CHU Sainte Justin Hospital
Montreal, Quebec, H3T 1C5, Canada
Location status: Recruiting
NCT Number: NCT04627766
Small connected devices monitoring vital signs do not exist in children although they could be very useful to monitor patients once they have left the pediatric intensive care unit (PICU) in order to early identify patients at risk of PICU readmission.
The main objective of this study is to assess the accuracy of a connected thoracic patch (VT-Patch) for the continuous monitoring of vital signs in PICU patients.
This monitoring device will be used to monitor children in PICU for 8 hours and results will be compared to the standard of care in the unit which is the Philips monitoring system.
The investigators will assess the device's accuracy for the monitoring of 4 vital signs (SpO2, heart rate, respiratory rate and temperature) and perform exploratory assessment of three additional parameters (EKG, blood pressure and movement capture). The skin colour of each participant will be categorized based on Fitzpatrick scale and non-invasive spectroscopic method.
Interested in participating?
Request Info0 year–18 year
All sexes
Interventional
Not applicable
Montreal, Quebec, H3T 1C5, Canada
Location status: Recruiting
Introduction:
Patients readmission to PICU is a major point of concern for physician because, although being a rare event, it has been strongly associated to more morbidity and death. Patients monitoring once they have left the PICU is drastically modified and small connected devices could be an option to early identify patients at risk of PICU readmission. Such devices have been manufactured for the monitoring of adults but they do not exist for children.
Besides, studies have recently shown that devices that use optical sensors may be less accurate in individuals with dark skin pigmentation since dark skin tones impose a limitation on optical biometric sensing. The wearable and medical device industry has not sufficiently addressed this issue yet. This is one of the main challenges that need to be addressed in this area to make sure the connected devices using this technology will work with all different skin colours including very dark skin.
Main objective:
Assess the accuracy of VT-Patch to measure oxygen saturation (SpO2) in children with diverse skin colour.
Secondary objectives:
Assess the accuracy of VT-Patch monitoring for 3 additional physiologic parameters (heart rate (HR), respiratory rate (RR) and temperature). Exploratory assessment of blood pressure (BP) monitoring, ECG signal and movement capture. The effect of the skin colour on the vital signs measurements will be investigated as well.
Design:
Prospective validation study of VT-Patch device for the monitoring of 4 physiologic parameters and exploratory assessment of three additional parameters.
Methods:
The study will be performed in two successive phases:
4 physiologic parameters (HR, SpO2, RR, temperature and BP) and ECG will be prospectively recorded during 8 hours with VT-Patch and the regular monitoring system implemented in the unit High-resolution database that will be the Gold standard for evaluation.
Participating center: 1 center (St. Justine's Hospital).
Perspectives: This validation step of the collected data is necessary to validate that VT-Patch is suitable for pediatric patient monitoring of all skin colors.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Patients will be monitored for 8 consecutive hours and vital signs will be recorded with VT-Patch and with the standard of care
Time frame: 8 hours
Percentage of the difference between the absolute value of SpO2 recorded by VT-Patch and the one recorded using standard of care
Time frame: 8 hours
Percentage of the difference between the absolute value of heart rate recorded by VT-Patch and the one recorded using the standard of care
Time frame: 8 hours
Percentage of the difference between the absolute value of respiratory rate recorded by VT-Patch and the one recorded using standard of care
Time frame: 8hours
Percentage of the difference between the absolute value of temperature recorded by VT-Patch and the one recorded using standard of care
Contact information is provided by the study sponsor or research team.
St. Justine's Hospital
Other
Validation of Continuous Monitoring of Vital Signs in Children With VTPatch Connected Device in the Pediatric Intensive Care Unit
Acronym: VITALPICU
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07322731
Dyssomnias, Female Urogenital Diseases and Pregnancy Complications
Van, Turkey, Turkey (Türkiye)
View Trial DetailsNCT06427564
Vital Signs
Milan, Italy
View Trial DetailsNCT06587321
Palliative Care, Remote Monitoring
Dublin, Ireland
View Trial DetailsNCT05451875
Children, Heart Rate
Montreal, Quebec, Canada
View Trial Details