Electrocardiogram recording
DeviceECG recording with investigational device
NCT Number: NCT06622265
The purpose of the study is to evaluate the performance of ECG classification algorithms and their ability to classify heart rhythms into multiple categories of rhythms and heart rates.
Looking for future studies?
Notify Me22 year and older
All sexes
Observational
HOPE Research Institute, Phoenix, Arizona, United States
Atrial fibrillation (Afib) is the most common cardiac arrhythmia in the United States, affecting up to one in four individuals across the lifespan, and is associated with substantial morbidity and mortality. Fifteen percent of strokes in the United States are attributable to Afib, and nearly 20% of these occur in individuals with no prior Afib history. Because Afib is often paroxysmal, one-time screening is unlikely to capture those at risk. Thus, there has been an increasing interest in leveraging monitoring for Afib via wearable devices, which provide a novel method to monitor Afib and determine the burden of Afib in the general population.
The primary objective of this study is to assess the sensitivity and specificity of the investigation device (WHOOP ECG Feature) for the detection of Afib as compared to a gold-standard reference (12-Lead ECG Holter Device).
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
-Subjects with an implantable pacemaker device or implantable cardioverter- defibrillator device.
ECG recording with investigational device
Time frame: 1 Day
Specificity is defined as the True Negatives/(True Negatives + False Positives). A true normal sinus rhythm reading will have a 12-lead ECG reading consistent with normal sinus rhythm. A true negative is defined as individuals who have normal sinus rhythm reported on their ECGs from both the WHOOP ECG Feature and the 12-lead ECG Holter Device. A false positive is defined as individuals who receive an ECG with atrial fibrillation by the WHOOP ECG Feature but their 12-lead ECG has an atrial fibrillation determined by adjudication.
Time frame: 1 Day
Sensitivity is defined as the True Positives/(True Positives + False Negatives). A true atrial fibrillation reading will have an 12-lead ECG reading consistent with atrial fibrillation. A true positive is defined as individuals who have atrial fibrillation reported on their WHOOP ECGs and on the 12-lead ECG from the Holter Device. A false negative is defined as individuals who receive an ECG with atrial fibrillation with the 12-lead ECG as determined by adjudication but their WHOOP ECG Feature determination is Normal Sinus Rhythm.
Whoop Inc.
Industry
Multi-phase, Prospective, Multicenter, Non-interventional, Pre-market, Non-randomized, Open-label Study to Evaluate the Clinical Performance of the WHOOP ECG Feature
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06676072
Arrhythmias, Cardiac, Atrial Arrhythmia
Scottsdale, Arizona, United States
View Trial DetailsNCT07014423
Arrhythmias, Cardiac, Atrial Fibrillation
Groningen, Netherlands
View Trial DetailsNCT07178691
Anxiety, Anxiety Disorders
Aalborg, Denmark
View Trial DetailsNCT06706518
Ablation of Atrial Fibrillation, Arrhythmias, Cardiac
Madrid, Spain
View Trial Details