Completed

NCT Number: NCT03033849

User Performance Evaluation of Contour® Next One, Accu-Chek® Aviva Connect, FreeStyle Freedom Lite, OneTouch® Verio and GlucoMen® Areo Blood Glucose Monitoring Systems Following ISO 15197:2013

The user performance evaluation shows whether people with diabetes are able to obtain accurate measurement results with a blood glucose monitoring system. In this study, user performance evaluation will be performed for Contour® Next One (Ascensia Diabetes Care GmbH), Accu-Chek® Aviva Connect (Roche Diabetes Care GmbH), FreeStyle Freedom Lite (Abbott Diabetes Care Inc.), OneTouch® Verio (LifeScan Europe) and GlucoMen® areo A. Menarini Diagnostics S.r.l. (based on ISO 15197:2013; EN ISO 15197:2015, clause 8.

For each BGMS, measurement procedures for user performance evaluation will be performed with 1 test meter and 1 reagent system lot by the study subjects.

The same meter and an additional test meter will be used for double measurements performed by study personnel (with the same reagent system lot used by subjects).

Completed

Looking for future studies?

Notify Me

Key information

Age range

18 year and older

Sex eligibility

All sexes

Study type

Interventional

Phase

Not applicable

Primary location

Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm

Ulm, Baden-Wurttemberg, 89081, Germany

Who can participate

Healthy volunteers accepted: Yes

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Male or female, with type 1 diabetes, type 2 diabetes or subjects without diabetes
  • Signed informed consent form
  • Minimum age of 18 years
  • Subjects are legally competent and capable to understand character, meaning and consequences of the study.
  • If blood glucose values < 80 mg/dl or > 300 mg/dl shall be measured after short term alteration in insulin therapy:
  • Male or female with type 1 diabetes and intensified insulin therapy or insulin pump therapy.
  • Signature of subjects to document consent with these procedures on informed consent form.

Exclusion criteria

  • Pregnancy or lactation period
  • Severe acute disease (at the study physician's discretion)
  • Severe chronic disease with potential risk during the test procedures (at the study physician's discretion)
  • Current constitution that compromises the subject's capability to participate in the study (at the study physician's discretion)
  • Being unable to give informed consent
  • < 18 years
  • Legally incompetent
  • Being committed to an institution (e.g. psychiatric clinic)
  • Language barriers potentially compromising an adequate compliance with study procedures
  • Dependent on investigator or sponsor
  • If blood glucose values < 80 mg/dl shall be measured after short term alteration in insulin therapy, subjects with type 1 diabetes, suffering from:
  • Coronary heart disease
  • Condition after myocardial infarction
  • Condition after cerebral events
  • Peripheral arterial occlusive disease
  • Hypoglycemia unawareness

Treatment and study plan

Blood glucose monitoring system for self-testing Contour® Next One

Device

This is a user performance study, that is intended to show whether people with diabetes are able to obtain accurate measurement results with a blood glucose monitoring system (BGMS).

The subjects' measurement technique (e.g. applying blood onto the system reagent, reading the result) will be observed and documented by study personnel.

Immediately after the measurement by the subject, the study personnel will perform a measurement with the same test meter.

Measurement results obtained by the subjects with the BGMS will be compared to measurement results obtained with the comparison method.

Sample collection for the comparison measurement will be performed by study personnel within 5 min after the subject's measurement with the test meter.

The samples will be collected from the subjects' skin puncture (if possible), an additional skin puncture (if required) will be documented.

An additional skin puncture will be performed by study personnel with single-use lancing devices.

Other names: Contour® Next One, Accu-Chek® Aviva Connect, FreeStyle Freedom Lite, OneTouch® Verio, GlucoMen® areo

Blood glucose monitoring system for self-testing Accu-Chek® Aviva Connect

Device

Blood glucose monitoring system for self-testing FreeStyle Freedom Lite

Device

Blood glucose monitoring system for self-testing OneTouch® Verio

Device

Blood glucose monitoring system for self-testing GlucoMen® areo

Device

Primary outcomes

  1. Acceptance criteria defined by ISO 15197:2013 (E) will be applied

    Time frame: For each subject, the experimental phase has an expected duration of up to 3 hours

    95 % of the individual glucose measured values shall fall within ± 15 mg/dl (0.83 mmol/l) of the measured values of the comparison measurement procedure at glucose concentrations < 100 mg/dl (5.55 mmol/l) and within ± 15 % at glucose concentrations ≥ 100 mg/dl (5.55 mmol/l).

Sponsors and collaborators

Lead sponsor

Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm

Other

Collaborators

  • Ascensia Diabetes Care

Registry information

Important dates

Study start
2017
Primary completion
2017
Study completion
2017
First posted
Jan 27, 2017
Registry last updated
Sep 12, 2017

OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.

View the official ClinicalTrials.gov record (opens in a new tab)

This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.

Published trials that share one or more normalized conditions with this study.