NCT Number: NCT04763278

Validation of User Needs of the Point Digit With Partial Hand Amputees

The objective of the device feasibility study will be to validate the user needs of the Point Digit system. This study will be a single group intervention model where one group of 5 partial hand amputees will be asked to perform several tasks. Successful completion of a task results in a fulfilled user need. Failure to complete a task results in an unfulfilled user need.

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

University of Colorado Denver | Anschutz Medical Campus, Aurora, Colorado, United States

Loading trial locations.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Partial hand amputees with absence of index and/or middle fingers and presence of thumb. Additional finger loss is acceptable if all other criterion are met.
  • Fluent in English
  • Individuals aged 18 or greater

Exclusion criteria

  • Significant cognitive deficits as determined upon clinical evaluation
  • Significant neurological deficits as determined upon clinical evaluation
  • Significant physical deficits of the residual limb impacting full participation in the study as determined upon clinical evaluation
  • Uncontrolled pain or phantom pain impacting full participation in the study as determined upon occupational therapist evaluation
  • Serious uncontrolled medical problems as judged by the project therapist.

Treatment and study plan

Point Digit

Device

The Point Digit partial hand prosthetic system consists of 1-4 ratcheting mechanical digits and a mounting bracket

Primary outcomes

  1. VAL-1: System Comfort Validation Test

    Time frame: Day 1 (8 hours)

    User wears Point Digit(s) for a full day (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall be comfortable for a full day's use. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.

  2. VAL-2: Unilateral Function Validation Test

    Time frame: Day 1 (10 minutes)

    User flexes and extends the Point Digit(s) without using contralateral hand (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall allow the user to flex and extend finger without their second hand. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.

  3. VAL-3.1: System Robustness Validation Test With 25 lb Bag

    Time frame: Day 1 (10 minutes)

    User lifts a 25 lb. bag with Point Digit(s) (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall be robust enough to operate in challenging environments. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.

  4. VAL-3.2: Robustness Validation Test With Hammer

    Time frame: Day 1 (10 minutes)

    User grips a hammer with Point Digit(s) and drives a nail into a piece of wood (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall be robust enough to operate in challenging environments. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.

  5. VAL-4: Object Release Validation Test

    Time frame: Day 1 (10 minutes)

    User releases Point Digit(s) while performing an active grasp by pressing the dorsal button (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall be releasable while performing active grasps. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.

  6. VAL-5: Position Function Validation Test

    Time frame: Day 1 (10 minutes)

    User positions Point Digit(s) into each of the locking positions (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall have a smooth ratcheting mechanism. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.

  7. VAL-6: Extension Function Validation Test

    Time frame: Day 1 (10 minutes)

    User activates spring-back mechanism by using (1) contralateral hand, (2) tabletop, and (3) side or thigh (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall have a reliable auto spring-back mechanism. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.

  8. VAL-7: Grip Functions Validation Test

    Time frame: Day 1 (10 minutes)

    User grips a 5 lb. cylindrical smooth object using Point Digit(s) for 10 seconds (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall have high grip surface friction. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.

  9. VAL-8: Fine Motor Function Validation Test

    Time frame: Day 1 (10 minutes)

    User picks up 1 coin from a smooth tabletop (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall have a pronounced fingernail. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.

  10. VAL-9: Overall Function & Robustness Validation Test

    Time frame: Day 1 (8 hours)

    User performs VAL-1 through VAL-8 (Pass/Fail). This validation method was used to confirm the following user need of the Point Digit has been met: Point Digit shall be robustly attached to mounting system. A simple pass/fail result was recorded for each user need for each participant. An overall passing rate above 80% across all user needs and participants is considered a success. Each participant attended a single-day experimental session. A day-use prosthetic socket was fabricated and Point Digits were mounted to the socket. Then, each participant attempted the validation test.

Sponsors and collaborators

Lead sponsor

Point Designs

Industry

Collaborators

  • Eunice Kennedy Shriver National Institute of Child Health and Human Development (NICHD)
  • University of Colorado, Denver

Registry information

Important dates

Study start
2021
Primary completion
2022
Study completion
2022
First posted
Feb 21, 2021
Registry last updated
Sep 27, 2024