Validation of the Dynamometer LABIN

The main objective of this work is to validate the dynamometer LABIN in comparison with the technique of gold-standard reference JAMAR, in a healthy population.

The infringement of the prehension is frequent and impacts on the social and professional quality of life of the patients. There are numerous instruments measuring the strength of prehension. The JAMAR is a mechanical, hydraulic dynamometer, exploring in kilograms the maximal strength of the prehension. It is at present Gold Standard for the measure of the strength of prehension. There is for the moment no electronic known instrument exploring the endurance of the strength of prehension during an effort of tightening. This parameter would indeed be interesting for the exploration in geriatrics and in the neuromuscular pathologies.

This study justifies itself by the fact that the dynamometer LABIN should allow the evaluation of the muscular endurance of the strength of hand as complement of the peak of strength

Completed

Looking for future studies?

Notify Me

Key information

Age range

20 year–60 year

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

CHU Clermont-Ferrand

Clermont-Ferrand, 63003, France

About this study

Measure of the strength of prehension with JAMAR and LABIN according to the following protocol:

  • Position: the subject sits comfortably on a chair without armrests, feet put on the ground, the pond at the bottom of the seat, the hips and the knees in 90 ° of flexion, shoulder in adduction and neutral rotation, the elbow bent in 90 °, forearm in neutral position, the wrist between 0 and 30 ° of extension and between 0 and 15 ° of slope ulnaire. The dynamometer is vertically presented and in front of the forearm to maintain the standardized positions.
  • Test: regulation of the zero by making a vacuous acquisition during 5 seconds. The order "Jamar then Labin" or Labin then Jamar is chosen by randomization. The operator encourages the subject to every test. The subject does not see the curve displaying.
  • Test: 3 clampings of a handle of a dynamometer LABIN by the right hand during 5 seconds, with a break of 30 seconds between the tests. After 2 minutes of break the subjects make 3 clampings with a left hand. After that , the same session is realized with the dynamometer JAMAR

The order "Jamar then Labin" or Labin then Jamar is chosen by randomization. The operator encourages the subject to every test. The subject does not see the curve displaying.

Total duration of participation in the study for the patient: 30 minutes

Main analysis: the calculation of the coefficient of correlation (Pearson or Spearman with regard to the statistical distribution), coefficient of linen concordance and the coefficient of intra-classy correlation.

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Age : > 20 years and <60 years
  • Professionally active
  • Right-handed(right-wing) or left-handed person
  • Volunteer
  • Subjects have given its consent in writing to participate in the study

Exclusion criteria

  • Subjects with a trauma or surgery of the upper limb
  • Subjects with a diabetes
  • Subjects with a dysthyroïdie
  • Subjects with a neurological disease
  • Subjects with a syndrome canalaire of the upper limb
  • Subjects with a cutaneous disorders(confusions) of the cognitive upper

Treatment and study plan

Dynamometer LABIN

Device

LABIN is a prototype of dynamometer containing a handle of metal strength, connected with a sensor of pressure registering(recording) the applied strength and transmitting the data in a computer showing an IT curve of the strength according to time.

Primary outcomes

  1. Measure of the strength of prehension (in Kg).

    Time frame: at day 1

    measure of the strength of prehension in kilogram by using of the JAMAR ( hydraulic dynamometer), and the LABIN (electronic dynamometer provided with a sensor of pressure).

Secondary outcomes

  1. The comfort of the using the dynamometer LABIN

    Time frame: at day 1

    A note of discomfort on Simple Verbal Scale from 0 to 10 (max.) is attributed by the subject

  2. muscular endurance, estimated by dynamometer LABIN preservation of the strength for the LABIN

    Time frame: at day 1

    Change from baselin in measure muscular endurance (The calculation of the horizontal length under the graphic curve which will be exploited by the colleagues of the IUT)

Sponsors and collaborators

Lead sponsor

University Hospital, Clermont-Ferrand

Other

Collaborators

  • Department Physical Measures

Registry information

Acronym: DYNAVAL

Important dates

Study start
2016
Primary completion
2016
Study completion
2016
First posted
Jul 28, 2016
Registry last updated
Sep 19, 2019

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.