Zio monitor
DeviceThe Zio monitor is a long-term continuous ambulatory patch ECG monitor which adheres to the patient's left pectoral region.
NCT Number: NCT07205367
This trial assesses the wear performance of the Zio monitor, a long-term continuous ambulatory patch ECG monitor, in a representative population by evaluating (1) the wear duration (up to 30 days) of the device, (2) the percentage of analyzable electrocardiographic time and (3) safety by reviewing all adverse events.
Looking for future studies?
Notify Me18 year–70 year
All sexes
Interventional
Not applicable
iRhythm Technologies - San Francisco, CA, San Francisco, California, United States
This is a prospective, multi-center, single-arm study to validate the design of the Zio monitor/Zio MCT device form factor in extended (21-day) wear when utilizing an updated skin prep process.
The trial is conducted in 2 phases. Phase I includes the standard skin preparation process of 40 strokes with an exfoliator in 4 different directions. Phase II incorporates a modified skin preparation process: only 20 strokes with an exfoliator in 2 directions are used. There are 75 participants in each phase.
Performance objectives include the following: (1) Mean wear duration of 21 days and (2) percent analyzable time of 80%.
Safety measures include the proportion of participants that experience clinically significant skin irritation through 30 days of wear, as well as the reporting of all adverse events.
Additional device measures include the following: (1) % signal artifact (defined as the proportion of ECG signal that is artifact over the wear duration, reported for each study device); (2) % device functionality (defined as the proportion of Zio monitors recording a continuous ECG signal through 21 days, and through 30 days) and (3) observational assessment of participant skin types (assessed by dryness levels, hyperhydrosis status, Fitzpatrick Skin Type and chest hair density).
All analyses will be performed on the population of enrolled participants on whom the Zio monitor application was attempted. The age, sex at birth, and application type (i.e., healthcare provider (HCP) designee vs. self-application) of participants will be monitored throughout the enrollment period to ensure representation across key user groups. Performance within subgroups including (but not limited to) sex at birth, age, and application type will be assessed.
Seventy-five (75) participants in Phase I and 75 participants in Phase II will be enrolled at 2 US iRhythm locations (San Francisco, CA and Deerfield, IL).
There are 3 on-location visits at Baseline (Visit 1/time of patch application), Day 21 (Visit 4) and Day 30 (Visit 5/following 30 days of patch application). There are 2 telephone/video calls at Day 7 (Visit 2) and Day 14 (Visit 3).
Participants who meet all inclusion criteria and no exclusion criteria will be enrolled.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
The Zio monitor is a long-term continuous ambulatory patch ECG monitor which adheres to the patient's left pectoral region.
Time frame: 21 - 30 days after baseline
This outcome measure is met if the mean wear duration is greater than 21 days
Time frame: Up to 30 days (Day 30)
This outcome measure is met if the percent analyzable time is greater than 80% in 90% of subjects.
Time frame: Up to 30 days (Day 30)
This includes the proportion of participants that experience clinically significant skin irritation through 30 days of wear. This will be assessed through the capture and reporting of all adverse events.
Time frame: Up to 30 days (Day 30)
Percent signal artifact is defined as the proportion of ECG signal that is artifact over the wear duration, reported for each study device. Signal artifact is a continuous value representing the percentage of the wear duration that was not interpretable. This variable will be calculated as the inverse of analyzable ECG percent. Signal artifact will be summarized using the population meeting protocol requirements and include means and 95% confidence intervals for these proportions.
Time frame: At 21 days (Day 21) and 30 days (Day 30)
Device functionality is defined as the number of participants with Zio monitors recording a continuous ECG signal through 21 and 30 days. The ability to record a continuous ECG signal is defined as device functionality. Device functionality will be determined by the number of participants with devices adhering to the skin, the number of participants with devices functioning/operating and the number of participants with devices that are recording and ECG signal.
Time frame: Baseline (Day 1)
This includes an assessment of the following skin types: dry, oily and normal.
Time frame: Baseline (Day 1)
This includes an assessment of the following hyperhydrosis status: primary, secondary or none.
Time frame: Baseline (Day 1)
This includes a 6-point scale to assess the following skin types: (1) pale skin; always burns, never tans; (2) fair skin; usually burns, tans minimally; (3) darker white skin; sometimes mild burn, tans uniformly; (4) light brown skin; burns minimally, always tans well; (5) brown skin; very rarely burns; tans very easily; (6) dark brown or black skin; never burns.
Time frame: Baseline (Day 1)
This includes an assessment of the following chest hair density types: smooth/no hair; some hair; moderately hairy; or very hairy in the target patch placement area.
iRhythm Technologies, Inc.
Industry
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT05298553
Arrhythmias, Cardiac, Atrial Fibrillation
Oxford, Oxfordshire, United Kingdom
View Trial DetailsNCT06172699
Ablation, Arrhythmias, Cardiac
San Diego, California, United States
View Trial DetailsNCT02241252
Arrhythmias, Cardiac, Cardiac Arrhythmia
Oklahoma City, Oklahoma, United States
View Trial DetailsNCT05772182
Arrhythmias, Cardiac, Cardiac Arrhythmia
Barcelona, Spain
View Trial Details