Vital USA is dedicated to developing and applying innovative electronic medical solutions that improve patient care in multiple clinical settings. Vital USA is a manufacturer of a multi-parameter patient monitor that is intended to be used in a variety of medical settings including hospitals, medical offices and patients' homes.
Looking for future studies?
Notify Me18 year–52 year
All sexes
Interventional
Not applicable
Louisville, Colorado, 80027, United States
Purpose of Clinical Investigation Respiratory Rate is an important physiological measurement in the healthcare setting. End Tidal Carbon Dioxide (EtCO2) is considered to be the highest level of safety and accuracy in Respiratory Rate monitoring by the American Society of Anesthesiology and the American Association of Respiratory Care.
The purpose of this study was to conduct a Respiratory Rate accuracy validation comparing the Vital Detect to an FDA cleared End Tidal Carbon Dioxide monitor Reference Standard (GE Datex-Ohmeda).
Respiratory Rate is an important physiological measurement in the healthcare setting. End Tidal Carbon Dioxide (EtCO2) is considered to be the highest level of safety and accuracy in Respiratory Rate monitoring by the American Society of Anesthesiology and the American Association of Respiratory Care.
The purpose of this study was to conduct a Respiratory Rate accuracy validation comparing the Vital Detect to an FDA cleared End Tidal Carbon Dioxide monitor Reference Standard (GE Datex-Ohmeda).
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Subject is considered as being morbidly obese (defined as BMI >39.5)
The Vital Detect is a noninvasive device designed for spot-checking physiological parameters, such as Noninvasive Blood Pressure, Oxygen Saturation, Pulse Rate, Body Temperature, and Respiratory Rate. This device is a finger-cuff technology which is applicable for use at-home or in healthcare facilities in/on individuals 18 years of age and older. It may be used by the individual themselves or an operator. The Vital Detect is not intended for continuous monitoring or for use with high frequency surgical equipment. It is non transit-operable. The Vital USA Vital Detect is composed of a sensor head mounted on a base connected to a phone or computer via Vital Detect App which is free to download.
Other names: VitalDetect
Time frame: 20 Seconds
The primary objective of this study was to compare the accuracy of device under test for the measurement of respiratory rate. The respiratory rate was measured simultaneously with EtCO2 (Reference) and the Vital USA Vital Detect monitor (Device Under Test).The endpoint of interest was accuracy as measured by the Accuracy root-mean-square (ARMS) difference between the Device Under Test (DUT) and the reference respiratory rate system (Ref) for all stable respiratory periods.
Vital USA, Inc.
Industry
Final Test Report for Vital USA Respiratory Rate Validation In Adults
Acronym: RR
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT05068622
Digestive System Diseases, Gastric Outlet Obstruction
Amiens, France
View Trial DetailsNCT06521346
Cardiac Disease, Cardiovascular Diseases
Damanhūr, Egypt
View Trial DetailsNCT06502002
Acute Lung Injury, Lung Diseases
Boston, Massachusetts, United States
View Trial DetailsNCT06149494
COPD, COPD Exacerbation Acute
London, United Kingdom
View Trial Details