Royal Children's Hospital
Parkville, Victoria, 3052, Australia
Location status: Recruiting
NCT Number: NCT06647225
The goal of this clinical trial is to learn if an ultrasound scan using artificial intelligence can accurately screen for hip dysplasia. Researchers will compare the artificial intelligence ultrasound results to the standard ultrasound measures to see if the artificial intelligence ultrasound scan can accurately screen for hip dysplasia.
It will also seek to understand how parents feel about their children undergoing this scan.
Participants will:
* Have an additional ultrasound performed on their child at their scheduled outpatient's appointment for hip dysplasia * Complete a short questionnaire about the experience of having the measurement performed on their child
Interested in participating?
Request Info4 week–20 week
All sexes
Interventional
Not applicable
Parkville, Victoria, 3052, Australia
Location status: Recruiting
Initial screening for Developmental Dysplasia of the Hip (DDH) in Australia is performed most often by general practitioners and paediatricians shortly after birth and by maternal child health care nurses (MCHN) throughout the first year of life. These physical examinations consist of the Ortolani and Barlow tests and the examination of the thigh and gluteal creases. A recent meta-analysis reported the sensitivity of these tests as 36%, which indicates there is potential for a large proportion of cases to go undetected when solely relying on these examinations. Moreover, there currently are no formalised processes by which standards of practice are taught, assessed, or maintained. Thus, there is a clear need for a less operator-dependent screening protocol that can be performed within the current models of infant care. While some countries utilise universal ultrasound screening, this too is limited by access to care, as devices are not portable and thus cannot be used in current care models. Furthermore, it requires a specialist operator, substantially increasing cost. The screening program's limited nature, combined with the need for more consensus among international healthcare providers regarding the best method for managing DDH, has produced highly mixed clinical practices.
One part of the solution is optimising screening protocols for DDH in existing care models. Each state in Australia has established MCHN care protocols that provide access care for young children. While physical screening for DDH in these visits is standard practice, there remains considerable scope for improvement in the accuracy and reliability of these screening methods. Selective screening relies on several clinical associations with DDH to identify which patients receive ultrasound screening. Still, it has been shown to detect only 50% of infants with dysplasia. The MCHN screening program relies on clinical examination alone to detect dysplasia, an inferior identification method. Universal screening has a higher rate of detection of dysplasia but is expensive, single point in time (so misses the development of dysplasia) and results in higher levels of treatment.
A possible solution is portable artificial intelligence (AI)-augmented ultrasound. Recently technology has been developed to support a portable ultrasound device to screen DDH that uses AI-enabled technology to screen for DDH rapidly and accurately. Prior data has demonstrated that physicians and nurses could operate the device following training from expert sonographers. With its low-cost and ease of operation (with simple training) by healthcare providers such as MCHNs, it could significantly augment the physical screening. Thus, there is clear potential for an affordable, repeatable, and accessible screening methodology to be translated into clinical care. Initial Canadian data is promising. Pilot data suggests that DDH detection rates with this technology is on par with the detection rates of orthopaedic specialists. However, as this study was performed in a community setting and only those participants referred to orthopaedic clinics had a standard ultrasound measure performed, this pilot was unable to compare this screening technique with current gold standard diagnostic measures across the whole cohort, nor determine device sensitivity or predictive values. To demonstrate that this technology is fit for purpose, it is imperative that the rate of false negatives is also understood, as this is what will lead to late presentation, - which is what screening ultimately endeavours to prevent. Moreover, in an Australian context an important consideration in a wider roll-out is whether this technology would be accepted for uptake by clinicians and parents.
The proposed project will seek to gather pilot data to assess the validity and feasibility of this technology within a population of infants aged 4-20 weeks flagged at risk for DDH and referred to the Royal Children's Hospital. This will enable the recruitment of a sufficient number of cases of DDH to determine the sensitivity of the device. While the sensitivity and specificity of the device in this at-risk population may not be generalizable to the wider community the information gathered will then inform and refine a larger study of this technology in a community setting such as tertiary (birthing hospitals) and primary (MCHN clinics) care. If it can be demonstrated that it is feasible to implement this technology into existing care models, there is clear scope for this technology to revolutionize DDH screening. Thus, this project seeks to determine how well the device performs (sensitivity, specificity and predictive value) and the the clinical acceptability of this measure within the patient population.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Participants will be excluded from enrolment if:
The hip ultrasound is performed using a handheld device (Exo Iris) that uses a pocket-sized ultrasound probe and is run through an application on an IoS (Apple mobile) operation system. . A real-time algorithm detects and records the anatomical landmarks. When there are enough images for analysis the operator is notified that the scan is complete.
Time frame: 1 day, both ultrasound scans will be performed on the same day
Artificial intelligence (AI) augmented ultrasound results will be compared to standard ultrasound imaging to calculate sensitivity ([number of true positive cases detected/(number of true positive cases detected + number of false negative cases detected)] X 100). Groups will be defined as follows:
Time frame: 1 day, both ultrasound scans will be performed on the same day
Artificial intelligence (AI) augmented ultrasound results will be compared to standard ultrasound imaging to calculate specificity ([number of true negative cases detected/(number of false positive cases detected + number of true negatives cases detected] X 100). Groups will be defined as follows:
Time frame: 1 day, both ultrasound scans will be performed on the same day
Artificial intelligence (AI) augmented ultrasound will be compared to standard ultrasound to calculate PPV ([number of true positive cases detected/(number of true positive cases detected + number of false positive cases predicted) X 100). Groups will be defined as follows:
Time frame: 1 day, both ultrasound scans will be performed on the same day
Artificial intelligence (AI) augmented ultrasound will be compared to standard ultrasound to calculate NPV [number of true negative cases detected/(number of false negatives detected + number of true negative cases detected) X 100. Groups will be defined as follows:
Time frame: 1 day, all data will be collected from day of scan
Sensitivity will be calculated between groups:
Time frame: 1 day, all data will be collected from day of scan
Analyses will stratified to look at differences in specificity between groups:
Time frame: 1 day, all data will be collected from day of scan
Positive predictive value will be compared between groups:
Time frame: 1 day, all data will be collected from day of scan
Negative predictive value will be compared between groups:
Time frame: 12 months or entire study duration
Device operators' reliability will be recorded as percentage of scans performed that return a suboptimal result. This will be done by graphing number of scans performed by operators (operator experience) (x axis) against proportion of sub-optimal scans (y axis) to visually identify if a steady state is achieved.
Time frame: 12 months or entire study duration
The total proportion of infants unable to be scanned with the Artificial Intelligence augmented ultrasound device and reasons why scans were unsuccessful will from the entire sample. A higher frequency of successful scan acquisition will indicate better device performance.
Time frame: 1 day, calculated at time of scan
Time to acquire the image will be calculated from the initiation of the scan to the time that the software indicates image acquisition is complete. The time to receive results will be calculated from the time of completion acquisition to the time the final recommendation is provided. A lower successful scan time will be indicative of higher feasibility.
Time frame: 1 day, caregivers will be asked to complete immediately following the scan
Caregivers will be asked to answer a purpose-built survey that has been piloted in Canadian studies (3 questions rated from 0-10, where 10 indicates a more positive experience) in addition to Australian-specific closed and open-ended questions.
Time frame: At the conclusion of their involvement in the study device (up to 12 months)
Operators will be asked to complete the 10-item System Usability Questionnaire which measures the perceived ease of using technological devices. Scores are calculated on a 5-point Likert scale where 1=Strongly disagree and 5=Strongly agree. A single composite score out of 100 is calculated from all 10 items and indicates the overall useability of the device, where a higher score indicates better useability. In addition to this, two open-ended questions (Are there any further comments you would like to make about what you liked about the device?" and "are there any further comments you would like to make about what you didn't like about the device?") will be asked.
Contact information is provided by the study sponsor or research team.
Melissa Formosa
CONTACT
Natalie K Hyde, BBiomedSc (Hons), PhD
CONTACT
Murdoch Childrens Research Institute
Other
Artificial Intelligence Augmented Ultrasound for Developmental Dysplasia of the Hip: a Validity Study
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07066852
Developmental Dysplasia of Hip, Hip Dislocation
Ankara, Turkey (Türkiye)
View Trial DetailsNCT07581717
Acetabular Dysplasia, Congenital Abnormalities
Istanbul, Turkey (Türkiye)
View Trial DetailsNCT04795843
Behavior, Congenital Abnormalities
Aarhus N, Denmark
View Trial DetailsNCT04090359
Dislocation, Hip, Hip Dislocation
New York, United States
View Trial Details