ECG measurement
Diagnostic TestSimultaneous recording of a single-lead ECG signal with the Withings BPM Core and ECG signal for the reference 12-lead ECG
NCT Number: NCT04464499
The proposed clinical study aims to validate the diagnostic performance, compared to a reference ECG, of the electrocardiographic function of the BPM Core developed by Withings for the automatic identification of atrial fibrillation (AF).
This study is active but is not currently recruiting participants.
Notify Me18 year–100 year
All sexes
Observational
ICPC Floréale, Bagnolet, France
BPM Core is a blood-pressure monitor with an build-in single-channel electrocardiogram similar to a lead I.
Primary objective:
The primary objective is to demonstrate that Withings BPM Core has acceptable sensitivity and specificity in the automatic identification of AF and SR in comparison with a gold-standard diagnosis from a 12-lead ECG by board certified cardiologists.
The algorithm classifies the device signals into four categories: normal sinus rhythm (RSN), AF, arrhythmia other than AF, or non-interpretable. The latter category includes signals of insufficient quality to be interpreted with confidence.
This classification will be compared to the diagnosis made by three independent, blinded cardiologists based on a 12-lead ECG. The final diagnosis retained will be decided by majority with adjudication by a consensus of 3 to 6 cardiologists. The sensitivity and specificity of the device will then be estimated.
Secondary objectives:
The first secondary objective is to assess the quality of tracings of the BPM Core armband thanks to a qualitative and quantitative comparison between BPM Core tracings and reference ECG ones, based on clinical evaluation criteria.
Firstly, three independent cardiologists who have not performed or read the reference ECG and do not know the patient's clinical files will make a diagnosis from the recordings from the BPM Core. There will be a majority vote to select the diagnosis. The cases where the three readers reached three different diagnosis will be reviewed by a panel of 3 to 6 board certified cardiologists who will decide by consensus. Sensitivity and specificity will be calculated relative to the reference ECG.
Secondly, quantitative analyses will be performed to assess the quality of ECG tracings provided by Withings BPM Core and its software, as follows:
The sponsor will make sure that the same cardiologist will not be asked to annotate both a Withings BPM Core signal and the corresponding lead I signal from the 12-channel ECG device. The sponsor will then centralize the annotations of Withings BPM Core tracings and of the ECG reference lead I signals and perform agreement calculations on synchronous pairs.
The second secondary objective is to verify the safety of use of the device by analyzing adverse effects.
The safety of use of the device will be assessed by the number of adverse effects.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Simultaneous recording of a single-lead ECG signal with the Withings BPM Core and ECG signal for the reference 12-lead ECG
Time frame: 1 year
sensitivity from the 2x2 confusion matrix AF vs 'non-AF'
Time frame: 1 year
sensitivity from the 2x2 confusion matrix NSR vs 'non-NSR'
Time frame: 1 year
sensitivity from the 2x2 confusion matrix AF vs 'non-AF'
Time frame: 1 year
sensitivity from the 2x2 confusion matrix NSR vs 'non-NSR'
Time frame: 1 year
(true positives + true negatives)/(sample size) from the 2x2 confusion matrix
Time frame: 1 year
(true positives + true negatives)/(sample size) from the 2x2 confusion matrix
Time frame: 1 year
(true positives + true negatives)/(sample size) from the 2x2 confusion matrix
Time frame: 1 year
(true positives + true negatives)/(sample size) from the 2x2 confusion matrix
Time frame: 1 year
(true positives + true negatives)/(sample size) from the 2x2 confusion matrix
Time frame: 1 year
(true positives + true negatives)/(sample size) from the 2x2 confusion matrix
Withings
Industry
Acronym: Core AF
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT04922281
Arrhythmias, Cardiac, Atrial Fibrillation
Rochester, New York, United States
View Trial DetailsNCT05452356
Arrhythmias, Cardiac, Atrial Fibrillation
London, Ontario, Canada
View Trial DetailsNCT06564012
Arrhythmias, Cardiac, Atrial Fibrillation
View Trial DetailsNCT05351775
Arrhythmias, Cardiac, Atrial Fibrillation
Turku, Finland
View Trial Details