NCT Number: NCT03566888

Validation of a Noninvasive Automated Blood Pressure Device

The investigator's objective is to validate the novel automated blood pressure device against the requirements set forth in the AAMI_ISO 81060-2:2010 standard in voluntarily consented study participants.

Completed

Looking for future studies?

Notify Me

Key information

Conditions

Age range

12 year–92 year

Sex eligibility

All sexes

Study type

Observational

Primary location

University of Southern California

Los Angeles, California, 90033, United States

About this study

A new prototype blood pressure cuff has been designed for the higi SH llc health stations. This new cuff accurately captures blood pressure (BP) by emulating a fitted cuff which is used in conjunction with a mercury sphygmomanometer, the accepted gold standard for measuring BP values in most clinical facilities. The cuff is driven by a motor and gearhead to automatically "wrap" the cuff around a user's arm when they initiate a BP test. A BP measurement is then taken by an Original Equipment Manufacturer (OEM) BP module using an algorithm developed for use with a fitted cuff; the same module one would expect to find in a clinic or doctor's office. For the device to pass the AAMI_ISO 81060-2:2010 standard statistical requirements and thus, be considered a valid device for measuring BP it will need to meet the standards criteria 1 and 2.

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • between the ages of 12 to 92 years of age.
  • generally healthy but can be on medication for blood pressure regulation.

Exclusion criteria

  • heart disease is present such as, coronary artery disease, cardiomegaly, irregular heart rhythm, atrial fibrillation, heart valve disease, congenital heart disease, cardiomyopathy, pericardial effusion, Marfan syndrome, and heart murmurs.
  • using a pacemaker to maintain a suitable heart rate.
  • missing their natural left arm.
  • exhibit a musculoskeletal disorder that may prevent them from sitting upright for a period of time (~40min) or prevent a BP reading taken from the left arm.
  • Special populations, for example pregnant women and patients with known arrhythmias.

Treatment and study plan

Blood pressure assessment

Device

Validation of an automated blood pressure device

Primary outcomes

  1. Systolic and Diastolic blood pressure (criteria 1 and 2)

    Time frame: 60 minutes

    Mean value of the differences of the paired (i.e., device under test - reference method) determinations for all study participants shall be within or equal to ± 5 mmHg with a standard deviation no greater than 8 mmHg (criterion 1). Standard deviation of the averaged paired determinations per participant shall meet the criterion 2 listed in the standard.

Sponsors and collaborators

Lead sponsor

University of Southern California

Other

Registry information

Important dates

Study start
2017
Primary completion
2018
Study completion
2018
First posted
Jun 25, 2018
Registry last updated
Nov 8, 2018