Visual and Refractive Outcomes of The ClarVista HARMONI™ Modular Intraocular Lens System (HMIOL)

The purpose of this study was to evaluate visual and refractive outcomes with the use of the HARMONI™ Modular Intraocular Lens (HMIOL) System implantation, assembly, and optic exchange in subjects undergoing cataract surgery.

Completed

Looking for future studies?

Notify Me

Key information

Age range

22 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

About this study

All subjects were implanted with the HARMONI™ Modular Intraocular Lens (HMIOL) System in the right eye, left eye, or both eyes based on the best interest of the subject and the clinical judgment the Investigator (Day 0, primary implantation) (Cohort 1). At Month 1 post primary implantation, subjects were given the option to undergo an optic exchange procedure. Subjects who underwent the optic exchange (Cohort 2) (Day 0, optic exchange) were followed for an additional month. Subjects who did not undergo an optic exchange were exited from the study, except at 1 site, which followed subjects up to Month 12.

Alcon Research, LLC, acquired ClarVista Medical in 2017. This study was designed and conducted by ClarVista Medical, Inc. The study results were collected, analyzed, and provided by ClarVista Medical, Inc. to Alcon Research, LLC.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Key Inclusion Criteria:

  • Able to understand and sign an approved Informed Consent Form (ICF);
  • Willing and able to return for scheduled treatment and follow-up examinations;
  • Planned removal of cataracts (cortical, nuclear, subcapsular, or a combination) by manual phacoemulsification cataract extraction;

Key Exclusion Criteria:

  • Any ocular conditions which could affect the stability of the IOL in either eye;
  • Traumatic or congenital cataract;
  • Pregnancy or planned pregnancy during the study period;
  • Medications that may confound the outcome or increase risk to the subject;
  • Any clinical finding or intraocular complication during primary cataract surgery or Implantation of the lens system likely to increase complications or risk to the subject;
  • Other protocol-specified exclusion criteria may apply.

Treatment and study plan

Harmoni Modular Intraocular Lens

Device

Two-component system consisting of a base and a separate optic that allows for an intraoperative exchange of the optic only without direct manipulation of the capsular bag.

Other names: HMIOL

Optic Exchange

Procedure

Removal of one optic and replacement with another for the purpose of improving refractive outcomes

Primary outcomes

  1. Percentage of Eyes With Uncorrected Distance Visual Acuity (UCDVA) by Visit - Cohort 1

    Time frame: Day 1 post primary implantation, Week 1 post primary implantation, Month 1 post primary implantation, Month 3 post primary implantation, Month 6 post primary implantation, Month 12 post primary implantation

    Visual acuity of the eye was tested while reading charts at 20-foot equivalent distance from the participant with an optical infinity adjustment of +0.25 diopter (D). UCDVA was recorded in Snellen, with 20/20 considered to be 'normal' vision. A visual acuity of 20/40 means the participant is able to read a certain size letter 20 feet away that a person with 'normal' vision would be able to read from 40 feet away. A larger denominator, therefore, indicates a lower visual acuity. No formal statistical hypothesis testing was planned.

  2. Percentage of Eyes With Uncorrected Distance Visual Acuity (UCDVA) by Visit - Cohort 2

    Time frame: Day 0 pre optic exchange, Day 1 post optic exchange, Week 1 post optic exchange, Month 1 post optic exchange

    Visual acuity of the eye was tested while reading charts at 20-foot equivalent distance from the participant with an optical infinity adjustment of +0.25 diopter (D). UCDVA was recorded in Snellen, with 20/20 considered to be 'normal' vision. A visual acuity of 20/40 means the participant is able to read a certain size letter 20 feet away that a person with 'normal' vision would be able to read from 40 feet away. A larger denominator, therefore, indicates a lower visual acuity. No formal statistical hypothesis testing was planned. The optic exchange occurred 1 month following primary implantation with HMIOL.

  3. Percentage of Eyes With Best Corrected Distance Visual Acuity (BCDVA) by Visit - Cohort 1

    Time frame: Baseline (Day -90 to Day 0 preoperative), Week 1 post primary implantation, Month 1 post primary implantation, Month 3 post primary implantation, Month 6 post primary implantation, Month 12 post primary implantation

    Visual acuity of the eye was tested while reading charts at 20-foot equivalent distance from the participant with subject manifest refraction in place. BCDVA was recorded in Snellen, with 20/20 considered to be 'normal' vision. A visual acuity of 20/40 means the participant is able to read a certain size letter 20 feet away that a person with 'normal' vision would be able to read from 40 feet away. A larger denominator, therefore, indicates a lower visual acuity. No formal statistical hypothesis testing was planned.

  4. Percentage of Eyes With Best Corrected Distance Visual Acuity (BCDVA) by Visit - Cohort 2

    Time frame: Baseline (Day 0 pre optic exchange), Week 1 post optic exchange, Month 1 post optic exchange

    Visual acuity of the eye was tested while reading charts at 20-foot equivalent distance from the participant with subject manifest refraction in place. BCDVA was recorded in Snellen, with 20/20 considered to be 'normal' vision. A visual acuity of 20/40 means the participant is able to read a certain size letter 20 feet away that a person with 'normal' vision would be able to read from 40 feet away. A larger denominator, therefore, indicates a lower visual acuity. No formal statistical hypothesis testing was planned.

  5. Mean Manifest Refraction Spherical Equivalent (MRSE) by Visit - Cohort 1

    Time frame: Baseline (Day -90 to Day 0 preoperative), Week 1 post primary implantation, Month 1 post primary implantation, Month 3 post primary implantation, Month 6 post primary implantation, Month 12 post primary implantation

    A manifest refraction (manual vision test) was conducted using letter charts and a phoropter. The subject was manually refracted to his/her best correction. The manifest refraction spherical equivalent (MRSE) was measured in diopters (D) and calculated as follows: sphere + 1/2 cylinder. A less negative value indicates a more accurate IOL power calculation. No formal statistical hypothesis testing was planned.

  6. Mean Manifest Refraction Spherical Equivalent (MRSE) by Visit - Cohort 2

    Time frame: Baseline (Day 0 pre optic exchange), Week 1 post optic exchange, Month 1 post optic exchange

    A manifest refraction (manual vision test) was conducted using letter charts and a phoropter. The subject was manually refracted to his/her best correction. The manifest refraction spherical equivalent (MRSE) was measured in diopters (D) and calculated as follows: sphere + 1/2 cylinder. A less negative value indicates a more accurate IOL power calculation. No formal statistical hypothesis testing was planned.

  7. Number of Surgeon Responses to Primary Surgery Question: How Would You Rate the Surgical Difficulty of the Base Implantation? - All HMIOL Cohort

    Time frame: Day 0 primary implantation

    The surgeon's response was collected on a study-specific questionnaire after each surgery. No formal statistical hypothesis testing was planned.

  8. Number of Surgeon Responses to Primary Surgery Question: How Would You Rate the Surgical Difficulty of the Optic Implantation and Assembly? - All HMIOL Cohort

    Time frame: Day 0 primary implantation

    The surgeon's response was collected on a study-specific questionnaire after each surgery. No formal statistical hypothesis testing was planned.

  9. Number of Surgeon Responses to Optic Exchange Surgery Question: How Would You Rate the Surgical Difficulty of the Optic Implantation and Assembly (Exchange)? - Cohort 2

    Time frame: Day 0 post-exchange

    The surgeon's response was collected on a study-specific questionnaire after each surgery. No formal statistical hypothesis testing was planned.

  10. Number of Surgeon Responses to Optic Exchange Surgery Question: How Would You Rate the Surgical Difficulty of the Optic Disassembly? - Cohort 2

    Time frame: Day 0 post-exchange

    The surgeon's response was collected on a study-specific questionnaire after each surgery. No formal statistical hypothesis testing was planned.

  11. Number of Surgeon Responses to Optic Exchange Surgery Question: How Would You Rate the Surgical Difficulty of the Optic Explantation? - Cohort 2

    Time frame: Day 0 post-exchange

    The surgeon's response was collected on a study-specific questionnaire after each surgery. No formal statistical hypothesis testing was planned.

  12. Number of Secondary Surgical Interventions (SSI) (Other Than Optic Exchange)

    Time frame: Cohort 1: Up to one month post primary implantation, with one site followed up to 12 months. Cohort 2: Up to one month post exchange, for a total of up to 2 months.

    A secondary surgical intervention (SSI) was defined as a surgical procedure that occurred after primary implantation and was conducted for the purpose of resolving residual refractive error (RRE) and optimizing visual outcomes, as determined by the investigator. No formal statistical hypothesis testing was planned.

  13. Number of Device Deficiencies

    Time frame: Cohort 1: Up to one month post primary implantation, with one site followed up to 12 months. Cohort 2: Up to one month post exchange, for a total of up to 2 months.

    A device deficiency was defined as a failure of the device to meet its performance specifications or expectations, or otherwise not perform as intended. This could include either a malfunction or damage to the device or any part thereof, regardless of the source of malfunction or damage, including user error, and regardless of the presence of injury (or lack thereof) to subject, user, or bystander. No formal statistical hypothesis testing was planned.

Sponsors and collaborators

Lead sponsor

ClarVista Medical

Industry

Registry information

Official study title

Evaluation of Visual and Refractive Outcomes of The ClarVista HARMONI™ Modular Intraocular Lens System

Important dates

Study start
2016
Primary completion
2017
Study completion
2017
First posted
Sep 24, 2018
Registry last updated