McMaster University
Hamilton, Ontario, L8S 4L8, Canada
NCT Number: NCT05886153
The VERDICT-2 verification study will examine the accuracy of the Cloud DX Vitaliti Continuous Vital Signs Monitor (Model: CVSM-1A) in an ambulatory context with healthy participants. The objective is to determine the accuracy of continuous non-invasive vital signs metrics including respiration, pulse rate, oxyhemoglobin saturation (SpO2), core temperature, heart rate, and cNIBP against standard comparator devices guided by consensus standards.
Looking for future studies?
Notify Me18 year–100 year
All sexes
Interventional
Not applicable
Hamilton, Ontario, L8S 4L8, Canada
Effective postoperative care is faced with several challenges- patients experience major complications following surgery, which often require re-hospitalization or utilization of healthcare services (i.e., emergency room visits) once discharged into the community. New remote automated monitoring (RAM) technologies that enable data integration, synthesis, and real-time alerts to clinicians have the potential to transform postoperative care, similar to how intraoperative care has evolved.
The Cloud DX Vitaliti Continuous Vital Signs Monitor (Model: CVSM-1A) is a continuous vital sign monitor intended to measure vital metrics including systolic and diastolic blood pressures (via ECG/ photoplethysmography [PPG]), pulse (via PPG), respiration rate (via impedance respiration waveform), core body temperature (via infrared sensor), and blood oxygen saturation levels (via PPG) of an adult individual using non-invasive techniques.
The VERDICT-2 verification study will examine the accuracy of the Vitaliti CVSM in an ambulatory context with healthy participants.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Measurements recorded by the Vitaliti CVSM device will be compared to the recordings from various standardized comparator devices.
Time frame: 44 Minutes
Determining the accuracy of the impedance respiration waveform on the Vitaliti device to measure respiration rate, compared to an appropriate standardized comparator device. The Vitaliti CVSM will be evaluated against performance criteria determined to be clinically relevant, in alignment with predicate devices, and as defined by the manufacturer, since there are no internationally recognized consensus standards.
Time frame: 27 Minutes
Determining the accuracy of the ECG/photoplethysmography (PPG) on the Vitaliti device to measure pulse rate, compared to an appropriate standardized comparator device. Detailed analysis will be conducted to verify accuracy of the Vitaliti device against performance criteria as defined in IEC 60601-2-27:2011/ (R2016).
Time frame: 27 Minutes
Determining the accuracy of the electrocardiogram (ECG)/photoplethysmography (PPG) on the Vitaliti device to measure blood oxygen saturation, compared to an appropriate standardized comparator device. A detailed analysis will be performed to compare the Vitaliti device against performance criteria as defined in ISO 80601-2-61:2017, MOD.
Time frame: 12 Minutes
Determining the accuracy of the infrared sensor on the Vitaliti device to measure core body temperature, compared to an appropriate standardized comparator device. A detailed analysis will be conducted to verify accuracy of the Vitaliti device against performance criteria as defined in ISO 80601-2-56:2017/AMD-1:2018.
Time frame: 36 Minutes
Determining the accuracy of the electrocardiogram (ECG) on the Vitaliti device to measure heart rate, compared to an appropriate standardized comparator device. A detailed analysis will be conducted to verify accuracy of the Vitaliti device against performance criteria as defined in IEC 60601-2-27:2011/ (R2016).
Time frame: 30 Minutes
Determining the accuracy of the electrocardiogram (ECG)/photoplethysmography (PPG) on the Vitaliti device to measure blood pressure, compared to an appropriate standardized comparator device. A detailed analysis will be conducted to verify accuracy against performance criteria as defined in ISO 81060-2: 2018/AMD-1:2020.
Hamilton Health Sciences Corporation
Other
Verification Protocol for the Cloud DX Vitaliti Continuous Vital Signs Monitor (Model CVSM-1A) for Clinical Use and Self Measurement in Ambulatory Settings
Acronym: VERDICT-2
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07274332
Autonomic Nervous System, Blood Pressure
Türkeli, Sinop, Turkey (Türkiye)
View Trial DetailsNCT06627920
Blood Pressure, Heart Rate
Valladolid, Castille and León, Spain
View Trial DetailsNCT05022264
Blood Pressure, Body Temperature
Ulricehamn, Sweden
View Trial DetailsNCT07497659
Blood Pressure, Heart Rate
Uberaba, Minas Gerais, Brazil
View Trial Details