Validation of Software for Assessment of Atrial Fibrillation From Data Acquired by a Wearable Smartwatch

The goal of this study is to validate the Fitbit ECG software algorithm's ability to detect atrial fibrillation using data derived from a Fitbit wrist-worn consumer device that features a single lead ECG. The performance of the software application will be measured against centralized Core Lab physician-adjudicated 12-lead ECG tracings from FDA cleared ECG monitors as reference standards. The Fitbit software is intended as a pre-screening technology. It should identify candidates who may benefit from further evaluation. The software is not intended to be a diagnostic system.

Completed

Looking for future studies?

Notify Me

Key information

Age range

22 year and older

Sex eligibility

All sexes

Study type

Observational

Primary location

Onassis Cardiac Surgery Center, Kallithea, Greece

Loading trial locations.

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Women and men 22 years of age or older
  • Capable of giving informed consent
  • Sufficient manual dexterity to capture an ECG from the Fitbit wrist-worn device.
  • Subjects in the atrial fibrillation (AF) cohort must have a known history of AF and must be in -AF at the time of enrollment based on the screening ECG

Exclusion criteria

  • Anti-arrhythmic (rhythm control) medication use (such as amiodarone or flecainide)
  • Cardiac pacemaker or implantable cardioverter-defibrillator
  • History of a life-threatening rhythm such as ventricular tachycardia or ventricular fibrillation
  • Subjects in the sinus rhythm (SR) cohort should not have any history of atrial fibrillation and should not be in atrial fibrillation or other atrial arrhythmia at the time of enrollment based on the screening ECG

Treatment and study plan

Electrocardiogram

Diagnostic Test

Diagnostic Test: Simultaneous assessment of heart rhythm by 12-lead ECG and wearable device characterization of heart rhythm.

Primary outcomes

  1. Evaluate the software algorithm's ability to classify a heart rhythm as sinus rhythm or atrial fibrillation from a waveform equivalent to a Lead I ECG.

    Time frame: Through study completion, an average of 4 months

    Sensitivity and specificity of the software algorithm in classifying sinus rhythm and atrial fibrillation from the equivalent of a Lead I ECG waveform, compared with physician's interpretation of a standard 12-lead ECG. Performance goals are 90% (sensitivity) and 92% (specificity)

Secondary outcomes

  1. Confirm the software's ability to produce a waveform clinically equivalent to a Lead I ECG from a clinical reference device (12-lead ECG).

    Time frame: Through study completion, an average of 4 months

    Qualitative assessment of the software produced waveform to reference Lead I from 12-lead ECG. Goal is >0.80 of strips appear to overlay to the unaided eye.

    Comparison of R-wave amplitudes between the Fitbit waveform and the clinical reference. Goal is >0.80 of R-wave amplitudes to be within 2mm.

Sponsors and collaborators

Lead sponsor

Fitbit LLC

Industry

Registry information

Important dates

Study start
2019
Primary completion
2019
Study completion
2020
First posted
Nov 25, 2019
Registry last updated
Apr 9, 2020

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.