Vdrive Lasso navigation
DeviceRemote robotic Lasso navigation
NCT Number: NCT01656772
This study will evaluate the use of a robotic device that is remotely controlled to maneuver a circular mapping catheter in the left atrium during Atrial Fibrillation (AF) ablation procedures.
Looking for future studies?
Notify Me18 year–80 year
All sexes
Interventional
Phase 3
ZNA Campus Middelheim Lindendreef 1, Antwerp, Antwerpen, Belgium
Key to the success of an electrophysiology procedure is the ability to accurately map, sense and pace the heart. Electrophysiology mapping catheters are available in a variety of shapes and electrode configurations. The majority of these catheters are manipulated manually by the physician. The Vdrive simplifies these manipulations by allowing an operator to mechanically advance, retract, rotate, deflect and maneuver a circular mapping catheter. The circular mapping catheter is not altered in structure, function or indication in the same manner as the predicate device (Cardiodrive Catheter Advancement System).
This is a prospective, randomized, non-blinded, controlled study comparing manual navigation of a circular mapping catheter with mechanical navigation by the Vdrive. A total of 120 patients are expected to complete the study: 80 patients in the Vdrive navigation arm and 40 patients in the manual navigation arm. Patients will be enrolled at a minimum of one site in the European Union where the product is currently marketed and the United States, where the Vdrive is considered to be investigational.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Remote robotic Lasso navigation
Manually maneuver a Lasso catheter
Time frame: Peri-procedural
Success was navigating and sensing by obtaining an EGM at the pre-specified targeted PVs. The null hypothesis was to be tested at the α = 0.05 significance level using a test statistic Z based on the Farrington and Manning likelihood score statistic with adjustment to take into account the correlation between multiple observations (PVs) on the same subject. The null hypothesis was to be rejected if Z > 1.645 or, equivalently, if the corresponding p-value was less than 0.05.
Time frame: 7 days Follow-up
The Vdrive will allow physicians to manipulate compatible circular mapping catheters while maintaining a safety profile that is not inferior to manual circular catheter navigation. The one-sided null hypothesis was to be tested at the α = 0.05 significance level using a standard unpooled asymptotically normal test statistic. The null hypothesis was to be rejected if Z < -1.7046 or, equivalently, if the corresponding p-value was less than 0.05.
Stereotaxis
Industry
Acronym: VERSATILE
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT05820295
Arrhythmias, Cardiac, Atrial Fibrillation
New York, United States
View Trial DetailsNCT05993897
Arrhythmias, Cardiac, Atrial Fibrillation
Aswān, Aswan Governorate, Egypt
View Trial DetailsNCT05357690
Arrhythmias, Cardiac, Atrial Fibrillation
Rochester, Minnesota, United States
View Trial DetailsNCT05164718
Arrhythmias, Cardiac, Atrial Fibrillation
Drammen, Norway
View Trial Details