Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm
Ulm, Baden-Wurttemberg, 89081, Germany
NCT Number: NCT03033849
The user performance evaluation shows whether people with diabetes are able to obtain accurate measurement results with a blood glucose monitoring system. In this study, user performance evaluation will be performed for Contour® Next One (Ascensia Diabetes Care GmbH), Accu-Chek® Aviva Connect (Roche Diabetes Care GmbH), FreeStyle Freedom Lite (Abbott Diabetes Care Inc.), OneTouch® Verio (LifeScan Europe) and GlucoMen® areo A. Menarini Diagnostics S.r.l. (based on ISO 15197:2013; EN ISO 15197:2015, clause 8.
For each BGMS, measurement procedures for user performance evaluation will be performed with 1 test meter and 1 reagent system lot by the study subjects.
The same meter and an additional test meter will be used for double measurements performed by study personnel (with the same reagent system lot used by subjects).
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Ulm, Baden-Wurttemberg, 89081, Germany
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
This is a user performance study, that is intended to show whether people with diabetes are able to obtain accurate measurement results with a blood glucose monitoring system (BGMS).
The subjects' measurement technique (e.g. applying blood onto the system reagent, reading the result) will be observed and documented by study personnel.
Immediately after the measurement by the subject, the study personnel will perform a measurement with the same test meter.
Measurement results obtained by the subjects with the BGMS will be compared to measurement results obtained with the comparison method.
Sample collection for the comparison measurement will be performed by study personnel within 5 min after the subject's measurement with the test meter.
The samples will be collected from the subjects' skin puncture (if possible), an additional skin puncture (if required) will be documented.
An additional skin puncture will be performed by study personnel with single-use lancing devices.
Other names: Contour® Next One, Accu-Chek® Aviva Connect, FreeStyle Freedom Lite, OneTouch® Verio, GlucoMen® areo
Time frame: For each subject, the experimental phase has an expected duration of up to 3 hours
95 % of the individual glucose measured values shall fall within ± 15 mg/dl (0.83 mmol/l) of the measured values of the comparison measurement procedure at glucose concentrations < 100 mg/dl (5.55 mmol/l) and within ± 15 % at glucose concentrations ≥ 100 mg/dl (5.55 mmol/l).
Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm
Other
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT01934660
Cardiovascular Disease, Cardiovascular Diseases
Bethesda, Maryland, United States
View Trial DetailsNCT00038727
Body Weight, CVD
Rockville, Maryland, United States
View Trial DetailsNCT06029517
Arterial Occlusive Diseases, Arteriosclerosis
Buffalo, New York, United States
View Trial DetailsNCT07728162
Diabetes Mellitus, Disease Attributes
Bolu, Turkey (Türkiye)
View Trial Details