, after which 1 ml of arterial blood was drawn into a heparinized syringe. The study period consisted of two rounds of hypoxia, and the volunteers were maintained on room air between each round. SaO2 measurements using a CO-oximeter were used as a reference for the SpO2 accuracy.

Primary outcomes

  1. Accuracy of SpO2

    Time frame: 2 day

    The root-mean-square of the difference between the observed pulse oximetry value (SpO2) and the carbon monoxide-oxygenometer arterial blood reference value (SRi) within the range of 70-100% of arterial oxygen saturation

Sponsors and collaborators

Lead sponsor

Seers Technology Co., Ltd.

Industry

Registry information

Important dates

Study start
2023
Primary completion
2023
Study completion
2023
First posted
Jun 8, 2023
Registry last updated
Jun 8, 2023

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.