Auto-Titrating EPAP
DeviceIn the Auto-titrating EPAP mode, the device automatically adjusted EPAP to the minimum level able to abolish tidal expiratory flow limitation.
NCT Number: NCT04725500
This study aimed to validate a novel automatic non-invasive ventilation (NIV) mode that continuously adjusts expiratory positive airway pressure (EPAP) to the lowest value that abolishes tidal expiratory flow limitation (EFL). The investigator conducted a prospective, non-randomized, study on stable chronic obstructive pulmonary disease (COPD) patients that may or may not be treated currently with NIV. Patients were studied in a sleep lab on a single night with the auto - titrating EPAP that adjusts to abolish tidal EFL. The primary endpoint was to evaluate the behavior of the EPAP during the night. Additionally, a sub-group of patients used the device at home for a 2 week period. EPAP behavior was assessed during this 2 week period.
Looking for future studies?
Notify Me40 year–80 year
All sexes
Interventional
Not applicable
Pulmonary Rehabilitation Associates, Youngstown, Ohio, United States
Tidal Expiratory Flow Limitation (EFLT) is the inability to increase expiratory flow despite increasing effort. Especially in patients with mild to severe COPD, EFL can present challenges when receiving mechanical ventilation. To overcome EFLT expiratory positive airway pressure (EPAP) is applied; however, a single level or fixed EPAP may not overcome the airflow obstruction. The current study was undertaken to explore the variability of EFLT determined using Forced Oscillation Technique (FOT) to dynamically measure lung reactance (∆Xrs) and to evaluate the ability of automatically adjusted EPAP (PEEPopt) to over come EFLT overnight and over a two week period.
In this prospective non-randomized trial, an unreleased noninvasive ventilator set in an S/T mode applied continuous oscillations (5 Hz, 1 cmH2O amplitude, 2 cmH20 peak to peak). Response to the oscillations was analyzed to calculate ∆Xrs and EPAP was adjusted automatically between 4 and 20 cmH2O. A fixed pressure support of 6 cmH2O was delivered. ∆Xrs was measured sitting and supine. Participants with evidence of EFLT (∆Xrs > 2.8) were asked to undergo overnight polysomnography (PSG) and then a sub-group of patients used the ventilator at home for two weeks. EFLT within and between participants was analyzed. EFLT behavior during PSG, the response of the device to dynamically abolish EFLT, as well as the impact of this therapy on sleep quality was studied. Objective ventilator adherence data were used to determine usage after the 2-week in-home use. Additionally, experienced NIV participants qualitatively rated therapy comfort compared to their current treatment.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
In the Auto-titrating EPAP mode, the device automatically adjusted EPAP to the minimum level able to abolish tidal expiratory flow limitation.
Time frame: Approximately 8 hours
The expiratory positive airway pressure (EPAP) is the parameter that is automatically adjusted. The ventilator automatically titrates EPAP to the minimum value able to abolish expiratory flow limitation.
Time frame: Approximately 8 hours
The difference between mean inspiratory and expiratory reactance is measured using the forced oscillation technique. The difference between mean inspiratory and expiratory reactance is an index of tidal expiratory flow limitation. Values > 2.8 cmH2O*s/L are indicative of tidal expiratory flow limitation.
Time frame: Approximately 8 hours nightly use over 14 days
The expiratory positive airway pressure (EPAP) is the parameter that is automatically adjusted. The ventilator automatically titrates EPAP to the minimum value able to abolish expiratory flow limitation.
Time frame: 2 weeks
The difference between mean inspiratory and expiatory reactance is measured using the forced oscillation technique. The difference between mean inspiratory and expiratory reactance is an index of tidal expiratory flow limitation. Values > 2.8 cmH2O*s/L are indicative of tidal expiratory flow limitation.
Time frame: Approximately 8 hours
Sleep quality is a key outcome during overnight studies. Nocturnal non-invasive ventilation may affect sleep quality in two ways: 1) it may improve it by reducing sleep-related respiratory problems; 2) it may worsen it if the ventilation mode causes discomfort to the patient. Sleep quality was evaluated by full laboratory polysomnography. Participants after there overnight answered a question of comfort and they answered as 1- being very uncomfortable 2- uncomfortable 3- neutral 4- comfortable 5- very comfortable.
Time frame: Nightly use for 14 days
Device usage every day, for 14 days based on data downloaded from the device.
Time frame: End of 2 weeks device take home study
Prior NIV users rated their comfort with the Forced Oscillation Technique (FOT) therapy over their prior non-invasive ventilation (NIV) therapy. Participants were asked the question "how comfortable did the pressure therapy feel compared to your current device" and they answered the questions on a scale of 0 to 5.
5-Very Comfortable 4-Comfortable 3-Neutral 2-Uncomfortable
1-Very Uncomfortable 0-Not applicable
Philips Clinical & Medical Affairs Global
Industry
Engineering Evaluation of the Vector NIV Device in Chronic Obstructive Pulmonary Disease (COPD) Patients
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT03927820
Asthma, Bronchial Diseases
Nashville, Tennessee, United States
View Trial DetailsNCT06614959
Bronchial Diseases, Bronchitis
Montreal, Quebec, Canada
View Trial DetailsNCT04754490
COPD, Chronic Disease
Abercynon, Wales, United Kingdom
View Trial DetailsNCT06336252
Arthritis, Basal Ganglia Diseases
Copenhagen, Denmark
View Trial Details